A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters.

W T Cecil1, K J Thorpe, E E Fibuch, G F Tuohy.   

Abstract

Two pulse oximeters, the Nellcor N-100 and the Ohmeda 3700, were compared with arterial blood values and with each other in a clinical evaluation of performance. Three hundred twenty-nine simultaneously sampled blood/oximeter data pairs from use of both makes of pulse oximeters on each of 152 test subjects were included in the comparison analysis for each oximeter. Among the patients, disease type and severity and hospital location varied widely. Basic descriptive statistics and linear regression analysis were employed to facilitate comparison. Both oximeters displayed a statistically significant but clinically insignificant bias when compared with arterial blood oxyhemoglobin: -0.31 (P = 0.023) and 0.59 (P = 0.001) for the Ohmeda 3700 pulse oximeter and the Nellcor N-100 pulse oximeter, respectively. Relative to arterial blood oxyhemoglobin, the 95% tolerance intervals were +4.84 to -5.45 (10.3) for the Ohmeda 3700 and +6.94 to -5.76 (12.7) for the Nellcor N-100. Regressed against [oxyhemoglobin + carboxyhemoglobin + methemoglobin] as x, the Nellcor N-100 read y = 0.85(x) + 12.5, r = 0.83, P less than 0.0001, and the Ohmeda 3700 read y = 1.02(x) - 5.3, r = 0.86, P less than 0.0001.

Entities:  

Mesh:

Year:  1988        PMID: 3339389     DOI: 10.1007/bf01618105

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  5 in total

1.  History of blood gas analysis. VII. Pulse oximetry.

Authors:  J W Severinghaus; Y Honda
Journal:  J Clin Monit       Date:  1987-04

2.  Evaluation of pulse oximetry.

Authors:  M Yelderman; W New
Journal:  Anesthesiology       Date:  1983-10       Impact factor: 7.892

3.  Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter.

Authors:  N Mackenzie
Journal:  J Clin Monit       Date:  1985-07

4.  Statistics and ethics in medical research: V--Analysing data.

Authors:  D G Altman
Journal:  Br Med J       Date:  1980-11-29

5.  Pulse oximetry in pediatric intensive care: comparison with measured saturations and transcutaneous oxygen tension.

Authors:  S Fanconi; P Doherty; J F Edmonds; G A Barker; D J Bohn
Journal:  J Pediatr       Date:  1985-09       Impact factor: 4.406

  5 in total
  10 in total

Review 1.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

Review 2.  Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.

Authors: 
Journal:  CMAJ       Date:  1992-03-01       Impact factor: 8.262

3.  Case-control studies of novel hemoglobin anomalies as differential diagnosis in sleep apnea syndrome.

Authors:  Berndt Zur; Michael Ludwig; Birgit Stoffel-Wagner
Journal:  Sleep Breath       Date:  2012-05-13       Impact factor: 2.816

4.  Pulse oximetry and finger blood pressure measurement during open-heart surgery.

Authors:  T S Kurki; N T Smith; T J Sanford; N Head
Journal:  J Clin Monit       Date:  1989-10

5.  Pulse Oximetry: The Working Principle, Signal Formation, and Applications.

Authors:  Timo Leppänen; Samu Kainulainen; Henri Korkalainen; Saara Sillanmäki; Antti Kulkas; Juha Töyräs; Sami Nikkonen
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 3.650

6.  Racial bias and reproducibility in pulse oximetry among medical and surgical inpatients in general care in the Veterans Health Administration 2013-19: multicenter, retrospective cohort study.

Authors:  Valeria S M Valbuena; Sarah Seelye; Michael W Sjoding; Thomas S Valley; Robert P Dickson; Steven E Gay; Dru Claar; Hallie C Prescott; Theodore J Iwashyna
Journal:  BMJ       Date:  2022-07-06

Review 7.  Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis.

Authors:  Ana M Cabanas; Macarena Fuentes-Guajardo; Katina Latorre; Dayneri León; Pilar Martín-Escudero
Journal:  Sensors (Basel)       Date:  2022-04-29       Impact factor: 3.847

8.  Failure rate of pulse oximetry in the postanesthesia care unit.

Authors:  B S Gillies; K Posner; P Freund; F Cheney
Journal:  J Clin Monit       Date:  1993-11

9.  A prospective study of intraoperative pulse oximetry failure.

Authors:  P R Freund; P T Overand; J Cooper; L Jacobson; S Bosse; B Walker; K L Posner; F W Cheney
Journal:  J Clin Monit       Date:  1991-07

Review 10.  Continuous Noninvasive Hemoglobin Monitoring: A Measured Response to a Critical Review.

Authors:  Steven J Barker; Aryeh Shander; Michael A Ramsay
Journal:  Anesth Analg       Date:  2016-02       Impact factor: 5.108

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.