Microbial Contamination Risk and Disinfection of Radiation Protective Garments.

Stephen Balter1, Michelle A Rodriguez2, Janett A Pike3, Norman J Kleiman4.   

Abstract

ABSTRACT: Healthcare-associated infections are a major public health concern for both patients and medical personnel. This has taken on greater urgency during the current COVID-19 pandemic. Radiation Personal Protective Equipment (RPPE) may contribute to risks of microbial contamination. This possibility was tested in 61 personal or shared-use lead aprons and thyroid collars at Columbia Presbyterian Irving Medical Center. Fifty percent tested positive for either bacterial or fungal contamination, mostly around the neckline of lead vests and thyroid collars. Repeated testing of garments some weeks to months later confirmed continued presence of microbial contamination. The possibility that hospital-approved disinfection agents could degrade the radio-protective features of these garments was also examined. Samples of identical construction to garments in regular use were subjected to either daily or weekly wipes with hypochlorite or alcohol-based hospital-approved cleaning agents for 6 mo. A third group of samples was maintained in contact with the cleaning agents for 6 mo. All samples were fluoroscoped four times during the study. None demonstrated any degradation in radioprotection. All samples were photographed monthly. Physical degradation of the outer plastic covering by concentrated hypochlorite and limited mechanical damage around stitched seams of the samples cleaned daily with alcohol was noted. Based on the high prevalence of microbial contamination, regular cleaning and disinfection protocols should be implemented. Regular cleaning with medical-facility-approved cleaning and disinfecting agents is likely to be effective at reducing the microbial load and unlikely to result in significant reduction in radioprotective properties of these garments.
Copyright © 2021 Health Physics Society.

Entities:  

Mesh:

Year:  2021        PMID: 33369969      PMCID: PMC7872069          DOI: 10.1097/HP.0000000000001387

Source DB:  PubMed          Journal:  Health Phys        ISSN: 0017-9078            Impact factor:   2.922


  18 in total

1.  Inspection of lead aprons: criteria for rejection.

Authors:  K Lambert; T McKeon
Journal:  Health Phys       Date:  2001-05       Impact factor: 1.316

2.  Infection control - a problem for patient safety.

Authors:  John P Burke
Journal:  N Engl J Med       Date:  2003-02-13       Impact factor: 91.245

3.  How protective are the lead aprons we use against ionizing radiation?

Authors:  Orhan Oyar; Arzu Kışlalıoğlu
Journal:  Diagn Interv Radiol       Date:  2011-09-24       Impact factor: 2.630

4.  Occupational health hazards in the interventional laboratory: time for a safer environment.

Authors:  Lloyd W Klein; Donald L Miller; Stephen Balter; Warren Laskey; David Haines; Alexander Norbash; Matthew A Mauro; James A Goldstein
Journal:  Radiology       Date:  2009-02       Impact factor: 11.105

5.  Potential infection risk from thyroid radiation protection.

Authors:  Siegfried Feierabend; Geoff Siegel
Journal:  J Orthop Trauma       Date:  2015-01       Impact factor: 2.512

6.  Surveillance of microbiological contamination and correct use of protective lead garments.

Authors:  V La Fauci; R Riso; A Facciolà; V Merlina; R Squeri
Journal:  Ann Ig       Date:  2016 Sep-Oct

7.  Occupational health hazards in the interventional laboratory: Time for a safer environment.

Authors:  Lloyd W Klein; Donald L Miller; Stephen Balter; Warren Laskey; Neil Naito; David Haines; Allan Ross; Matthew A Mauro; James A Goldstein
Journal:  Catheter Cardiovasc Interv       Date:  2018-01-04       Impact factor: 2.692

8.  Cross infection in the high-touch areas of dental radiology clinics.

Authors:  M Rahmatulla; K Almas; N al-Bagieh
Journal:  Indian J Dent Res       Date:  1996 Jul-Sep

9.  Isolation gowns: a false sense of security?

Authors:  S A Lovitt; R L Nichols; J W Smith; A C Muzik; P F Pearce
Journal:  Am J Infect Control       Date:  1992-08       Impact factor: 2.918

10.  Evaluation of bacterial presence on lead X-ray aprons utilised in the operating room via IBIS and standard culture methods.

Authors:  Sameer Jain; Rebecca A Rajfer; Rachel Melton-Kreft; Laura Nistico; Mark C Miller; Paul Stoodley; Daniel T Altman; Gregory T Altman
Journal:  J Infect Prev       Date:  2019-04-15
View more
  2 in total

1.  European consensus on patient contact shielding.

Authors:  Peter Hiles; Patrick Gilligan; John Damilakis; Eric Briers; Cristian Candela-Juan; Dario Faj; Shane Foley; Guy Frija; Claudio Granata; Hugo de Las Heras Gala; Ruben Pauwels; Marta Sans Merce; Georgios Simantirakis; Eliseo Vano
Journal:  Insights Imaging       Date:  2021-12-23

2.  A phantom study of a protective trolley for neonatal radiographic imaging: new equipment to protect the operator from scatter radiation.

Authors:  Xinhong Wang; Mengxi Xu; Cong Chen; Zhongkun Bao; Hongzhu Wang; Jizhou Zhang; Zhen Wang; Haipeng Liu
Journal:  Pediatr Radiol       Date:  2022-08-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.