Marie Lambrecht1, Jan-Jakob Sonke2, Ursula Nestle3,4, Heike Peulen1, Damien C Weber5, Marcel Verheij2, Coen W Hurkmans1.
Abstract
BACKGROUND ANDEntities:
Keywords: 4DCT; Clinical trial; EORTC; Lung cancer; Radiotherapy quality assurance; Tumour motion
Year: 2018 PMID: 33458418 PMCID: PMC7807606 DOI: 10.1016/j.phro.2018.10.003
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Scanner models and acquisition parameters used in the 11 institutions participating in the Lungtech trial RTQA site visit.
| Institute | Scanner brand | FOV Acquisition diameter (mm)# | FOV Reconstruction diameter (mm) | bin technique & | kVp+ (V) | Expos-ure (mA/s) | Rotation speed (s) | Slice thickness (mm) | Scan length (mm) | CTDI vol(mGy) | Tube current (mA) | Pitch * |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| A | Toshiba Acquilion/LB | 550 | 550 | 10, P | 120 | 76 | 0.5 | 2 | 170 | NA | 250 | 0.12 |
| 170 | NA | 250 | 0.12 | |||||||||
| 170 | NA | 250 | 0.075 | |||||||||
| B | Philips Brillance big bore | 600 | 480 | 10, A | 120 | 600 | 0.5 | 2 | 162 | NA | 180 | 0.15 |
| 162 | NA | 132 | 0.11 | |||||||||
| 162 | NA | 90 | 0.075 | |||||||||
| C | Philips Gemini TF Big Bore | 600 | 600 | 10, P | 120 | 250 | 0.5 | 2 | 180 | 13.2 | 45 | 0.09 |
| 180 | 13.2 | 45 | 0.09 | |||||||||
| 180 | 13.2 | 45 | ||||||||||
| D | Philips Brilliance big bore | 600 | 568 | 10, A | 120 | 801 | 0.44 | 3 | 240 | NA | 74 | 0.08 |
| 240 | NA | 74 | 0.041 | |||||||||
| 240 | NA | 74 | 0.041 | |||||||||
| E | Siemens Sensation open | 500 | 500 | 8, P | 120 | 400 | 0.5 | 1.5 | 79.5 | NA | 80 | 0.16 |
| 79.5 | NA | 80 | 0.1 | |||||||||
| 120 | NA | 40 | ||||||||||
| F | Siemens sensation open | 500 | 500 | 8, P | 120 | 400 | 0.5 | 3 | 186 | 35.6 | 80 | 0.1 |
| 189 | 35.6 | 80 | 0.1 | |||||||||
| 276 | 36.0 | 40 | ||||||||||
| G | Philips Brilliance big bore | 600 | 568 | 10, A | 120 | 801 | 0.44 | 3 | 171 | 47.4 | 74 | 0.08 |
| 171 | 47.4 | 132 | 0.041 | |||||||||
| 171 | 47.4 | 90 | 0.041 | |||||||||
| H | Philips Brilliance big bore | 600 | 480 | 10, A | 120 | 600 | 0.5 | 2 | 172 | 35.5 | 180 | 0.15 |
| 172 | 35.5 | 132 | 0.11 | |||||||||
| 172 | 25.5 | 90 | 0.075 | |||||||||
| I | Philips Brilliance big bore | 600 | 500 | 10, P | 120 | 350 | 0.5 | 2 | 226 | 20.7 | 105 | 0.15 |
| 226 | 20.7 | 53 | 0.11 | |||||||||
| 226 | 20.7 | 77 | ||||||||||
| J | Toshiba Acquilion/LB | 550 | 550 | 10, P | 120 | 46 | 0.33 | 2 | 161 | 81.3 | 150 | 0.075 |
| 161 | 108.4 | 150 | 0.075 | |||||||||
| 161 | 108.4 | 150 | ||||||||||
| K | Siemens Somatom Definition AS | 500 | 500 | 10, P | 120 | 80 | 0.5 | 3 | 176 | NA | 160 | NA |
| 176 | NA | 160 | NA | |||||||||
| 176 | NA | 80 | NA |
*Where 3 values are given, different values are used for breathing periods of 3, 4 and 6 s, respectively. The pitch values that are larger than pmax are given in bold. NA: Not available. $Number of bins are given and type of binning: P = phase binning, A = amplitude binning, #FOV = field of view, +KVP = kilovolt peak
Fig. 1Boxplot of volume deviation expressed in % of the true volume per sphere and motion type (minus signs mean an underestimation).
Mean and range deviations in volume (in % of the true volume) and amplitude (in cm) across institutions per sphere size and motion tested (minus signs mean an underestimation).
| Motion | Sphere radius (mm) | Volume deviations (% of true volume) | Amplitude deviations (cm) | |||
|---|---|---|---|---|---|---|
| End of inspiration | End of expiration | Mid-ventilation | Average CT | |||
| I: 15 mm/3 s | 7.5 | 14.1% | 2.2% | 16.2% | 43.2% | −0.08 |
| (−5.6, 40.8) | (−8.4, 20.2) | (3.3, 35.4) | (22.1, 66.4) | (−0.25, −0.01) | ||
| 12.5 | 7.7% | 1.1% | 11.3% | 41.4% | −0.08 | |
| (−2.3, 33.9) | (−1.6, 5.6) | (3.5, 36.0) | (26.3, 62.4) | (−0.26, −0.01) | ||
| II: 15 mm/6 s | 7.5 | 3.8% | 1.0% | 11.2% | 12.4% | −0.18 |
| (−98.7, 39.7) | (−1.5, 6.0) | (−33.8, 29.9) | (−33.8, 36.3) | (−0.60, 0.01) | ||
| 12.5 | 21.5% | 2.1% | 13.2% | 31.8% | −0.13 | |
| (−3.5, 68.5) | (−1.5, 3.5) | (0.1, 28.9) | (13.7, 45.9) | (−0.28, −0.01) | ||
| III: 25 mm/4 s | 7.5 | 21.3% | −1.0% | 21.4% | 22.8% | −0.14 |
| (−13.8, 64.8) | (−12.9, 5.9) | (−3.9, 51.1) | (−2.1, 44.6) | (−0.35, −0.02) | ||
Fig. 2Coronal views of the phantom with the 7.5 mm radius sphere and the 15 mm/6 sec motion II at (a) static (b) end of expiration phase and (c) end of inspiration phase.
P-values of the Wilcoxon test between the volume deviations observed in each chosen phase.
| Motion | Sphere radius (mm) | End of inspiration/End of expiration | Mid-ventilation/End of expiration | End of inspiration/Mid-ventilation |
|---|---|---|---|---|
| I: 15 mm/3 s | 7.5 | p > 0.05 | ||
| II: 15 mm/6 s | p > 0.05 | p > 0.05 | p > 0.05 | |
| III: 25 mm/4 s | p > 0.05 | |||
| I: 15 mm/3s | 12.5 | p > 0.05 | ||
| II: 15 mm/6s | p > 0.05 |
Fig. 3Deviation of the peak to peak amplitude (minus signs mean an underestimation of the amplitude).