| Literature DB >> 33937120 |
Milovan Savanović1,2, François Gardavaud3, Dražan Jaroš4, Bénédicte Lonkuta5, Matthias Barral6, François Henri Cornelis7, Jean-Noël Foulquier8.
Abstract
BACKGROUND: The use of imaging is indispensable in modern radiation therapy, both for simulation and treatment delivery. For safe and sure utilization, dose delivery from imaging must be evaluated.Entities:
Keywords: Computed Tomography; Cone-Beam Computed Tomography; Four-Dimensional Computed Tomography; Lung; Organs at Risk Exposure; Stereotactic Body Radiation Therapy
Year: 2021 PMID: 33937120 PMCID: PMC8064136 DOI: 10.31661/jbpe.v0i0.2009-1173
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Patient characteristics
| Characteristics | Raw data |
|---|---|
| 100 | |
| 70 (38 – 90) | |
| 58 | |
| 42 | |
| 68 | |
| 32 | |
| 32 (17 – 50) | |
| 37 (21 – 49) | |
| 63.5 (39.0 – 106.0) | |
| 1.68 (1.48 – 1.86) | |
| 23.0 (15.2 – 35.9) | |
| 4.3 (1.7 – 9.6) | |
| 54 (29 – 94) | |
| 5 (4 – 8) | |
BMI: Body mass index, 4D-CT: Four-Dimensional Computed Tomography
Parameters of helical computed tomography (CT) and Four-Dimensional Computed Tomography (4D-CT) acquisitions
| Parameters | Helical CT | 4D-CT |
|---|---|---|
| 0.7 | 0.7 | |
| 0.938 | / | |
| / | 16 | |
| 20 | 20 | |
| 1.25 | 1.25 | |
| Large Body | Large Body | |
| 120 | 120 | |
| 10 | 10 | |
| 440 | 440 | |
| 9.83 | 9.83 | |
| Yes / Yes | Yes / Yes |
CT: Computed tomography, 4D-CT: Four-Dimensional Computed Tomography, SFOV: Scan field of view, kV: kilo voltage, min mA: minimum value of milliampere modulation, max mA: maximum value of milliampere modulation
Figure 1Scan area (in blue box) for helical computed tomography (CT) (a), Four-Dimensional Computed Tomography (4D-CT) acquisition for tumor located in the upper lobe (UL) (b), and 4D-CT acquisition for tumor located in the lower lobe (LL) (c)
Figure 2National Cancer Institute dosimetry system for CT (NCICT) software Graphical User Interface (GUI) calculating dose delivered by Four-Dimensional Computed Tomography (4D-CT) to organs at risk (OARs) in the scan area illustrated with the red rectangle for upper lobe (UL) (a) and lower lobe (LL) regions (b)
Mean values and standard deviation of the dose delivered to the organs at risk (OARs), effective dose and imaging total dose were evaluated from different modalities depending on scanned regions
| Technique | Zone | Spinal cord (mGy) | Esophagus (mGy) | Lungs (mGy) | Heart (mGy) | Deff (mSv) |
|---|---|---|---|---|---|---|
| WT | 11.54±3.05 | 21.08±4.30 | 29.02±5.40 | 29.40±5.76 | 14.29±2.43 | |
| WT | 7.14±1.46 | 3.52±1.66 | 4.50±2.11 | 6.86±2.41 | 1.33±0.63 | |
| UL | 11.80±5.40 | 32.07±8.19 | 46.45±12.67 | 38.44±12.07 | 13.93±4.11 | |
| LL | 17.77±6.64 | 21.67±6.77 | 30.20±9.42 | 39.69±12.46 | 22.43±6.04 | |
| UL | 74.40±15.72 | 94.47±13.87 | 130.14±19.19 | 135.97±20.89 | 50.28±6.62 | |
| LL | 77.26±16.59 | 80.29±14.77 | 109.57±18.67 | 131.91±23.59 | 56.95±8.66 |
CT: Computed tomography, kV-CBCT: kilovoltage Cone-Beam computed tomography, 4D-CT: Four-Dimensional Computed Tomography, WT: Whole thorax, UL: Upper lobe, LL: Lower lobe, mGy: milli-gray, Deff: Effective dose
Figure 3Violin plots of dose delivered to the organs at risk (OARs) evaluated from Helical computed tomography (CT) from whole thorax (a); kilovoltage Cone-Beam CT (kV-CBCT) from whole thorax (b); cine Four-Dimensional Computed Tomography (4D-CT) in upper lobe (UL) region (c); and cine 4D-CT in lower lobe (LL) area (d). The dashed lines near the middle of the plots indicate the medians and the dot lines indicate the quartiles
The results of mean doses and maximum doses in volume (0.1 cc for the spinal cord and 0.5 cc for the heart and esophagus) evaluated from treatment planning system (TPS) for the spinal cord, esophagus, lungs and heart, depending on the lobe (upper lobe (UL) and lower lobe (LL) regions)
| Lobe | Dose | Spinal cord | Esophagus | Lungs | Heart | |
|---|---|---|---|---|---|---|
| V20<10 % | V16<20 % | |||||
| Dmean (Gy) | 1.24±0.93 | 2.84±2.42 | 3.49±1.84 | 6.05±3.96 | 0.93±1.86 | |
| Dmax (Gy) | 10.19±6.31 | 14.93±11.97 | 4.62±3.13 | 8.74±16.18 | ||
| Dmean (Gy) | 1.54±1.27 | 2.52±1.33 | 3.88±2.37 | 6.84±4.42 | 3.27±3.35 | |
| Dmax (Gy) | 11.02±6.20 | 10.94±3.64 | 5.11±3.46 | 16.54±14.51 | ||
UL: Upper lobe, LL: Lower lobe, Dmean: mean dose, Dmax: maximum dose
Comparison between treatment beam mean doses and imaging total mean dose delivered to the organs at risk (OARs) depending on upper lobe (UL) and lower lobe (LL) areas with associated relative gap rations and p-values were presented
| OARs | Upper lobe | Lower lobe | ||||||
|---|---|---|---|---|---|---|---|---|
| TPS | Imaging | RGR | p | TPS | Imaging | RGR | p | |
| (mGy) | (mGy) | (mGy) | (mGy) | |||||
| 1240±930 | 74.4±15.7 | 16.7 | <0.0001 | 11020±6200 | 77.3±16.6 | 121 | <0.0001 | |
| 2840±2420 | 94.5±13.9 | 30.1 | <0.0001 | 10940±3640 | 80.3±14.8 | 137.6 | <0.0001 | |
| 3490±760 | 130.1±19.2 | 26.8 | <0.0001 | 6840±4420 | 109.6±18.7 | 52.9 | <0.0001 | |
| 930±1840 | 136.0±20.9 | 6.8 | <0.0001 | 16540±14510 | 131.9±23.6 | 106.9 | <0.0001 | |
OARs: Organs at risk, Sc: Spinal cord, TPS: Treatment planning system, RGR: Relative GAP