Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part II: System modeling, scatter correction, and optimization.

Adam Wang1, Alexander Maslowski1, Philippe Messmer1, Mathias Lehmann1, Adam Strzelecki1, Elaine Yu1, Pascal Paysan1, Marcus Brehm1, Peter Munro1, Josh Star-Lack1, Dieter Seghers1.   

Abstract

PURPOSE: To correct for scatter in kV cone-beam CT (CBCT) projection data on a clinical system using a new tool, Acuros® CTS, that estimates scatter images rapidly and accurately by deterministically solving the linear Boltzmann transport equation.
METHODS: Phantom and patient CBCT scans were acquired on TrueBeam® radiotherapy machines. A first-pass reconstruction was used to create water and bone density maps of the imaged object, which was updated to include a more accurate representation of the patient couch. The imaging system model accounted for the TrueBeam x-ray source (polychromatic spectrum, beam filtration, bowtie filter, and collimation hardware) and x-ray detection system (antiscatter grid, flat-panel imager). Acuros CTS then used the system and object models to estimate the scatter component of each projection image, which was subtracted from the measured projections. The corrected projections were then reconstructed to produce the final result. We examined the tradeoff between run time and accuracy using a Pareto optimization of key parameters, including the voxel size of the down-sampled object model, the number of pixels in the down-sampled detector, and the number of scatter images (angular down-sampling). All computations and reconstructions were performed on a research workstation containing two graphics processing units (GPUs). In addition, we established a method for selecting a subset of projections for which scatter images were calculated. The projections were selected to minimize interpolation errors in the remaining projections. Image quality improvement was assessed by measuring the accuracy of the reconstructed phantom and patient images.
RESULTS: The Pareto optimization yielded a set of parameters with an average run time of 26 seconds for scatter correction while maintaining high accuracy of scatter estimation. This was achieved in part by means of optimizing the projection angles that were processed, thus favoring the use of more angles in the lateral (i.e., horizontal) direction and fewer angles in the AP direction. In a 40 cm solid water phantom reconstruction, nonuniformities were decreased from 217 HU without scatter correction to 51 HU with conventional (kernel-based) scatter correction to 17 HU with Acuros CTS-based scatter correction. In clinical pelvis scans, nonuniformities in the bladder were reduced from 85 HU with conventional scatter correction to 14 HU with Acuros CTS.
CONCLUSIONS: Acuros CTS is a promising new tool for fast and accurate scatter correction for CBCT imaging. By carefully modeling the imaging chain and optimizing several parameters, we achieved high correction accuracies with computation times compatible with the clinical workflow. The improvement in image quality enables better soft-tissue visualization and potentially enables applications such as adaptive radiotherapy.
© 2018 American Association of Physicists in Medicine.

Entities:  

Keywords:  Monte Carlo; cone-beam CT; scatter

Mesh:

Year:  2018        PMID: 29509973     DOI: 10.1002/mp.12849

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  Characterization of a prototype rapid kilovoltage x-ray image guidance system designed for a ring shape radiation therapy unit.

Authors:  Bin Cai; Eric Laugeman; Thomas R Mazur; Justin C Park; Lauren E Henke; Hyun Kim; Geoffrey D Hugo; Sasa Mutic; Hua Li
Journal:  Med Phys       Date:  2019-02-13       Impact factor: 4.071

2.  Acuros CTS: A fast, linear Boltzmann transport equation solver for computed tomography scatter - Part I: Core algorithms and validation.

Authors:  Alexander Maslowski; Adam Wang; Mingshan Sun; Todd Wareing; Ian Davis; Josh Star-Lack
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

3.  Fast shading correction for cone-beam CT via partitioned tissue classification.

Authors:  Linxi Shi; Adam Wang; Jikun Wei; Lei Zhu
Journal:  Phys Med Biol       Date:  2019-03-13       Impact factor: 3.609

4.  Clinical implementation of HyperArc.

Authors:  Felix H C Wong; Puleng A Moleme; Omer A Ali; Koki V Mugabe
Journal:  Phys Eng Sci Med       Date:  2022-04-11

5.  Validation of a deterministic linear Boltzmann transport equation solver for rapid CT dose computation using physical dose measurements in pediatric phantoms.

Authors:  Sara Principi; Yonggang Lu; Yu Liu; Adam Wang; Alex Maslowski; Todd Wareing; John Van Heteren; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2021-10-29       Impact factor: 4.071

6.  A Comparison Study Between CNN-Based Deformed Planning CT and CycleGAN-Based Synthetic CT Methods for Improving iCBCT Image Quality.

Authors:  Bo Yang; Yankui Chang; Yongguang Liang; Zhiqun Wang; Xi Pei; Xie George Xu; Jie Qiu
Journal:  Front Oncol       Date:  2022-05-30       Impact factor: 5.738

7.  Deterministic linear Boltzmann transport equation solver for patient-specific CT dose estimation: Comparison against a Monte Carlo benchmark for realistic scanner configurations and patient models.

Authors:  Sara Principi; Adam Wang; Alexander Maslowski; Todd Wareing; Petr Jordan; Taly Gilat Schmidt
Journal:  Med Phys       Date:  2020-10-20       Impact factor: 4.071

8.  Evaluation of scatter rejection and correction performance of 2D antiscatter grids in cone beam computed tomography.

Authors:  Yeonok Park; Timur Alexeev; Brian Miller; Moyed Miften; Cem Altunbas
Journal:  Med Phys       Date:  2021-03-04       Impact factor: 4.071

9.  A unified scatter rejection and correction method for cone beam computed tomography.

Authors:  Cem Altunbas; Yeonok Park; Zhelin Yu; Anant Gopal
Journal:  Med Phys       Date:  2021-02-06       Impact factor: 4.071

10.  Improvements in CBCT Image Quality Using a Novel Iterative Reconstruction Algorithm: A Clinical Evaluation.

Authors:  Stephen J Gardner; Weihua Mao; Chang Liu; Ibrahim Aref; Mohamed Elshaikh; Joon K Lee; Deepak Pradhan; Benjamin Movsas; Indrin J Chetty; Farzan Siddiqui
Journal:  Adv Radiat Oncol       Date:  2019-01-10