Rick Keesman1, Erik van der Bijl1, Tomas M Janssen1, Tineke Vijlbrief1, Floris J Pos1, Uulke A van der Heide1.
Abstract
BACKGROUND &Entities:
Keywords: Distortion; Implant; MR-linac; MRIgRT; MRgRT; Prosthesis
Year: 2020 PMID: 33458331 PMCID: PMC7807622 DOI: 10.1016/j.phro.2020.07.010
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Proposed workflow to decide on eligibility for MRIgRT for patients with metallic implants, based on planning CT and MR (including B0 map).
Acquisition parameters for the sequences used for anatomical imaging and acquisition of B0 maps during MR simulation and MRIgRT.
| acq. voxel size (mm3) | 0.7x0.7x3.0 | 2.0x2.0x6.0 | 1.2x1.2x2.4 | 4.0x4.0x4.0 |
| field-of-view (mm3) | 200x281x75 | 262x253x69 | 400x448x250 | 430x430x60 |
| TE (ms) | 120 | 2.2 | 128 | 4.6 |
| TR (ms) | 3960 | 6.6 | 1300 | 11 |
| flip angle (˚) | 90 | 21 | 90 | 15 |
| readout bandwidth (Hz/px) | 291 | 869 | 373 | 556 |
| scan time (min:sec) | 3:42 | 1:55 | 2:53 | 0:34 |
Fig. 2The B0 maps and T2 TSE images, acquired during MR simulation (3 T system) and the first treatment fraction of MRIgRT (1.5 T system), are shown together with the planning CT to which they were registered. For direct comparison, the field distortions are shown relative to field strength (range ± 1.7 ppm). Black and white arrows point to phase wrapping inside the CTV due to rectal air and metallic implants, respectively. Bright streak artefacts throughout the whole imaging plane were a result of the implant in CT. In the T2 TSE images acquired with MRI, signal voids, inside and nearby the implant as a consequence of magnetic field disturbances, were apparent.
Fig. 3B0 maps acquired during MR simulation and each treatment fraction of MRIgRT are shown for each patient. During the first treatment fraction, patient #1 and #2 had significant rectal filling resulting in phase wrapping near and inside the CTV (black arrows). Phase wrapping, sourced by the implant (white arrows), either occurs always or never inside the CTV, with the exception for patient #1 during simulation where the phase wrapping stops just outside of the CTV.
Metrics on the field-distortions ( in ppm) and derived expected geometric distortions (mm) for the 3D T2 TSE image used for plan adaptation during MRIgRT inside the CTVs are reported as (minimum|mean ± standard deviation|maximum). The maximum distortion inside the CTV () is calculated by including an estimate of the number of phase wraps sourced by the implant.
| #1 | simulation | (−1.36|−0.16 ± 0.34|0.72) | (−0.28|−0.03 ± 0.07|0.15) | 0 | 0.43 |
| fraction #1 | (−1.71|−0.44 ± 0.95|1.71) | (−0.35|−0.09 ± 0.20|0.35) | 1 | 1.40 | |
| #2 | (−1.70|−0.51 ± 0.74|1.71) | (−0.35|−0.10 ± 0.15|0.35) | 1 | 1.40 | |
| #3 | (−1.71|−0.55 ± 0.69|1.71) | (−0.35|−0.11 ± 0.14|0.35) | 1 | 1.40 | |
| #4 | (−1.71|−0.45 ± 0.83|1.71) | (−0.35|−0.09 ± 0.17|0.35) | 1 | 1.40 | |
| #5 | (−1.71|−0.53 ± 0.55|1.70) | (−0.35|−0.11 ± 0.11|0.35) | 1 | 1.40 | |
| #2 | simulation | (−0.29| 0.13 ± 0.19|0.64) | (−0.06| 0.03 ± 0.04|0.13) | 0 | 0.19 |
| fraction #1 | (−1.71|−0.28 ± 0.44|1.70) | (−0.35|−0.06 ± 0.09|0.35) | 0 | 0.70 | |
| #2 | (−0.68|−0.10 ± 0.14|0.23) | (−0.14|−0.02 ± 0.03|0.05) | 0 | 0.19 | |
| #3 | (−0.53| 0.05 ± 0.16|0.43) | (−0.11| 0.01 ± 0.03|0.09) | 0 | 0.20 | |
| #4 | (−0.81| 0.01 ± 0.25|0.82) | (−0.17| 0.00 ± 0.05|0.17) | 0 | 0.33 | |
| #5 | (−0.75|−0.17 ± 0.14|0.13) | (−0.15|−0.04 ± 0.03|0.03) | 0 | 0.18 | |
| #3 | simulation | (−1.70|−0.19 ± 0.81|1.70) | (−0.35|−0.04 ± 0.17|0.35) | 3 | 2.79 |
| fraction#1 | (−1.70|−0.06 ± 0.85|1.70) | (−0.35|−0.01 ± 0.17|0.35) | 2 | 2.10 | |
| #2 | (−1.71|−0.06 ± 0.84|1.70) | (−0.35|−0.01 ± 0.17|0.35) | 2 | 2.09 | |
| #3 | (−1.71|−0.13 ± 0.85|1.70) | (−0.35|−0.03 ± 0.17|0.35) | 1 | 1.40 | |
| #4 | (−1.66|−0.02 ± 0.84|1.70) | (−0.34| 0.00 ± 0.17|0.35) | 2 | 2.09 | |
| #5 | (−1.68|−0.01 ± 0.86|1.68) | (−0.34| 0.00 ± 0.18|0.35) | 1 | 1.39 |