Literature DB >> 33413529

Accuracy of intraoperative estimation of femoral stem anteversion in cementless total hip arthroplasty by using a digital protractor and a spirit level.

Anuwat Pongkunakorn1, Nawakun Phetpangnga2, Narawit Kananai2.   

Abstract

BACKGROUND: The femoral component anteversion during surgery is traditionally assessed by a visual assessment of the surgeon and has proven to be imprecise. We sought to determine the accuracy of a digital protractor and a spirit level to measure the stem anteversion during cementless THA.
METHODS: A prospective study was conducted among 107 patients (114 hips) who underwent primary cementless THA via posterolateral approach. A pipe with a spirit level was attached to the tibial tubercle and intermalleolar midpoint. While the leg was held perpendicularly to the floor, stem anteversion was estimated by 3 methods: method A by visual assessment; method B by a digital protractor alone; and method C by a digital protractor combined with a spirit level. The angles were compared with the true anteversion measured by postoperative CT scan.
RESULTS: The average anteversion by method C (22.8° ± 6.9°, range -2° to 40°) was significantly lower than method A (24.6° ± 5.2°, range 0° to 30°) (p=0.033), but not different from the true anteversion (22.1° ± 8.2°, range -5.4° to 43.1°) (p=0.445). There were no significant differences between method B (23.2° ± 8.2°, range -4° to 45°) and method A, C or the true anteversion. The mean deviation of the intraoperative estimation from the true anteversion was 0.8° ± 3.7° (range -7.1° to 8.0°) by method C; 1.2° ± 5.1° (range -8.8° to 14.3°) by method B; and 2.5° ± 7.4° (range -19.0° to 16.0°) by method A. Estimation error within 5° was found in 107 hips (93.9%) with method C; 86 hips (75.4%) with method B; and 59 hips (51.8%) with method A.
CONCLUSION: Accurate estimation of stem anteversion during cementless THA can be determined intraoperatively by the use of a digital protractor and a spirit level. TRIAL REGISTRATION: Thai Clinical Trials Registry ( TCTR 20180326003 ). Registered on 20 March 2018. Retrospectively registered.

Entities:  

Keywords:  Femoral stem anteversion; cementless total hip arthroplasty; digital protractor; spirit level

Mesh:

Substances:

Year:  2021        PMID: 33413529      PMCID: PMC7792124          DOI: 10.1186/s13018-020-02183-7

Source DB:  PubMed          Journal:  J Orthop Surg Res        ISSN: 1749-799X            Impact factor:   2.359


  15 in total

1.  Variability of the location of the tibial tubercle affects the rotational alignment of the tibial component in kinematically aligned total knee arthroplasty.

Authors:  Stephen M Howell; Justin Chen; Maury L Hull
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-04-05       Impact factor: 4.342

2.  Evaluation of methods that locate the center of the ankle for computer-assisted total knee arthroplasty.

Authors:  Robert A Siston; Aaron C Daub; Nicholas J Giori; Stuart B Goodman; Scott L Delp
Journal:  Clin Orthop Relat Res       Date:  2005-10       Impact factor: 4.176

3.  Combined anteversion technique for total hip arthroplasty.

Authors:  Lawrence D Dorr; Aamer Malik; Manish Dastane; Zhinian Wan
Journal:  Clin Orthop Relat Res       Date:  2008-11-01       Impact factor: 4.176

4.  Can the use of an inclinometer improve the positioning of the acetabular component in total hip arthroplasty?

Authors:  B Darrith; J A Bell; C Culvern; C J Della Valle
Journal:  Bone Joint J       Date:  2018-07       Impact factor: 5.082

5.  Computed tomography measurement of the accuracy of component version in total hip arthroplasty.

Authors:  Andrew P Wines; David McNicol
Journal:  J Arthroplasty       Date:  2006-08       Impact factor: 4.757

6.  The difference between the radiographic and the operative angle of inclination of the acetabular component in total hip arthroplasty: use of a digital protractor and the circumference of the hip to improve orientation.

Authors:  G Meermans; I Goetheer-Smits; R F Lim; W J Van Doorn; J Kats
Journal:  Bone Joint J       Date:  2015-05       Impact factor: 5.082

7.  A comparison of surgeon estimation and computed tomographic measurement of femoral component anteversion in cementless total hip arthroplasty.

Authors:  Lawrence D Dorr; Zhinian Wan; Aamer Malik; Jinjun Zhu; Manish Dastane; Prashant Deshmane
Journal:  J Bone Joint Surg Am       Date:  2009-11       Impact factor: 5.284

8.  A comparison of five approaches to measurement of anatomic knee alignment from radiographs.

Authors:  G McDaniel; K L Mitchell; C Charles; V B Kraus
Journal:  Osteoarthritis Cartilage       Date:  2009-10-24       Impact factor: 6.576

9.  Validity of the intra-operative measurement of stem anteversion and factors for the erroneous estimation in cementless total hip arthroplasty using postero-lateral approach.

Authors:  Y K Lee; J W Kim; T Y Kim; Y C Ha; K H Koo
Journal:  Orthop Traumatol Surg Res       Date:  2018-02-16       Impact factor: 2.256

10.  Use of a Digital Protractor and a Spirit Level to Determine the Intraoperative Anteversion of Femoral Component during Cemented Hip Hemiarthroplasty: a Prospective Clinical Trial.

Authors:  Anuwat Pongkunakorn; Patanapong Palawong; Swist Chatmaitri; Nawakun Phetpangnga
Journal:  Arch Bone Jt Surg       Date:  2019-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.