| Literature DB >> 28786168 |
Jacek M Chojnowski1,2, Michael P Barnes3,4, Jonathan R Sykes2,5, David I Thwaites2.
Abstract
Modern day Stereotactic treatments require high geometric accuracy of the delivered treatment. To achieve the required accuracy the IGRT imaging isocenter needs to closely coincide with the treatment beam isocenter. An influence on this isocenter coincidence and on the spatial positioning of the beam itself is the alignment of the treatment beam focal spot with collimator rotation axis. The positioning of the focal spot is dependent on the linac beam steering and on the stability of the monitor chamber and beam steering servo system. As such, there is the potential for focal spot misalignment and this should be checked on a regular basis. Traditional methods for measuring focal spot position are either indirect, inaccurate, or time consuming and hence impractical for routine use. In this study a novel, phantomless method has been developed using the EPID (Electronic Portal Imaging Device) that utilizes the different heights of the MLC and jaws. The method has been performed on four linear accelerators and benchmarked against an alternate ion chamber-based method. The method has been found to be reproducible to within ±0.012 mm (1 SD) and in agreement with the ion chamber-based method to within 0.001 ± 0.015 mm (1 SD). The method could easily be incorporated into a departmental routine linac QA (Quality Assurance) program.Entities:
Keywords: zzm321990EPIDzzm321990; focal spot; quality assurance
Mesh:
Year: 2017 PMID: 28786168 PMCID: PMC5875833 DOI: 10.1002/acm2.12147
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Diagram of a Varian linac head (schematic and not to scale) and illustration of radiation focal spot position determination using the EPID. Vertical black line represents the collimator rotation axis. Red line represents center of jaw defined field with 180° collimator rotation. Blue line represents center of MLC defined field with 180° collimator rotation.
Figure 2A setup of (a) the dosimetry tool attached to a collimator on a Varian Clinac® 6EX (b). The illustration of obscuring of the focal spot by the jaw and the geometric relationship of the lengths of similar triangles (schematic and not to scale).
Reproducibility measurements of the 6MV focal spot offset using the EPID on four different Varian linacs
| Measurement No. | Linac 1 (6EX) | Linac 2 (21iX) | Linac 3 (6EX) | Linac 4 (TrueBeam) | ||||
|---|---|---|---|---|---|---|---|---|
| Crossplane | Inplane | Crossplane | Inplane | Crossplane | Inplane | Crossplane | Inplane | |
| (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | |
| 1 | −0.002 | 0.099 | 0.069 | 0.021 | −0.027 | −0.445 | 0.081 | −0.020 |
| 2 | −0.008 | 0.084 | 0.087 | 0.058 | −0.026 | −0.419 | 0.082 | −0.018 |
| 3 | −0.006 | 0.098 | 0.081 | 0.097 | −0.032 | −0.433 | 0.085 | −0.046 |
| Average | −0.005 | 0.094 | 0.079 | 0.059 | −0.029 | −0.433 | 0.083 | −0.028 |
| 1SD | 0.003 | 0.008 | 0.009 | 0.038 | 0.003 | 0.013 | 0.002 | 0.016 |
Validation of the phantomless method of focal spot offset measurement with the ionization chamber method
| Measurements | Linac 1 (6EX) | Linac 2 (21iX) | Linac 3 (6EX) | Linac 4 (TrueBeam) | ||||
|---|---|---|---|---|---|---|---|---|
| Crossplane | Inplane | Crossplane | Inplane | Crossplane | Inplane | Crossplane | Inplane | |
| (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | (mm) | |
| By ion chamber | −0.005 | 0.094 | 0.097 | 0.055 | −0.036 | −0.458 | 0.106 | −0.024 |
| By EPID | −0.005 | 0.094 | 0.079 | 0.059 | −0.029 | −0.433 | 0.083 | −0.028 |
| Difference | 0.000 | 0.000 | 0.018 | −0.004 | −0.007 | −0.025 | 0.023 | 0.004 |