An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool.

Yangguang Ma1, Xuemin Wang2, Rizhen Mai3, Tao Wang1, Yuntong Pei1, Shuaipeng Liu1, Yuexin Guo1.   

Abstract

PURPOSE: To develop an efficient and economic daily quality research tool (DQRT) for daily check of multiplatform linear accelerators (LINACs) with flattening filter (FF) and flattening filter-free (FFF) photon beams by using an Electronic Portal Image Device (EPID).
MATERIALS AND METHODS: After EPID calibration, the monitored parameters were analyzed from a 10 cm × 10 cm open and 60° wedge portal images measured by the EPID with 100 MU exposure. Next, the repeatability of the EPID position accuracy, long-term stability, and linearity between image gray value and exposure were verified. Output and beam quality stability of the 6-MV FF and FFF beams measured by DQRT with the introduced setup errors of EPID were also surveyed. Besides, some test results obtained by DQRT were compared with those measured by FC65-G and Matrixx. At last, the tool was evaluated on three LINACs (Synergy, VersaHD, TrueBeam) for 2 months with two popular commercial QA tools as references.
RESULTS: There are no differences between repeatability tests for all monitored parameters. Image grayscale values obtained by EPID and exposure show good linearity. Either 6 MV FF or FFF photon beam shows minimal impact to the results. The differences between FC65-G, Matrixx and DQRT results are negligible. Monitor results of the two commercial tools are consistent with the DQRT results collected during the 2-month period.
CONCLUSION: With a shorter time and procedure, the DQRT is useful to daily QA works of LINACs, producing a QA result quality similarly to or more better than the traditional tools and giving richer contents to the QA results. For hospitals with limited QA time window available or lack of funding to purchase commercial QA tools, the proposed DQRT can provide an alternative and economic approach to accomplish the task of daily QA for LINACs.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  electronic portal image device; quality assurance; quick daily check

Mesh:

Year:  2021        PMID: 33410254      PMCID: PMC7856503          DOI: 10.1002/acm2.13055

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  25 in total

1.  Assessment of flatness and symmetry of megavoltage x-ray beam with an electronic portal imaging device (EPID).

Authors:  G Liu; T van Doorn; E Bezak
Journal:  Australas Phys Eng Sci Med       Date:  2002-07       Impact factor: 1.430

2.  Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager.

Authors:  G J Budgell; Q Zhang; R J Trouncer; R I Mackay
Journal:  Med Phys       Date:  2005-11       Impact factor: 4.071

3.  Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID.

Authors:  G J Budgell; R Zhang; R I Mackay
Journal:  Phys Med Biol       Date:  2007-02-27       Impact factor: 3.609

4.  Electron beam quality control using an amorphous silicon EPID.

Authors:  J A Beck; G J Budgell; D A Roberts; P M Evans
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

5.  Task Group 142 report: quality assurance of medical accelerators.

Authors:  Eric E Klein; Joseph Hanley; John Bayouth; Fang-Fang Yin; William Simon; Sean Dresser; Christopher Serago; Francisco Aguirre; Lijun Ma; Bijan Arjomandy; Chihray Liu; Carlos Sandin; Todd Holmes
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

6.  A cine-EPID based method for jaw detection and quality assurance for tracking jaw in IMRT/VMAT treatments.

Authors:  Todsaporn Fuangrod; Pejman Rowshanfarzad; Peter B Greer; Richard H Middleton
Journal:  Phys Med       Date:  2014-11-28       Impact factor: 2.685

7.  EPID-based in vivo dosimetry for stereotactic body radiotherapy of non-small cell lung tumors: Initial clinical experience.

Authors:  R Consorti; A Fidanzio; V Brainovich; F Mangiacotti; M De Spirito; M A Mirri; A Petrucci
Journal:  Phys Med       Date:  2017-09-28       Impact factor: 2.685

8.  Evaluation of the Machine Performance Check application for TrueBeam Linac.

Authors:  Alessandro Clivio; Eugenio Vanetti; Steven Rose; Giorgia Nicolini; Maria F Belosi; Luca Cozzi; Christof Baltes; Antonella Fogliata
Journal:  Radiat Oncol       Date:  2015-04-21       Impact factor: 3.481

9.  Beam focal spot position: The forgotten linac QA parameter. An EPID-based phantomless method for routine Stereotactic linac QA.

Authors:  Jacek M Chojnowski; Michael P Barnes; Jonathan R Sykes; David I Thwaites
Journal:  J Appl Clin Med Phys       Date:  2017-08-08       Impact factor: 2.102

10.  An investigation of gantry angle data accuracy for cine-mode EPID images acquired during arc IMRT.

Authors:  Peter M McCowan; Daniel W Rickey; Pejman Rowshanfarzad; Peter B Greer; William Ansbacher; Boyd M McCurdy
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.