Marie Lambrecht1, Christos Melidis2, Jan-Jakob Sonke3, Sonja Adebahr4,5, Ronald Boellaard6, Marcel Verheij7, Matthias Guckenberger8, Ursula Nestle9,10, Coen Hurkmans11.
Abstract
BACKGROUND: The EORTC has launched a phase II trial to assess safety and efficacy of SBRT for centrally located NSCLC: The EORTC 22113-08113-Lungtech trial. Due to neighbouring critical structures, these tumours remain challenging to treat. To guarantee accordance to protocol and treatment safety, an RTQA procedure has been implemented within the frame of the EORTC RTQA levels. These levels are here expanded to include innovative tools beyond protocol compliance verification: the actual dose delivered to each patient will be estimated and linked to trial outcomes to enable better understanding of dose related response and toxicity.Entities:
Mesh:
Year: 2016 PMID: 26791788 PMCID: PMC4721147 DOI: 10.1186/s13014-015-0567-5
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Combinations of treatment planning systems, treatment devices and motion management options present within the institutions in which a site-visit has been performed. For some centres several combinations are possible according to their facility questionnaire and preferred combination will be known during the site visit
| Treatment planning system | Treatment device | 4D option* |
|---|---|---|
| Accuray Multiplan | Accuray Cyberknife | Tracking (1) |
| Brainlab Iplan | Varian Novalis | Gating (1) |
| Varian Eclipse | Varian Truebeam | ITV (3) |
| Varian Eclipse | Varian Clinac | ITV (2) |
| Elekta Monaco | Elekta Synergy | ITV (1) |
| Accuray Tomotherapy | Accuray Tomo HD | ITV (1) |
| Philips Pinnacle | Elekta Synergy | ITV (1) |
* number of centres using this combination is given between brackets
Trial dose specification
| Ideal solution 8 × 7.5 Gy | Reduced prescription to meet OARs constraints 8 × 7 Gy | Underdosed PTV to meet OARs constraints 8 × 7.5 Gy | ||||
|---|---|---|---|---|---|---|
| Volume | 99 % | 95 % | 99 % | 95 % | 99 % | 95 % |
| PTV | ≥54 Gy (90 %) | ≥60 Gy (100 %) | ≥50.4 Gy (90 %) | ≥56 Gy (100 %) | NA | ≥48 Gy (80 %) |
| CTV | NA | NA | ≥54 Gy (90 %) ≥60 Gy (100 %) | |||
| GTV | NA | NA | Acceptable variation | |||
| 66–78Gy (110–130 %) | ||||||
Fig. 1a CIRS 008A body - b rod simulating a breathing lung - c 15 mm diameter sphere ionization chamber insert - d 15 mm diameter sphere film insert
Fig. 2a Customized PET CIRS phantom plugged on the breathing platform - b PET rod simulating the lung compartment - c 15 mm diameter sphere simulating a lung lesion
Combination of breathing parameters
| Test 1: Spheres | Test 2: Spheres | Test 3: Sphere | |
|---|---|---|---|
| 15 and 25 mm | 15 and 25 mm | 15 mm | |
| Breathing Period (s) | 6 | 3 | 4 |
| Peak to peak Amplitude (mm) | 15 | 15 | 25 |
PET stored parameters
| PET | Static | 15 mm/3 s | 15 mm/6 s | 25 mm/4 s |
|---|---|---|---|---|
| Type | 3D | 4D:10.phases | 4D:10 phases | 4D:10 phases |
| Acquisition time | 12:30 | 12:48 | 13:04 | 13:20 |
| TOF | YES | |||
| Binning type | X | Phase binning | ||
| Number of bed position | 2 | 1 | ||
| Overlap bed | 90 mm | |||
| Time per bed position | 2 min | 10 min | ||
| Data corrected for: | decay corrected : aquisition start time | |||
| attenuation corrected | ||||
| scatter corrected | ||||
| dead time corrected | ||||
| randoms corrected | ||||
| non-uniform radial sampling corrected | ||||
| detector normalizationu | ||||
| Type of reconstruction algorithm | BLOB-OS-TF | |||
| Energy window | 440–665 KeV | |||
| Slice thickness | 4 mm | |||
| Image matrix | 144 × 144 | |||
| FOV | 903 mm | |||
Fig. 3CIRS phantom with the 15 mm sphere size animated with a 15 mm/6 s breathing signal (a) Superposition of 10 % and 40 % phases (the most distant sphere positions) PET-CT images - (b) Average PET-CT
Fig. 4Volume recovery coefficients VRC50 (VRC50 = VOI50/VOI static) expressed in % of VOI static. Single site results aquired on a Gemini TF PET-CT Big bore
Fig. 5Activity recovery coefficients (RCmax, RC50, RC70) of the 15 mm sphere calculated in static and dynamic conditions (static PET and 4D PET). Single site results aquired on a Gemini TF PET-CT Big bore
Fig. 6Average 4DCT of the CIRS phantom with the 25 mm sphere animated with a 15 mm/6 s breathing motion a transversal b sagittal c coronal
Fig. 7Volume variation induced by the three breathing motions and expressed in per cent of the static volume for the 15 mm sphere. Single site results aquired on a Gemini TF PET-CT Big bore using a phase binning approach
True peak to peak amplitude of the 15 mm diameter sphere versus measured on CT from one of the participating institutions
| LAT | ANT-POST | SUP-INF | ||||
|---|---|---|---|---|---|---|
| true | Measured | true | Measured | true | Measured | |
| A:15 mm | 0 | 0.02 | 0 | 0.04 | 1.5 | 1.4 |
| C:3 s | ||||||
| A:15 mm | 0 | 0.06 | 0 | 0.01 | 1.5 | 0.87 |
| C:6 s | ||||||
| A:25 mm | 0 | 0.07 | 0 | 0.07 | 2.5 | 2.21 |
| C:4 s | ||||||