Literature DB >> 3831256

Comparison of a pulse oximeter with an ear oximeter and an in-vitro oximeter.

N Mackenzie.   

Abstract

A pulse oximeter was compared with an ear oximeter for measurement of arterial hemoglobin saturation within the range of 70 to 100% in 11 healthy volunteer subjects. Two hundred seventy-seven pooled data points were obtained, and analysis was performed by means of linear regression. The accuracy was 3% (95% confidence limits). The pulse oximeter was also compared with an in-vitro oximeter, and accuracy was within 2%. The pulse oximeter was easy to use because, unlike the ear oximeter, it required no time-consuming instrument calibration or site preparation. In addition, the delays involved in taking an invasive sample, transporting it to the blood gas laboratory, and waiting for the results were eliminated. Saturation values were continually available, and placement and use of the pulse oximeter sensor caused no discomfort to the volunteer subjects.

Mesh:

Substances:

Year:  1985        PMID: 3831256     DOI: 10.1007/bf02832255

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  2 in total

1.  Evaluation of pulse oximetry.

Authors:  M Yelderman; W New
Journal:  Anesthesiology       Date:  1983-10       Impact factor: 7.892

2.  Spectrophotometric monitoring of arterial oxygen saturation in the fingertip.

Authors:  I Yoshiya; Y Shimada; K Tanaka
Journal:  Med Biol Eng Comput       Date:  1980-01       Impact factor: 2.602

  2 in total
  6 in total

Review 1.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

2.  A clinical evaluation of the accuracy of the Nellcor N-100 and Ohmeda 3700 pulse oximeters.

Authors:  W T Cecil; K J Thorpe; E E Fibuch; G F Tuohy
Journal:  J Clin Monit       Date:  1988-01

3.  Pulse oximetry: a "which" hunt?

Authors:  M K Sykes
Journal:  J Clin Monit       Date:  1989-04

4.  Evaluation of the Datascope ACCUSAT pulse oximeter in healthy adults.

Authors: 
Journal:  J Clin Monit       Date:  1989-01

5.  Pulse oximetry during cardiac catheterization in children with congenital heart disease.

Authors:  A M Lynn; A Bosenberg
Journal:  J Clin Monit       Date:  1986-10

6.  Pulse oximetry in the accident and emergency department.

Authors:  C J Holburn; M J Allen
Journal:  Arch Emerg Med       Date:  1989-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.