The single-monitor trial: an embedded CADe system increased adenoma detection during colonoscopy: a prospective randomized study.

Peixi Liu1, Pu Wang2, Jeremy R Glissen Brown3, Tyler M Berzin3, Guanyu Zhou1, Weihui Liu1, Xun Xiao1, Ziyang Chen1, Zhihong Zhang1, Chao Zhou1, Lei Lei1, Fei Xiong1, Liangping Li1, Xiaogang Liu1.   

Abstract

BACKGROUND: Computer-aided detection (CADe) of colon polyps has been demonstrated to improve colon polyp and adenoma detection during colonoscopy by indicating the location of a given polyp on a parallel monitor. The aim of this study was to investigate whether embedding the CADe system into the primary colonoscopy monitor may serve to increase polyp and adenoma detection, without increasing physician fatigue level.
METHODS: Consecutive patients presenting for colonoscopies were prospectively randomized to undergo routine colonoscopy with or without the assistance of a real-time polyp detection CADe system. Fatigue level was evaluated from score 0 to 10 by the performing endoscopists after each colonoscopy procedure. The main outcome was adenoma detection rate (ADR).
RESULTS: Out of 790 patients analyzed, 397 were randomized to routine colonoscopy (control group), and 393 to a colonoscopy with computer-aided diagnosis (CADe group). The ADRs were 20.91% and 29.01%, respectively (OR = 1.546, 95% CI 1.116-2.141, p = 0.009). The average number of adenomas per colonoscopy (APC) was 0.29 and 0.48, respectively (Change Folds = 1.64, 95% CI 1.299-2.063, p < 0.001). The improvement in polyp detection was mainly due to increased detection of non-advanced diminutive adenomas, serrated adenoma and hyperplastic polyps. The fatigue score for each procedure was 3.28 versus 3.40 for routine and CADe group, p = 0.357.
CONCLUSIONS: A real-time CADe system employed on the primary endoscopy monitor may lead to improvements in ADR and polyp detection rate without increasing fatigue level during colonoscopy. The integration of a low-latency and high-performance CADe systems may serve as an effective quality assurance tool during colonoscopy. www.chictr.org.cn number, ChiCTR1800018058.
© The Author(s), 2020.

Entities:  

Keywords:  artificial intelligence; colonoscopy; computer-aided diagnosis; polyp

Year:  2020        PMID: 33403003      PMCID: PMC7745558          DOI: 10.1177/1756284820979165

Source DB:  PubMed          Journal:  Therap Adv Gastroenterol        ISSN: 1756-283X            Impact factor:   4.409


Background and objectives

Colonoscopy remains the gold standard examination for colorectal cancer (CRC) screening. However, colonoscopy can be technically demanding, and there is significant variation in how colonoscopy is performed. This variation in performance has been linked to important outcome measures. For example, interval cancers are more common in low adenoma detectors as compared with high adenoma detectors. Evidence has shown that each 1% increase in adenoma detection rate (ADR) predicts 3% decrease in interval CRC.[1,2] Computer-aided detection (CADe)[3] of colon polyps during colonoscopy has been demonstrated to improve colon polyp detection rates (PDRs) and ADRs by alarming polyps on a parallel monitor.[4] Introducing a second monitor might unnecessarily add physician fatigue by changing gaze patterns and requiring that an endoscopist watches both screens, an issue that has come up in the study of similar technologies.[5] Moreover, interestingly, some briefly visible polyps, especially those that appear on the edge of the screen or are partially obscured, might still be missed during the time interval when the endoscopist turns to look at the second monitor. A recent study showed that more than 43% of those polyps and adenomas initially missed by physicians were on the edge of the visual field.[6] Thus, vigilance remains crucially important so that brief alarm flashes on the edge of the visual field may otherwise be missed (supplemental material Video S1, Video S2). This degree of vigilance may be easier if the CADe is embedded into the primary colonoscopy monitor so that the endoscopist’s attentions are not divided. Thanks to recent improvements in graphics processing unit (GPU) computation power and improvements in processing time, a CADe system may be embedded into the primary endoscopy monitor without introducing noticeable latency. A recent study[7] showed an increase of adenoma detection using fully integrated CADe in the endoscopy workflow, avoiding a second display to show the artificial intelligence (AI) detection. However, the ADR of the control group was based on only high-risk CRC populations without representation of more extensive normal patients. Furthermore, false detections and missed polyps by the CADe and initially missed polyps by the endoscopist were not recorded nor analyzed in the recent one-monitor study. The concrete contributions and distractions caused by the real-time CADe system are crucially important to reveal the actual efficacy and risk of such systems. In this study, we further hypothesized that this may also lead to decrease physician fatigue during endoscopy. The aim of this study was to investigate whether the integration of a CADe system into the primary monitor used during colonoscopy may increase polyp and adenoma detection without increasing physician fatigue. False detections and missed polyps by the CADe system and initially missed polyps by the endoscopist were recorded, analyzed and compared with the previous double-blinded study,[6] in order to gain a concrete understanding of how a real-time CADe system makes efficacy during real-life colonoscopy.

Method

Study design

CADe system integration (EndoScreener, Shanghai Wision AI Co., Ltd. China)

The real-time automatic polyp detection system (Figure S1, supplemental material) was developed on a deep-learning architecture,[8] which was previously validated to have a per-image sensitivity of 94.38%, per-image specificity of 95.92% and an area under the receiver operating characteristic curve of 0.984 to detect colon polyps in colonoscopy report images. Moreover, the algorithm was demonstrated to have a per-polyp sensitivity of 100.00% (per-image sensitivity of 91.64%) and a per-image specificity of 95.40% in real-world colonoscopy videos.[3] By changing the deep-learning framework from Caffe (Berkeley, CA) to TensorRT (Nvidia, CA), and combining some layers in SegNet, the system processes >30 frames per second on Ge-Force-1080ti (Nvidia, CA). To enhance user experience regarding latency, the colonoscopy video was unprocessed and displayed directly on the primary monitor without waiting for its detection result (the detection box), which was then placed on subsequent frames with a latency of 20.33 ± 0.14 ms. CPU/GPU was used for the overlap and display of colonoscopy video frames, which takes 5.99 ± 0.24 ms to overlap the detection box in the original colonoscopy video and display on the monitor. If FPGA or ASIC were used for the overlapping and displaying instead of CPU/GPU, colonoscopy video with detection can be displayed with a latency less than 1.00 ms (Figure 1).
Figure 1.

Latency diagram.

Latency diagram.

Prospective randomized study

The prospective study was designed as a non-blinded randomized trial. This study was conducted in the Endoscopy Center of Sichuan Provincial People’s Hospital, China. Consecutive patients who underwent colonoscopy from September 2018 to February 2019 were eligible for enrollment (Figure 2). Written informed consent was obtained from all individual participants enrolled in the study. Bowel preparation method was 2 L of polyethylene glycol with 6 ml simethicone solution, given in split doses. Colonoscopies were performed with last-generation high image-quality colonoscopes (Olympus CF-Q260 and CF-H260) and high-definition monitors. We excluded patients with a history of inflammatory bowel disease (IBD), CRC, any polyposis syndromes, colorectal surgery and patients with a contraindication for biopsy. Patients with prior failed colonoscopy and high suspicion of polyposis syndromes, IBD and typical advanced CRC were also excluded.
Figure 2.

Flow diagram of enrollment.

Flow diagram of enrollment. Baseline demographic characteristics for each patient were recorded (Table S1, supplemental material). Any complication during the procedure or recovery was also recorded. Eleven physicians from the division of gastroenterology participated in the study, including eight experienced endoscopists (four seniors, four mid-level) and three junior endoscopists. Each patient was prospectively randomized into one of two groups by the research assistant, using a digital random number generator before the colonoscopy procedure. In the control group, a routine colonoscopy was performed. In the research group (CADe group), the real-time automatic polyp detection system was used to assist the endoscopist. The system was embedded in the endoscopy processor, processed each frame of the video stream synchronously and reported the detected polyp location with a hollow blue tracing box directly in the endoscopy monitor with a simultaneous sound alarm (Video S3). The system was employed during withdrawal only. For any area detected by the CADe system, the endoscopist was required to check and verify the area using his or her own clinical judgment. The staff assistant records whether the CADe system detects polyps before endoscopists every time encountering a polyp; this is based on the subjective judgment of the operating endoscopist. In this case, the CADe system detects a polyp prior to the endoscopist, which indicates not only the advantage of low latency but also the possibility of avoiding a missed diagnosis. If the endoscopist deemed that a polyp was first identified by himself or simultaneously with the CADe system, then it was not credited to the CADe system. After each procedure, the endoscopist rated his or her fatigue level from 0 to 10 using a 10-point Likert scale where 0 represented no fatigue and 10 represented extreme fatigue. When a polyp was identified either by the endoscopist first or by CADe system first, the nurse assisted in performing cold forceps biopsy for histology and the staff assistant recorded the location, size and morphological features according to the Paris classification. In the CADe group, missed polyps by the system and consistent false detections were also recorded. A missed polyp was defined as a polyp confirmed by the endoscopist but undetected by the system. A consistent false detection was defined as a detected lesion, which was continuously traced by the system, deemed by the endoscopist not to be a polyp.[6] This study was approved by the local review board and registered with the Chinese Clinical Trial Registry. All authors had access to the study data and reviewed and approved the final manuscript.

Study endpoint

The primary endpoint was ADR, which was defined as proportion of individuals undergoing a complete colonoscopy who have one or more adenomas detected. The secondary endpoints were PDR, which was defined as the proportion of individuals undergoing a complete colonoscopy who have one or more polyps detected. Polyps per colonoscopy (PPC) and adenomas per colonoscopy (APC) were calculated by dividing the total number of polyps or adenomas detected by the total number of colonoscopies performed.

Statistical analysis

We prospectively designed this study to allow for 80% power or more to detect a 10% difference (30% versus 20%), in ADR, between colonoscopy procedures using a two-group Chi-square test with a two-sided level of 0.05. A sample size of 702 participants was needed, and the overall participant enrollment goal was 850 to allow for potential exclusions or dropouts. Statistical analysis was performed with R studio V.3.4.0 or higher. Comparison of baseline clinical and demographic characteristics between the CADe and the control group was performed using a Chi-square test for categorical variables and using the two-sample t-test for continuous variables. Regarding the ADR and PDR, a logistic regression was performed to evaluate the effect of computer-assisted diagnosis for colonoscopy on the adenoma/polyp detection rate. The response variable was the binary outcome of whether an adenoma/polyp was detected or not. The covariate was the group variable indicating whether the patient belonged to the computer-assisted group. Regarding the number of detected adenomas and polyps, a Poisson regression was applied to evaluate the effect of computer-assisted diagnosis for colonoscopy. A two-sided p-value of 0.05 was used as the threshold for statistical significance. In the event of any baseline clinical and demographic characteristics showing a statistically significant difference between the two comparison groups, additional covariate adjusted logistics/Poisson regression models were built to address the possible confounding effect by adding those significant characteristics into the models as covariates.

Results

Patient enrollment and baseline data

A total of 846 consecutive patients were enrolled, among which 56 patients were excluded during colonoscopy due to meeting exclusion criteria. A total of 790 eligible patients were analyzed, with 397 patients randomized into the control group and 393 into the CADe group. Baseline characteristics are presented in Table 1. There were no statistically significant differences between the two groups in terms of demographic data and adenoma detection risk factors (Table S1). There were no complications reported. Withdrawal times when biopsy time was excluded from analysis were 6.62 min and 6.71 min in the control and CADe groups, respectively (p = 0.450).
Table 1.

Baseline information.

CharacteristicsRoutine colonoscopy (n = 397)CADe colonoscopy (n = 393)p-value[*]
Age, mean (SD)48.79 (13.00)49.84 (13.11)0.253
BMI, mean (SD)23.13 (3.02)23.08 (3.08)0.821
Withdrawal time, mean (SD)6.94 (1.53)7.29 (1.98)0.007
Withdrawal time excluding biopsy time6.62 (1.22)6.71 (1.63)0.450
Total time, mean (SD)13.01 (4.64)13.37 (5.15)0.315
Insertion time, mean (SD)6.07 (4.40)6.08 (4.70)0.975
No polyp withdrawal time, mean (SD)6.54 (1.13)6.41 (1.27)0.249
Indication0.231
Screening, n (%)84 (21.16)98 (24.94)
Symptomatic, n (%)313 (78.84)295 (75.06)
Sex0.407
Female, n (%)203 (51.13)213 (54.20)
Male, n (%)194 (48.87)180 (45.80)
BMI category0.920
<25, n (%)296 (74.56)290 (73.79)
25 ⩽ BMI <30, n (%)93 (23.43)96 (24.43)
⩾30, n (%)8 (2.02)7 (1.78)
Procedure time0.334
AM, n (%)197 (49.62)209 (53.18)
PM, n (%)200 (50.38)184 (46.82)
Endoscope0.127
H-260183 (46.10)202 (51.40)
Q-260214 (53.90)191 (48.60)
Anesthesia[#]0.835
No, n (%)18 (4.53)19 (4.83)
Yes, n (%)379 (95.47)374 (95.17)
Boston Score, mean (SD)6.71 (1.38)6.72 (1.52)0.911
Boston Score Rank0.067
Inadequate (Sum < 6.0 or anyone < 2.0), n (%)63 (15.87)82 (20.87)
Adequate (Sum ⩾ 6.0 and everyone ⩾ 2.0), n (%)334 (84.13)311 (79.13)
Doctor level0.787
Senior127 (31.99)120 (30.53)
Mid-level234 (58.94)241 (61.32)
Junior36 (9.07)32 (8.14)

No Polyp Withdrawal time: withdrawal time during those colonoscopies where no polyp was detected or removed.

p-value from χ2 test (or Fisher’s exact test, as appropriate) or t-test.

Anesthesia was administered with midazolam, fentanyl by an anesthesiologist there to monitor for complications.

BMI, body mass index.

Baseline information. No Polyp Withdrawal time: withdrawal time during those colonoscopies where no polyp was detected or removed. p-value from χ2 test (or Fisher’s exact test, as appropriate) or t-test. Anesthesia was administered with midazolam, fentanyl by an anesthesiologist there to monitor for complications. BMI, body mass index.

Adenoma characteristics, adenomas detected per colonoscopy and ADR

There were 304 (48.64%) adenomas and 12 (1.92%) serrated adenomas detected. The ADR were 20.91% and 29.01% in the control and CADe arms, respectively [odds ratio (OR) = 1.546, 95% confidence interval (CI) 1.116–2.141, p = 0.009]. The APCs were 0.29 and 0.48, respectively (Change Folds = 1.64, 95% CI 1.299–2.063, p < 0.001). The average number of sessile serrated lesions (SSLs) was 0.0049 versus 0.0261 in the control versus CADe groups, p = 0.021 (Tables 2 and 3).
Table 2.

Polyp and adenoma characteristics.

CharacteristicsRoutine colonoscopy (n = 204)CADe colonoscopy (n = 421)p-value[*]
Pathology0.006
Carcinoma, n (%)0 (0.00)0 (0.00)1.000
SSL, n (%)1 (0.49)11 (2.61)0.021
Adenoma, n (%)
Advanced adenoma, n (%)8 (3.92)6 (1.43)0.607
Others, n (%)108 (52.94)182 (43.23)<0.001
Benign lesions, n (%)
Hyperplastic and inflammatory87 (42.65)222 (52.73)<0.001
Hamartoma0 (0.00)0 (0.00)1.000
Normal colon mucosa0 (0.00)0 (0.00)1.000
Polyp location0.505
Cecum, n (%)5 (2.45)8 (1.90)0.400
Ascending, n (%)37 (18.14)87 (20.67)<0.001
Transverse, n (%)48 (23.53)76 (18.05)0.011
Descending, n (%)25 (12.25)45 (10.69)0.017
Sigmoid, n (%)38 (18.63)98 (23.28)<0.001
Rectum, n (%)51 (25.00)107 (25.42)<0.001
Polyp shape0.016
Pedunculated, n (%)18 (8.82)22 (5.23)0.507
Sessile, n (%)182 (89.22)398 (94.54)<0.001
LST, n (%)4 (1.96)1 (0.24)0.218
Polyp size (mm), mean (SD)4.97 (3.03)3.85 (2.05)<0.001
Polyp size category0.001
0–5 mm, n (%)149 (73.04)359 (85.27)<0.001
6–10 mm, n (%)47 (23.04)56 (13.30)0.349
>10 mm, n (%)8 (3.92)6 (1.43)0.607
Adenoma location0.253
Cecum, n (%)4 (3.45)4 (2.13)0.989
Ascending, n (%)25 (21.55)52 (27.66)0.002
Transverse, n (%)31 (26.72)33 (17.55)0.771
Descending, n (%)16 (13.79)25 (13.30)0.154
Sigmoid, n (%)22 (18.97)49 (26.06)0.002
Rectum, n (%)18 (15.52)25 (13.30)0.273
Adenoma shape0.093
Pedunculated, n (%)17 (14.66)19 (10.11)0.716
Sessile, n (%)97 (83.62)169 (89.89)<0.001
LST, n (%)2 (1.72)0 (0.00)0.996
Adenoma size (mm), mean (SD)5.83 (3.15)4.55 (2.26)<0.001
Adenoma size category0.004
0–5 mm, n (%)69 (59.48)146 (77.66)<0.001
6–10 mm, n (%)41 (35.34)37 (19.68)0.683
>10 mm, n (%)6 (5.17)5 (2.66)0.776

p-value from χ2 test (or Fisher’s exact test, as appropriate).

LST, laterally spreading tumor; SSL, sessile serrated lesion.

Table 3.

Main outcomes.

Routine colonoscopy (n = 397)CAD colonoscopy (n = 393)p-value[**]FC/OR95% CI
PDR0.33250.4707<0.0011.786[*]1.339–2.381
ADR0.20910.29010.0091.546[*]1.116–2.141
SDR0.00250.00760.3363.0460.316–29.404
PPC0.51391.0712<0.0012.085[#]1.764–2.464
APC0.29220.4784<0.0011.637[#]1.299–2.063
Fatigue level3.28213.40200.3571.037[*]0.960–1.119

OR, odds ratio.

FC, fold change.

p-value from χ2 test (or Fisher’s exact test, as appropriate) or t-test.

ADR, adenoma detection rate; APC, adenoma per colonoscopy; PDR, polyp detection rate; PPC, polyp per colonoscopy; SDR, sessile serrated lesion detection rate.

Polyp and adenoma characteristics. p-value from χ2 test (or Fisher’s exact test, as appropriate). LST, laterally spreading tumor; SSL, sessile serrated lesion. Main outcomes. OR, odds ratio. FC, fold change. p-value from χ2 test (or Fisher’s exact test, as appropriate) or t-test. ADR, adenoma detection rate; APC, adenoma per colonoscopy; PDR, polyp detection rate; PPC, polyp per colonoscopy; SDR, sessile serrated lesion detection rate. The increase in detection was mainly due to non-advanced diminutive adenomass, serrated adenoma and hyperplastic polyps. Among which, the detection of SSL was significantly higher in CADe group (2.61%, 11/421) than that of routine group (0.49%, 1/204) (P=0.021). Moreover, in CADe group, 6 out 11 SSLs were detected in right colon, the remaining were detected in sigmoid colon and rectum, whereas there was only one SSL detected in right colon in routine group.

Polyp characteristics, polyps detected per colonoscopy and PDR

A total of 625 polyps were detected. The PDRs of the control and CADe groups were 33.25% and 47.07%, respectively (OR = 1.786, 95% CI 1.339–2.381, p < 0.001). The PPCs were 0.51 and 1.07, respectively (Change Folds = 2.09, 95% CI 1.764–2.464, p < 0.001) (Tables 2 and 3). There was no statistically significant difference between the two groups in terms of baseline clinical and demographic variables. Thus, covariate adjusted models were not considered to address the potential confounding effect.

Consistent false detections by the CADe system

There was a total of 29 consistent false detections (based on endoscopist judgment of non-polyp objects which were continuously traced) in the CADe group, averaging at 0.074 per colonoscopy, mostly wrinkled mucosa. One extra consistent false detection was found after histology exam as normal colon mucosa (Table 4).
Table 4.

Statistical results of false-alarms and missed polyp.

CAD colonoscopy[*]
False-alarm 29 (100.00)
Bubble3 (10.34)
Feces1 (3.45)
Undigested debris3 (10.34)
Wrinkled mucosa12 (41.38)
Local inflammation6 (20.69)
Local bleeding0 (0.00)
Rounded drug capsules3 (10.34)
Other (circular blood vessel, scar, diverticulum, etc)1 (3.45)
Missed Polyp 0 (0.00)

n (%).

Statistical results of false-alarms and missed polyp. n (%). Of all the detected polyps in the CADe group, none was missed by the CADe system.

Polyps detected first by CADe system before endoscopists in the CADe group

In total, 165 polyps, including 73 adenomas and one SSLs, averaging 0.19 adenomas and 0.42 polyps per patient, were detected by CADe system prior to endoscopists in the CADe group (Table 5). These polyps were generally small in size [mean polyp size 3.55 mm (SD 1.58); mean adenoma size 3.82 mm (SD 1.23)], isochromatic [149 (90.3%) of 165 polyps; 61 (83.6%) of 73 adenomas], flat in shape [122 (73.9%) polyps; 48 (65.8%) adenomas; Video S3], unclear boundary [33 (20%) polyps; 11 (15.1%) adenomas], partly behind colon folds [32 (19.4%) polyps; 16 (21.9%) adenomas], and on the edge of the visual field [78 (47.3%) polyps; 36 (49.3%) adenomas; Video S1].
Table 5.

Characteristics of polyps missed at endoscopy among patients in computer-aided detection group.

Endoscopist-missed polyp in CADe colonoscopyPolyp (n = 165)Adenoma (n = 73)SSL (n = 1)
Polyp characteristics
 Isochromatic149 (90.3)61 (83.6)0 (0.0)
 Flat122 (73.9)48 (65.8)0 (0.0)
 Unclear boundary33 (20.0)11 (15.1)0 (0.0)
 Partially occluded by colon folds32 (19.4)16 (21.9)1 (100.0)
 On the edge of visual field78 (47.3)36 (49.3)1 (100.0)
Colon condition
 Insufficient air inflation27 (16.4)8 (11.0)1 (100.0)
 Partially occluded by Liquid feces or debris35 (21.2)18 (24.7)1 (100.0)
Endoscopist
 Withdraw too fast20 (12.1)7 (9.6)0 (0.0)
 Obvious polyp missed diagnosis11 (6.7)6 (8.2)0 (0.0)
Endoscopy
 Overexposure112 (67.9)52 (71.2)1 (100.0)
 Blurred lens20 (12.1)7 (9.6)0 (0.0)
 Insufficient light condition21 (12.7)12 (16.4)0 (0.0)
Characteristics of polyps missed at endoscopy among patients in computer-aided detection group.

Fatigue level

The fatigue score for each procedure was 3.28 versus 3.40 for routine and CADe groups, p = 0.357.

Discussion

With the assistance of CADe system, ADR was found significantly increased from 20.91% in routine group to 29.01% in CADe group (P=0.009), APC was also found significantly increased from 0.29 in routine group to 0.48 in CADe group (P<0.001) without introducing more fatigue to endoscopists, which was demonstrated by similar fatigue scores in routine and CADe group (3.28 vs 3.40, p=0.357). Missed polyps during colonoscopy might lead to subsequent CRC, which is one of the leading causes of cancer-related death.[9,10] From a quality-control perspective, small increments in the quality of screening colonoscopies could have significant impact on the net gains of large-scale CRC screening programs.[11] Furthermore, truthful awareness of adenoma numbers for each patient leads to better net gains from screening colonoscopy, including prompt resection of precancerous lesions and a better understanding of the surveillance time.[6,12] Studies have shown that missed polyps fall into one of two categories: those polyps that remain outside the visual field and those polyps that are in the visual field but are unrecognized by the endoscopist.[13] Although polyps that remain outside of the visual field contribute to the majority of missed polyps, polyps that remain unrecognized are also a major issue, and until now only second-observer strategies seemed helpful in increasing the PDR via this pathway.[14-16] However, it is likely that adding additional human observers beyond one would lead to limited gain in ADR and would not fully address the many factors that affect intra-observer variability for a given colonoscopy such as “inattentional blindness”[17,18] and “change blindness.”[19] However, a high-performance CADe[20] system has the advantages of high reproducibility, fidelity and uniformity. The CADe system[3] is thus an ideal way to address unrecognized polyps and has shown good performance in our previous study,[4] especially for those polyps that appear only briefly on the screen, those on the edge of the screen or those that remain partially occluded. The system has also demonstrated good performance in the detection of polyps with subtle visual features including flat and isochromatic polyps, and polyps with unclear boundaries. However, for these polyps, gaze patterns that rely on shuttling back and forth between two screens may lead to missing polyps that do appear in the visual field. Integration of the CADe system into the primary endoscopy monitor may mitigate these issues.[21] A dual-screen set-up allows for specific study designs in a research setting, but it is not conducive to clinical practice. Dual-monitor set-ups in general may lead to increased fatigue for physician participants, and dual-monitor set-ups have been shown to increase fatigue via eye fatigue[22] and overall energy and attention consumption. Similarly, in a vigilance task study,[23] more workload and tense arousal were reported when participants performed actively in a dual-task research scenario. In addition, there appears to be a positive correlation between a center-looking visual gaze pattern (VGP) and ADR,[24] and wider VGP has been demonstrated to allow an increased PDR.[25] These also indicate that frequent shift of sight might harm the best VGP and thus lead to a decrease of ADR or PDR. Therefore, it is crucially important to minimize latency. There are three independent components of latency for a CADe system: the Display Latency of the colonoscopy video overlapped with detection box; the Computation Latency that the convolutional neural network (CNN) takes to produce the detection box for a certain video frame; and the Algorithm Latency due to per-frame sensitivity of the AI algorithm (time discrepancy between the first frame of polyp appearance and the first frame of detection on the polyp). For single-monitor set-up, the display latency should be imperceptible. How fast the CADe system reacts to a polyp appearance mainly depends on the Algorithm Latency due to per-frame sensitivity and Computation Latency. In this study, the Display Latency of the colonoscopy video overlapped with detection box is 5.99 ± 0.24 ms, the Computation Latency is 20.33 ± 0.14 ms, and the per-frame sensitivity is 95.4% as validated in the preclinical study.[3] In this study, the endoscopists could identify the blue box promptly without having to switch between two monitors. Results demonstrated an equal fatigue level between standard endoscopy and endoscopy using the integrated CADe system. Furthermore, the rare consistent false detections, attributed to the high specificity of the CADe system, did not increase the fatigue level of the operating endoscopists in this study. This was also demonstrated by an equal withdrawal time when biopsy time was excluded in two groups. In this study, ADR (20.91% versus 29.01%, p = 0.009) and APC (0.29 versus 0.48, p < 0.001) were shown to have similar increases from the control to the experimental arm as our previous study.[4] The ADR in the control group is much higher than that reported extensively on a low ADR population.[26-30] That indicates the CADe system can contribute to an increased detection rate even when the baseline ADR of participating endoscopists was already qualified, instead of being only useful to endoscopists with low detection rate. The population is representative for normal people with various reasons to receive a colonoscopy compared with a recent study that only included high-risk older populations, mostly patients with CRC symptoms and positive results from a fecal immunochemical test.[7] There was likely a lower limit in ADR in both the control and experimental groups, as the study site utilized older generation Olympus CF-Q260 colonoscopes; new colonoscopes may increase baseline ADR.[31] An integrated CADe system should be further investigated in newer colonoscopy models. As in previous studies, the increase in ADR seen in the experimental group was largely due to the increased detection of diminutive adenomas and hyperplastic polyps. However, in this study, the detection of SSLs was higher in the CADe group in both proximal and distal colon, a fact which may be attributed to the nature of an integrated system. Such a single-monitor system may allow endoscopists to promptly check any quick, subtle alarms and identify more subtle features associated with SSL in real time, whereas in the two-monitor set-up those SSL might be missed during the time interval when the endoscopist turns to look at the second monitor. Consistent false detections by the CADe system (Table 4) and polyps detected first by CADe system before endoscopist in the CADe group (Table 5) were consistent with the previous double-blinded study,[6] which provide concrete evidence of how CADe system makes efficacy during colonoscopy. Moreover, data show that 26.4% (165/625) polyps, 24.0% (73/304) adenomas and 8.3% (1/12) SSLs were detected first by CADe system which was prior to the reaction of the endoscopist’s eye, which indicates not only that the glass-to-glass latency is imperceptibly low, but also proves the high sensitivity of the algorithm that could capture subtle features[32] of a polyp even when it appears briefly on the screen or in suboptimal conditions including insufficient light, blurred lens and so on (Table 5). Also, the super-fast detection by the CADe system indicates a possibility that fewer missed diagnoses would happen with the aid of such a high-performance CADe system. This study has several limitations. First, the study was unblinded, and awareness of the CADe system may have led participating endoscopists to either put more attention into either endoscopy with a “competitive spirit,”[11] or they might relax and unduly rely on the CADe system for polyp detection. Both cases might affect the ADR and lead to a slight overestimation or underestimation of the effectiveness of such CADe system. Similarly, endoscopist scoring of fatigue may also have been affected by the unblinded nature of the study. Second, the control group in this study was routine colonoscopy instead of CADe-assisted colonoscopy with a second monitor, and we thus cannot draw any direct conclusion that an integrated CADe system is more effective or less fatiguing than a dual-monitor system. Third, the fatigue score was subjective and susceptible to factors other than the visual alarms, including insertion difficulty, bowel preparation level, case load per day and so on. Although randomization could balance the overall baseline fatigue factors, larger sample sizes (and a blinded component) might be needed to make the result more robust. Fourth, whether a polyp was first detected by CADe before the endoscopist was based on the operating endoscopist’s own judgment; this might introduce two subjective biases: first man may be reluctant to admit his failure to AI, especially when the operating endoscopist saw the polyp quickly after CADe’s detection box. Second, it is difficult for the endoscopist to tell if the CADe system is a little ahead when they identify a polyp almost at the same time. Thus, the result of how many polyps were initially detected by the CADe system before the endoscopists might be underestimated. Fifth, The fact that the CADe system detected a polyp prior to endoscopists does not necessarily means that the endoscopists would have missed that lesion, and that is hard to be objectively measured, however, faster detection capability of AI is always believed a strong guarantee for endoscopist to miss less. In conclusion, real-time visual alarms provided by a high-performance CADe system embedded into the primary colonoscopy monitor, with nearly unnoticeable latency, have been shown to cause a significant improvement in ADR due to an increased detection of diminutive adenomas without increasing physician fatigue level during colonoscopy. Implementation of a single-screen, augmented-reality CADe system may be an important step toward quality assurance during screening and surveillance colonoscopy. Click here for additional data file. Supplemental material, sj-pdf-1-tag-10.1177_1756284820979165 for The single-monitor trial: an embedded CADe system increased adenoma detection during colonoscopy: a prospective randomized study by Peixi Liu, Pu Wang, Jeremy R. Glissen Brown, Tyler M. Berzin, Guanyu Zhou, Weihui Liu, Xun Xiao, Ziyang Chen, Zhihong Zhang, Chao Zhou, Lei Lei, Fei Xiong, Liangping Li and Xiaogang Liu in Therapeutic Advances in Gastroenterology Click here for additional data file. Supplemental material, sj-pdf-2-tag-10.1177_1756284820979165 for The single-monitor trial: an embedded CADe system increased adenoma detection during colonoscopy: a prospective randomized study by Peixi Liu, Pu Wang, Jeremy R. Glissen Brown, Tyler M. Berzin, Guanyu Zhou, Weihui Liu, Xun Xiao, Ziyang Chen, Zhihong Zhang, Chao Zhou, Lei Lei, Fei Xiong, Liangping Li and Xiaogang Liu in Therapeutic Advances in Gastroenterology Click here for additional data file. Supplemental material, sj-pdf-3-tag-10.1177_1756284820979165 for The single-monitor trial: an embedded CADe system increased adenoma detection during colonoscopy: a prospective randomized study by Peixi Liu, Pu Wang, Jeremy R. Glissen Brown, Tyler M. Berzin, Guanyu Zhou, Weihui Liu, Xun Xiao, Ziyang Chen, Zhihong Zhang, Chao Zhou, Lei Lei, Fei Xiong, Liangping Li and Xiaogang Liu in Therapeutic Advances in Gastroenterology Click here for additional data file. Supplemental material, sj-pdf-4-tag-10.1177_1756284820979165 for The single-monitor trial: an embedded CADe system increased adenoma detection during colonoscopy: a prospective randomized study by Peixi Liu, Pu Wang, Jeremy R. Glissen Brown, Tyler M. Berzin, Guanyu Zhou, Weihui Liu, Xun Xiao, Ziyang Chen, Zhihong Zhang, Chao Zhou, Lei Lei, Fei Xiong, Liangping Li and Xiaogang Liu in Therapeutic Advances in Gastroenterology
  30 in total

1.  Effect of a deep-learning computer-aided detection system on adenoma detection during colonoscopy (CADe-DB trial): a double-blind randomised study.

Authors:  Pu Wang; Xiaogang Liu; Tyler M Berzin; Jeremy R Glissen Brown; Peixi Liu; Chao Zhou; Lei Lei; Liangping Li; Zhenzhen Guo; Shan Lei; Fei Xiong; Han Wang; Yan Song; Yan Pan; Guanyu Zhou
Journal:  Lancet Gastroenterol Hepatol       Date:  2020-01-22

2.  Participation by experienced endoscopy nurses increases the detection rate of colon polyps during a screening colonoscopy: a multicenter, prospective, randomized study.

Authors:  Chang Kyun Lee; Dong Il Park; Suck-Ho Lee; Young Hwangbo; Chang Soo Eun; Dong Soo Han; Jae Myung Cha; Bo-In Lee; Jeong Eun Shin
Journal:  Gastrointest Endosc       Date:  2011-09-01       Impact factor: 9.427

3.  Standard forward-viewing colonoscopy versus full-spectrum endoscopy: an international, multicentre, randomised, tandem colonoscopy trial.

Authors:  Ian M Gralnek; Peter D Siersema; Zamir Halpern; Ori Segol; Alaa Melhem; Alain Suissa; Erwin Santo; Alan Sloyer; Jay Fenster; Leon M G Moons; Vincent K Dik; Ralph B D'Agostino; Douglas K Rex
Journal:  Lancet Oncol       Date:  2014-02-20       Impact factor: 41.316

4.  The impact of regulatory fit on performance in an inattentional blindness paradigm.

Authors:  Daniel Memmert; Christian Unkelbach; Steffen Ganns
Journal:  J Gen Psychol       Date:  2010 Apr-Jun

5.  Lower Adenoma Miss Rate of Computer-Aided Detection-Assisted Colonoscopy vs Routine White-Light Colonoscopy in a Prospective Tandem Study.

Authors:  Pu Wang; Peixi Liu; Jeremy R Glissen Brown; Tyler M Berzin; Guanyu Zhou; Shan Lei; Xiaogang Liu; Liangping Li; Xun Xiao
Journal:  Gastroenterology       Date:  2020-06-17       Impact factor: 22.682

Review 6.  Effectiveness, training and quality assurance of colonoscopy screening for colorectal cancer.

Authors:  Douglas J Robertson; Michal F Kaminski; Michael Bretthauer
Journal:  Gut       Date:  2015-03-24       Impact factor: 23.059

7.  Consensus on the Prevention, Screening, Early Diagnosis and Treatment of Colorectal Tumors in China: Chinese Society of Gastroenterology, October 14-15, 2011, Shanghai, China.

Authors:  Jing-Yuan Fang; Shu Zheng; Bo Jiang; Mao-De Lai; Dian-Chun Fang; Ying Han; Qian-Jiu Sheng; Jing-Nan Li; Ying-Xuan Chen; Qin-Yan Gao
Journal:  Gastrointest Tumors       Date:  2014-05-09

8.  Gaze patterns hold key to unlocking successful search strategies and increasing polyp detection rate in colonoscopy.

Authors:  Mariam Lami; Harsimrat Singh; James H Dilley; Hajra Ashraf; Matthew Edmondon; Felipe Orihuela-Espina; Jonathan Hoare; Ara Darzi; Mikael H Sodergren
Journal:  Endoscopy       Date:  2018-02-07       Impact factor: 10.093

9.  Impact of preprocedure simethicone on adenoma detection rate during colonoscopy: a multicenter, endoscopist-blinded randomized controlled trial.

Authors:  Yu Bai; Jun Fang; Sheng-Bing Zhao; Dong Wang; Yan-Qing Li; Rui-Hua Shi; Zi-Qin Sun; Ming-Jun Sun; Feng Ji; Jian-Min Si; Zhao-Shen Li
Journal:  Endoscopy       Date:  2017-10-06       Impact factor: 10.093

10.  Computer aided detection for laterally spreading tumors and sessile serrated adenomas during colonoscopy.

Authors:  Guanyu Zhou; Xun Xiao; Mengtian Tu; Peixi Liu; Dan Yang; Xiaogang Liu; Renyi Zhang; Liangping Li; Shan Lei; Han Wang; Yan Song; Pu Wang
Journal:  PLoS One       Date:  2020-04-21       Impact factor: 3.240

View more
  9 in total

Review 1.  Artificial Intelligence in Endoscopy.

Authors:  Alexander Hann; Alexander Meining
Journal:  Visc Med       Date:  2021-11-01

Review 2.  Effect of artificial intelligence-aided colonoscopy for adenoma and polyp detection: a meta-analysis of randomized clinical trials.

Authors:  Ding Huang; Jingyi Shen; Jiaze Hong; Yi Zhang; Senjie Dai; Nannan Du; Mengting Zhang; Daxin Guo
Journal:  Int J Colorectal Dis       Date:  2021-11-11       Impact factor: 2.571

Review 3.  Current Status and Future Perspectives of Artificial Intelligence in Colonoscopy.

Authors:  Yu Kamitani; Kouichi Nonaka; Hajime Isomoto
Journal:  J Clin Med       Date:  2022-05-22       Impact factor: 4.964

Review 4.  Scoping out the future: The application of artificial intelligence to gastrointestinal endoscopy.

Authors:  Scott B Minchenberg; Trent Walradt; Jeremy R Glissen Brown
Journal:  World J Gastrointest Oncol       Date:  2022-05-15

5.  Artificial intelligence-assisted detection and classification of colorectal polyps under colonoscopy: a systematic review and meta-analysis.

Authors:  Aling Wang; Jiahao Mo; Cailing Zhong; Shaohua Wu; Sufen Wei; Binqi Tu; Chang Liu; Daman Chen; Qing Xu; Mengyi Cai; Zhuoyao Li; Wenting Xie; Miao Xie; Motohiko Kato; Xujie Xi; Beiping Zhang
Journal:  Ann Transl Med       Date:  2021-11

Review 6.  Artificial intelligence in colonoscopy: A review on the current status.

Authors:  Solveig Linnea Veen Larsen; Yuichi Mori
Journal:  DEN open       Date:  2022-03-23

Review 7.  Randomized Controlled Trials of Artificial Intelligence in Clinical Practice: Systematic Review.

Authors:  Thomas Y T Lam; Max F K Cheung; Yasmin L Munro; Kong Meng Lim; Dennis Shung; Joseph J Y Sung
Journal:  J Med Internet Res       Date:  2022-08-25       Impact factor: 7.076

8.  Randomized Clinical Trials of Machine Learning Interventions in Health Care: A Systematic Review.

Authors:  Deborah Plana; Dennis L Shung; Alyssa A Grimshaw; Anurag Saraf; Joseph J Y Sung; Benjamin H Kann
Journal:  JAMA Netw Open       Date:  2022-09-01

9.  Real-time colorectal polyp detection using a novel computer-aided detection system (CADe): a feasibility study.

Authors:  E Soons; T Rath; Y Hazewinkel; W A van Dop; D Esposito; P A Testoni; P D Siersema
Journal:  Int J Colorectal Dis       Date:  2022-09-27       Impact factor: 2.796

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.