A cine-EPID based method for jaw detection and quality assurance for tracking jaw in IMRT/VMAT treatments.

Todsaporn Fuangrod1, Pejman Rowshanfarzad2, Peter B Greer3, Richard H Middleton4.   

Abstract

A new tool with the potential to verify and track jaw position during delivery has been developed. The method should be suitable for independent quality assurance for jaw position during jaw tracking dynamic IMRT and VMAT treatments. The jaw detection and tracking algorithm developed consists of five main steps. Firstly, the image is enhanced by removing a normalised predicted EPID image (that does not include the collimator transmission) from each cine EPID image. Then, using a histogram clustering technique a global intensity threshold level was determined. This threshold level was used to classify each pixel of the image as either under the jaws or under the MLC. Additionally, the collimator angle was automatically detected and used to rotate the image to vertical direction. Finally, this rotation allows the jaw positions to be determined using vertical and horizontal projection profiles. Nine IMRT fields (with static jaws) and a single VMAT clinical field (with dynamic jaws) were tested by determining the root mean square difference between planned and detected jaw positions. The test results give a detection accuracy of ±1 mm RMS error for static jaw IMRT treatments and ±1.5 mm RMS error for the dynamic jaw VMAT treatment. This method is designed for quality assurance and verification in modern radiation therapy; to detect the position of static jaws or verify the position of tracking jaws in more complex treatments. This method uses only information extracted from EPID images and it is therefore independent from the linear accelerator.
Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  EPID; Jaws; QA; Real-time verification

Mesh:

Year:  2014        PMID: 25442072     DOI: 10.1016/j.ejmp.2014.11.002

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool.

Authors:  Yangguang Ma; Xuemin Wang; Rizhen Mai; Tao Wang; Yuntong Pei; Shuaipeng Liu; Yuexin Guo
Journal:  J Appl Clin Med Phys       Date:  2021-01-07       Impact factor: 2.102

Review 2.  Novel methodologies for dosimetry audits: Adapting to advanced radiotherapy techniques.

Authors:  Marlies Pasler; Victor Hernandez; Núria Jornet; Catharine H Clark
Journal:  Phys Imaging Radiat Oncol       Date:  2018-03-19

3.  Validation of new transmission detector transmission factors for online dosimetry: an experimental study.

Authors:  So-Yeon Park; Jong Min Park; Jung-In Kim; Sungyoung Lee; Chang Heon Choi
Journal:  Radiat Oncol       Date:  2018-08-24       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.