EPID-based in vivo dosimetry for stereotactic body radiotherapy of non-small cell lung tumors: Initial clinical experience.

R Consorti1, A Fidanzio2, V Brainovich1, F Mangiacotti1, M De Spirito3, M A Mirri4, A Petrucci1.   

Abstract

PURPOSE: EPID-based in vivo dosimetry (IVD) has been implemented for stereotactic body radiotherapy treatments of non-small cell lung cancer to check both isocenter dose and the treatment reproducibility comparing EPID portal images.
METHODS: 15 patients with lung tumors of small dimensions and treated with volumetric modulated arc therapy were enrolled for this initial experience. IVD tests supplied ratios R between in vivo reconstructed and planned isocenter doses. Moreover a γ-like analysis between daily EPID portal images and a reference one, in terms of percentage of points with γ-value smaller than 1, Pγ<1, and mean γ-values, γmean, using a local 3%-3mm criteria, was adopted to check the treatment reproducibility. Tolerance levels of 5% for R ratio, Pγ<1 higher than 90% and γmean lower than 0.67 were adopted.
RESULTS: A total of 160 EPID images, two images for each therapy session, were acquired during the treatment of the 15 patients. The overall mean of the R ratios was equal to 1.005±0.014 (1 SD), with 96.9% of tests within±5%. The 2D image γ-like analysis showed an overall γmean of 0.39±0.12 with 96.1% of tests within the tolerance level, and an average Pγ<1 value equal to 96.4±3.6% with 95.4% of tests with Pγ<1>90%. Paradigmatic discrepancies were observed in three patients: a set-up error and a patient morphological change were identified thanks to CBCT image analysis whereas the third discrepancy was not fully justified.
CONCLUSIONS: This procedure can provide improved patient safety as well as a first step to integrate IVD and CBCT dose recalculation.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  EPID; In vivo dosimetry; Lung; SBRT

Mesh:

Year:  2017        PMID: 29173910     DOI: 10.1016/j.ejmp.2017.09.133

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool.

Authors:  Yangguang Ma; Xuemin Wang; Rizhen Mai; Tao Wang; Yuntong Pei; Shuaipeng Liu; Yuexin Guo
Journal:  J Appl Clin Med Phys       Date:  2021-01-07       Impact factor: 2.102

2.  Voxel-Level BED Corrected Dosimetric and Radiobiological Assessment of 2 Kinds of Hybrid Radiotherapy Planning Methods for Stage III NSCLC.

Authors:  Hao Wang; Ying Huang; Hua Chen; Yan Shao; Yanhua Duan; Aihui Feng; Hengle Gu; Xiurui Ma; Zhiyong Xu; Qing Kong; Yongkang Zhou
Journal:  Technol Cancer Res Treat       Date:  2022 Jan-Dec

Review 3.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

4.  Verification of an optimizer algorithm by the beam delivery evaluation of intensity-modulated arc therapy plans.

Authors:  Tamas Pocza; Domonkos Szegedi; Tibor Major; Csilla Pesznyak
Journal:  Radiol Oncol       Date:  2021-11-19       Impact factor: 2.991

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.