Clinical evaluations of an amplitude-based binning algorithm for 4DCT reconstruction in radiation therapy.

Hua Li1, Camille Noel, Jose Garcia-Ramirez, Daniel Low, Jeffrey Bradley, Clifford Robinson, Sasa Mutic, Parag Parikh.   

Abstract

PURPOSE: Phase-binning algorithms are commonly utilized in 4DCT image reconstruction for characterization of tumor or organ shape and respiration motion, but breathing irregularities occurring during 4DCT acquisition can cause considerable image distortions. Recently, amplitude-binning algorithms have been evaluated as a potential improvement to phase-binning algorithms for 4DCT image reconstruction. The purpose of this study was to evaluate the performance of the first commercially available on-line retrospective amplitude-binning algorithm for comparison to the traditional phase-binning algorithm.
METHODS: Both phantom and clinical data were used for evaluation. A phantom of known geometry was mounted on a 4D motion platform programmed with seven respiratory waves (two computer generated and five patient trajectories) and scanned with a Philips Brilliance Big bore 16-slice CT simulator. 4DCT images were reconstructed using commercial amplitude- and phase-binning algorithms. Image quality of the amplitude- and phase-binned image sets was compared by evaluation of shape and volume distortions in reconstructed images. Clinical evaluations were performed on 64 4DCT patient image sets in a blinded review process. The amplitude- and phase-binned 4DCT maximum intensity projection (MIP) images were further evaluated for 28 stereotactic body radiation therapy (SBRT) cases of total 64 cases. A preliminary investigation of the effects of respiratory amplitude and pattern irregularities on motion artifact severity was conducted.
RESULTS: The phantom experiments illustrated that, as expected, maximum inhalation occurred at the 0% amplitude and maximum exhalation occurred at the 50% amplitude of the amplitude-binned 4DCT image sets. The phantom shape distortions were more severe in the images reconstructed from the phase-binning algorithm. In the clinical study, compared to the phase-binning algorithm, the amplitude-binning algorithm yielded fewer or less severe motion artifacts in 37.5% of the cases (24∕64), comparable artifacts in 54.7% of the cases (35∕64), and slightly greater artifacts in 7.8% of the cases (5∕64). Evaluation of SBRT cases demonstrated that the reconstructed tumor sizes and locations were comparable in 96% (1∕28) of the MIP image pairs generated from both amplitude- and phase-binning algorithms. In this case the amplitude-binned image set rendered a smaller tumor size, which was likely due to very shallow respiratory amplitudes occurring over several breathing cycles.
CONCLUSIONS: Overall, the amplitude-binning algorithm for 4DCT reconstruction reduced the severity of tumor distortion and image artifacts compared to the phase-binning algorithm. However, the full range of motion may not be characterized using amplitude-binning algorithms. Despite superior performance, amplitude binning can still be susceptible to motion artifacts caused by large variations in amplitude of respiratory waves.

Entities:  

Mesh:

Year:  2012        PMID: 22320802     DOI: 10.1118/1.3679015

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Objective function to obtain multiple representative waveforms for a novel helical CT scan protocol.

Authors:  Dan Ruan; David Thomas; Daniel A Low
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

2.  Image quality in thoracic 4D cone-beam CT: a sensitivity analysis of respiratory signal, binning method, reconstruction algorithm, and projection angular spacing.

Authors:  Chun-Chien Shieh; John Kipritidis; Ricky T O'Brien; Zdenka Kuncic; Paul J Keall
Journal:  Med Phys       Date:  2014-04       Impact factor: 4.071

3.  Assessing and accounting for the impact of respiratory motion on FDG uptake and viable volume for liver lesions in free-breathing PET using respiration-suspended PET images as reference.

Authors:  Guang Li; C Ross Schmidtlein; Irene A Burger; Carole A Ridge; Stephen B Solomon; John L Humm
Journal:  Med Phys       Date:  2014-09       Impact factor: 4.071

4.  A comparison of amplitude-based and phase-based positron emission tomography gating algorithms for segmentation of internal target volumes of tumors subject to respiratory motion.

Authors:  Shyam S Jani; Clifford G Robinson; Magnus Dahlbom; Benjamin M White; David H Thomas; Sergio Gaudio; Daniel A Low; James M Lamb
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-11-01       Impact factor: 7.038

5.  Characterization of optical-surface-imaging-based spirometry for respiratory surrogating in radiotherapy.

Authors:  Guang Li; Jie Wei; Hailiang Huang; Qing Chen; Carl P Gaebler; Tiffany Lin; Amy Yuan; Andreas Rimner; James Mechalakos
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

6.  Cyberknife® stereotactic radiation therapy for stage I lung cancer and pulmonary metastases: evaluation of local control at 24 months.

Authors:  Myriam Khadige; Julia Salleron; Vincent Marchesi; Guillaume Oldrini; Didier Peiffert; Véronique Beckendorf
Journal:  J Thorac Dis       Date:  2018-08       Impact factor: 2.895

7.  Comparative analysis of respiratory motion tracking using Microsoft Kinect v2 sensor.

Authors:  Evan Silverstein; Michael Snyder
Journal:  J Appl Clin Med Phys       Date:  2018-03-25       Impact factor: 2.102

8.  Ventilation Series Similarity: A Study for Ventilation Calculation Using Deformable Image Registration and 4DCT to Avoid Motion Artifacts.

Authors:  Geoffrey G Zhang; Kujtim Latifi; Vladimir Feygelman; Kuei-Ting Chou; Tzung-Chi Huang; Thomas J Dilling; Bradford A Perez; Eduardo G Moros
Journal:  Contrast Media Mol Imaging       Date:  2017-09-17       Impact factor: 3.161

9.  Commissioning a four-dimensional Computed Tomography Simulator for minimum target size due to motion in the Anterior-Posterior direction: a procedure and treatment planning recommendations.

Authors:  Marcus Sonier; Brandon Vangenderen; Dallas Visagie; Cameron Appeldoorn; Te-Chih Archie Chiang; Lindsay Mathew; Stefan Reinsberg; Jim Rose; Ramani Ramaseshan
Journal:  J Appl Clin Med Phys       Date:  2020-07-15       Impact factor: 2.102

10.  Quality assurance of four-dimensional computed tomography in a multicentre trial of stereotactic body radiotherapy of centrally located lung tumours.

Authors:  Marie Lambrecht; Jan-Jakob Sonke; Ursula Nestle; Heike Peulen; Damien C Weber; Marcel Verheij; Coen W Hurkmans
Journal:  Phys Imaging Radiat Oncol       Date:  2018-12-07
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.