Technical Note: Simulation of 4DCT tumor motion measurement errors.

Tai H Dou1, David H Thomas1, Dylan O'Connell1, Jeffrey D Bradley2, James M Lamb1, Daniel A Low1.   

Abstract

PURPOSE: To determine if and by how much the commercial 4DCT protocols under- and overestimate tumor breathing motion.
METHODS: 1D simulations were conducted that modeled a 16-slice CT scanner and tumors moving proportionally to breathing amplitude. External breathing surrogate traces of at least 5-min duration for 50 patients were used. Breathing trace amplitudes were converted to motion by relating the nominal tumor motion to the 90th percentile breathing amplitude, reflecting motion defined by the more recent 5DCT approach. Based on clinical low-pitch helical CT acquisition, the CT detector moved according to its velocity while the tumor moved according to the breathing trace. When the CT scanner overlapped the tumor, the overlapping slices were identified as having imaged the tumor. This process was repeated starting at successive 0.1 s time bin in the breathing trace until there was insufficient breathing trace to complete the simulation. The tumor size was subtracted from the distance between the most superior and inferior tumor positions to determine the measured tumor motion for that specific simulation. The effect of the scanning parameter variation was evaluated using two commercial 4DCT protocols with different pitch values. Because clinical 4DCT scan sessions would yield a single tumor motion displacement measurement for each patient, errors in the tumor motion measurement were considered systematic. The mean of largest 5% and smallest 5% of the measured motions was selected to identify over- and underdetermined motion amplitudes, respectively. The process was repeated for tumor motions of 1-4 cm in 1 cm increments and for tumor sizes of 1-4 cm in 1 cm increments.
RESULTS: In the examined patient cohort, simulation using pitch of 0.06 showed that 30% of the patients exhibited a 5% chance of mean breathing amplitude overestimations of 47%, while 30% showed a 5% chance of mean breathing amplitude underestimations of 36%; with a separate simulation using pitch of 0.1 showing, respectively, 37% overestimation and 61% underestimation.
CONCLUSIONS: The simulation indicates that commercial low-pitch helical 4DCT processes potentially yield large tumor motion measurement errors, both over- and underestimating the tumor motion.

Entities:  

Mesh:

Year:  2015        PMID: 26429283      PMCID: PMC4592437          DOI: 10.1118/1.4931416

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  Acquiring a four-dimensional computed tomography dataset using an external respiratory signal.

Authors:  S S Vedam; P J Keall; V R Kini; H Mostafavi; H P Shukla; R Mohan
Journal:  Phys Med Biol       Date:  2003-01-07       Impact factor: 3.609

2.  Respiration-correlated spiral CT: a method of measuring respiratory-induced anatomic motion for radiation treatment planning.

Authors:  E C Ford; G S Mageras; E Yorke; C C Ling
Journal:  Med Phys       Date:  2003-01       Impact factor: 4.071

3.  Acquiring 4D thoracic CT scans using a multislice helical method.

Authors:  P J Keall; G Starkschall; H Shukla; K M Forster; V Ortiz; C W Stevens; S S Vedam; R George; T Guerrero; R Mohan
Journal:  Phys Med Biol       Date:  2004-05-21       Impact factor: 3.609

4.  4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT.

Authors:  Tinsu Pan; Ting-Yim Lee; Eike Rietzel; George T Y Chen
Journal:  Med Phys       Date:  2004-02       Impact factor: 4.071

5.  Reduction of irregular breathing artifacts in respiration-correlated CT images using a respiratory motion model.

Authors:  Agung Hertanto; Qinghui Zhang; Yu-Chi Hu; Oleksandr Dzyubak; Andreas Rimner; Gig S Mageras
Journal:  Med Phys       Date:  2012-06       Impact factor: 4.071

6.  Effects of breathing variation on gating window internal target volume in respiratory gated radiation therapy.

Authors:  Jing Cai; Robert McLawhorn; Paul W Read; James M Larner; Fang-fang Yin; Stanley H Benedict; Ke Sheng
Journal:  Med Phys       Date:  2010-08       Impact factor: 4.071

7.  Quantifying ITV instabilities arising from 4DCT: a simulation study using patient data.

Authors:  Sara St James; Pankaj Mishra; Fred Hacker; Ross I Berbeco; John H Lewis
Journal:  Phys Med Biol       Date:  2012-02-17       Impact factor: 3.609

8.  A novel fast helical 4D-CT acquisition technique to generate low-noise sorting artifact-free images at user-selected breathing phases.

Authors:  David Thomas; James Lamb; Benjamin White; Shyam Jani; Sergio Gaudio; Percy Lee; Dan Ruan; Michael McNitt-Gray; Daniel Low
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-03-07       Impact factor: 7.038

9.  A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing.

Authors:  Daniel A Low; Michelle Nystrom; Eugene Kalinin; Parag Parikh; James F Dempsey; Jeffrey D Bradley; Sasa Mutic; Sasha H Wahab; Tareque Islam; Gary Christensen; David G Politte; Bruce R Whiting
Journal:  Med Phys       Date:  2003-06       Impact factor: 4.071

10.  Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy.

Authors:  Yvette Seppenwoolde; Hiroki Shirato; Kei Kitamura; Shinichi Shimizu; Marcel van Herk; Joos V Lebesque; Kazuo Miyasaka
Journal:  Int J Radiat Oncol Biol Phys       Date:  2002-07-15       Impact factor: 7.038

View more
  5 in total

1.  Initial clinical observations of intra- and interfractional motion variation in MR-guided lung SBRT.

Authors:  David H Thomas; Anand Santhanam; Amar U Kishan; Minsong Cao; James Lamb; Yugang Min; Dylan O'Connell; Yingli Yang; Nzhde Agazaryan; Percy Lee; Daniel Low
Journal:  Br J Radiol       Date:  2018-01-22       Impact factor: 3.039

2.  Preoperative Imaging Overestimates the Tumor Size in Pancreatic Neuroendocrine Neoplasms Associated with Multiple Endocrine Neoplasia Type 1.

Authors:  V Polenta; E P Slater; P H Kann; M B Albers; J Manoharan; A Ramaswamy; A H Mahnken; D K Bartsch
Journal:  World J Surg       Date:  2018-05       Impact factor: 3.352

3.  Quality assurance of four-dimensional computed tomography in a multicentre trial of stereotactic body radiotherapy of centrally located lung tumours.

Authors:  Marie Lambrecht; Jan-Jakob Sonke; Ursula Nestle; Heike Peulen; Damien C Weber; Marcel Verheij; Coen W Hurkmans
Journal:  Phys Imaging Radiat Oncol       Date:  2018-12-07

4.  4DCT and VMAT for lung patients with irregular breathing.

Authors:  Rhydian Caines; Naomi K Sisson; Carl G Rowbottom
Journal:  J Appl Clin Med Phys       Date:  2021-11-24       Impact factor: 2.102

5.  Dosimetric Study of Deep Learning-Guided ITV Prediction in Cone-beam CT for Lung Stereotactic Body Radiotherapy.

Authors:  Shujun Zhang; Bo Lv; Xiangpeng Zheng; Ya Li; Weiqiang Ge; Libo Zhang; Fan Mo; Jianjian Qiu
Journal:  Front Public Health       Date:  2022-03-22
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.