Pulse oximetry and finger blood pressure measurement during open-heart surgery.

T S Kurki1, N T Smith, T J Sanford, N Head.   

Abstract

Pulse oximeter arterial hemoglobin oxygen saturation (SpO2) and finger arterial pressure (FINAP) were continuously monitored before, during, and after cardiopulmonary bypass in 15 male patients. SpO2 was monitored simultaneously with two pulse oximeters, a Nellcor N-100 and an Ohmeda Biox III. The readings obtained from the two pulse oximeters were compared with arterial blood measurements obtained using a CO-oximeter. FINAP was monitored by a prototype device (Finapres) based on the Penaz volume-clamp method. FINAP was correlated with intraarterial pressure (IAP). Both pulse oximeters functioned well before cardiopulmonary bypass. The correlations with CO-oximeter values were 0.927 for the N-100 and 0.921 for the Biox III. Immediately after the onset of cardiopulmonary bypass, the N-100 pulse oximeter stopped displaying values. The Biox III pulse oximeter continued to display values during the cardiopulmonary bypass period; the correlation with CO-oximeter values was 0.813. After cardiopulmonary bypass, the N-100 began displaying values in 2 to 10 minutes. After cardiopulmonary bypass the correlation with CO-oximeter values was 0.792 for the N-100 and 0.828 for the Biox III pulse oximeter. The Finapres finger blood pressure device functioned well in 13 of 15 patients before cardiopulmonary bypass. The mean bias +/- precision of FINAP-IAP for mean pressure was 8.3 +/- 10.2 mm Hg (SD) and the correlation coefficient was 0.814. During cardiopulmonary bypass, the Finapres device functioned well in 10 of 15 patients. The mean bias precision of FINAP-IAP, for mean pressure in these 10 patients was 6.6 +/- 8.7 mm Hg and the correlation coefficient was 0.902.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1989        PMID: 2809666     DOI: 10.1007/bf01618251

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  24 in total

1.  Minimizing interference and false alarms from electrocautery in the Nellcor N-100 pulse oximeter.

Authors:  F E Block; G J Detko
Journal:  J Clin Monit       Date:  1986-07

2.  Effects of intravenously administered dyes on pulse oximetry readings.

Authors:  M S Scheller; R J Unger; M J Kelner
Journal:  Anesthesiology       Date:  1986-11       Impact factor: 7.892

3.  Venous admixture to radial artery blood samples during cardiopulmonary bypass.

Authors:  S Aronson; W A Neff; S Slogoff; A S Keats
Journal:  Chest       Date:  1987-11       Impact factor: 9.410

4.  Oxygen saturation during preinduction placement of monitoring catheters in the cardiac surgical patient.

Authors:  F A Hensley; D L Dodson; D E Martin; R A Stauffer; D R Larach
Journal:  Anesthesiology       Date:  1987-06       Impact factor: 7.892

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Infrared heat lamps interfere with pulse oximeters.

Authors:  T D Brooks; D A Paulus; W E Winkle
Journal:  Anesthesiology       Date:  1984-11       Impact factor: 7.892

7.  Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a finger.

Authors:  N T Smith; K H Wesseling; B de Wit
Journal:  J Clin Monit       Date:  1985-01

8.  Noninvasive continuous blood pressure measurement from the finger: optimal measurement conditions and factors affecting reliability.

Authors:  T Kurki; N T Smith; N Head; H Dec-Silver; A Quinn
Journal:  J Clin Monit       Date:  1987-01

9.  Accuracy of response of six pulse oximeters to profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

10.  Availability of intraarterial pressure waveforms from catheter-manometer systems during surgery.

Authors:  K H Wesseling; N T Smith
Journal:  J Clin Monit       Date:  1985-01
View more
  3 in total

Review 1.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

2.  Intraoperative Oxygen Practices in Cardiac Surgery: A National Survey.

Authors:  Anthony Calhoun; Ameeka Pannu; Ariel L Mueller; Omar Elmadhoun; Juan D Valencia; Megan L Krajewski; Brian P O'Gara; Anastasia Katsiampoura; Sean T O'Connor; Louis Chu; Erika Monteith; Puja Shankar; Kyle Spear; Shahzad Shaefi
Journal:  J Cardiothorac Vasc Anesth       Date:  2022-01-19       Impact factor: 2.894

3.  Feasibility of Brachial Occlusion Technique for Beat-to-Beat Pulse Wave Analysis.

Authors:  Lukas Matera; Pavol Sajgalik; Vratislav Fabian; Yegor Mikhailov; David Zemanek; Bruce D Johnson
Journal:  Sensors (Basel)       Date:  2022-09-26       Impact factor: 3.847

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.