Continuous Glucose Monitoring in the Intensive Care Unit During the COVID-19 Pandemic.

Shivani Agarwal1,2, Justin Mathew2, Georgia M Davis3, Alethea Shephardson2, Ann Levine2, Rita Louard4,2, Agustina Urrutia3, Citlalli Perez-Guzman3, Guillermo E Umpierrez3, Limin Peng3, Francisco J Pasquel3.   

Abstract

OBJECTIVE: Real-time continuous glucose monitoring (rtCGM) in critically ill hospitalized patients holds promise; however, real-world data are needed. RESEARCH DESIGN AND METHODS: We placed Dexcom G6 CGM on intensive care unit (ICU) patients at Montefiore Medical Center with confirmed coronavirus disease 2019 (COVID-19) infection and glycemic variability. We analyzed inpatient CGM accuracy using point-of-care (POC) glucose-CGM matched pairs and included patients for analysis regardless of clinical status.
RESULTS: We included 11 patients with CGM: 8 on continuous insulin infusion (CII), 8 on vasopressors, 8 intubated, 4 on high-dose glucocorticoids, 6 on renal replacement therapy, and 2 with anasarca. Accuracy was 12.58% for mean and 6.3% for median absolute relative difference. CGM reduced POC testing by ∼60% for patients on CII.
CONCLUSIONS: In this real-world preliminary analysis of rtCGM during critical illness, we demonstrate early feasibility, considerable accuracy, and meaningful reduction in the frequency of POC glucose testing.
© 2020 by the American Diabetes Association.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 33361145      PMCID: PMC7896255          DOI: 10.2337/dc20-2219

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


Introduction

The coronavirus disease 2019 (COVID-19) pandemic has highlighted the urgent need to develop innovative care solutions to manage hyperglycemia in hospitalized patients while preserving resources. CGM use in the critically ill patient in the intensive care unit (ICU) offers unique benefits by allowing real-time tracking of glucose levels to detect often unrecognized severe hypo- or hyperglycemic events, while decreasing the high burden of frequent point-of-care (POC) testing during clinically tenuous states and continuous insulin infusion (CII) (1). Recent studies of real-time (rt)CGM use in non-ICU patients have demonstrated promising results in inpatient settings (2–4). However, these studies excluded critically ill patients. During the pandemic, use of CGM in the ICU could have transformative results in enabling optimized glycemic control, reducing exposure time of staff, and decreasing resource use. Nevertheless, scant data exist on the use of modern rtCGM devices in the ICU. We implemented CGM in the ICU at Montefiore Medical Center during the COVID-19 pandemic and report our results on the performance and accuracy in a real-world setting. We purposely studied critically ill patients, who received therapies historically excluded from inpatient CGM studies, to provide vital early data on CGM use in the ICU.

Research Design and Methods

Following the U.S. Food and Drug Administration statement of nonobjection to inpatient CGM use during the pandemic (1,5), hospital leadership allowed clinical use of CGM in the ICU at Montefiore Medical Center. This retrospective analysis of clinical data was approved by the Albert Einstein College of Medicine Institutional Review Board (Bronx, NY).

Patient Eligibility and Selection

Patients were eligible if they had multiple blood glucose readings in the hyperglycemic (>180 mg/dL) or hypoglycemic range (<70 mg/dL) and were hospitalized in the ICU during the COVID-19 pandemic from 1 April to 30 May 2020 at Montefiore Medical Center. Patients remained eligible if they were receiving mechanical ventilation, vasopressors, CII, or renal replacement therapy (RRT). Exclusion criteria involved receiving high-dose ascorbic acid or acetaminophen >4 g/day. Selection of patients was based on requests for CGM placement by hospital staff and review of inclusion/exclusion criteria by an internal inpatient CGM committee (authors S.A., R.L., A.S., and A.L.).

CGM Placement and Documentation

After patient selection, a trained diabetes nurse practitioner or nurse specialist placed a G6 sensor and transmitter onto the patient’s abdomen or upper arm in cases of prone positioning. CGM receivers were placed outside of the patient’s door for visualization of real-time glucose data. Recorded blood glucose values were monitored in real-time by the inpatient CGM committee for validation of sensor and POC glucose readings according to Food and Drug Administration guidelines (±20% of each other [for blood glucose ≥100 mg/dL] or ±20 mg/dL of each other [for POC blood glucose ≤100 mg/dL]) (6,7). Alarm triggers were set to <100 and ≥250 mg/dL and 3 mg/dL/min rate of glucose rise or drop. After validation, CGM values were used to monitor glycemic trends, adjust insulin dosing, and alarm for severe hypo- or hyperglycemic events, which triggered confirmatory POC testing. At minimum, once-daily POC measurement was required to ensure maintained CGM validation.

Statistical Analysis

Descriptive data are presented as mean ± SD for continuous variables (age, HbA1c, time in range) and count (%) for categorical variables (race/ethnicity, diabetes type, comorbidities). We collected CGM and POC glucose matched pairs (n = 493) to calculate the mean and median average relative difference (MARD and median ARD %) between measurements. We also calculated the proportion of CGM values within ±15, 20, and 30% of POC reference values for glucose levels >100 mg/dL and ±15, 20 or 30 mg/dL for POC glucose levels ≤100 mg/dL (%15/15, %20/20, %30/30) (8,9). Clarke error grid analysis was used to assess clinical reliability of matched POC-CGM glucose pairs (9).

Results

Overall (n = 11), mean age was 56 ± 17 years, six patients were male, and eight belonged to underrepresented racial-ethnic groups (Hispanic or non-Hispanic Black). Six patients had preexisting type 2 diabetes, three had type 1 diabetes, and two had no known diabetes. Mean HbA1c was 8.7 ± 2.5%. At the time of CGM use, nine patients were on CII, eight received vasopressors, eight were intubated, four received high-dose glucocorticoids, six were on RRT, and two had anasarca. Median number of days on CGM was 9.0 (3.0–10.0). Mean time in range was 46.1 ± 15.8%, above range was 53.3 ± 15.8%, and below range was 0.6 ± 0.8%. Clarke error grid analysis and MARD are reported in Fig. 1. In sum, 493 paired POC-CGM measurements were used for analysis. Regarding accuracy, MARD and median ARD were 12.58% and 6.3%, respectively. We found 77.7% of values fell within zone A (CGM within 20% of POC), 20.5% zone B (>20% difference, no incorrect treatment), 0.2% zone C (hyperglycemia or hypoglycemia leading to inappropriate treatment), 1.2% zone D (undetected hypoglycemia or hyperglycemia needing treatment), and 0.4% zone E (hypoglycemia mistaken for hyperglycemia, and vice versa) (Fig. 1).
Figure 1

Clarke error grid analysis showing the relationship between CGM and POC glucose values in critically ill patients with COVID-19 (n = 11).

Clarke error grid analysis showing the relationship between CGM and POC glucose values in critically ill patients with COVID-19 (n = 11). To calculate reduction in POC testing while using CGM, we calculated the mean number of hours/potential POC tests while on CII versus actual POC tests (n = 8 patients). Mean number of hours/potential POC tests on CII was 72.125 h/∼72 tests, compared with mean number of actual POC tests on CGM of 28, representing an estimated 60% reduction in POC tests during rtCGM use.

Conclusions

We present our firsthand real-world experience of rtCGM in the ICU during the COVID-19 pandemic in New York City. Our findings show that use of rtCGM in the ICU setting is feasible, acceptable, and reliable as an adjunctive modality to POC glucose measurements. CGM use resulted in a meaningful reduction in POC testing frequency for patients requiring CII. Overall, our study underscores the great potential of CGM to aid in glucose testing for critically ill patients. While other studies have systematically excluded the sickest patients, we included patients with multiple critical illnesses and therapies, which is a particular strength of our study. Overall, 98% of all sensor readings had clinically acceptable correlation (zones A + B). Currently, there are few reports of remote monitoring using the G6 device in the ICU. Compared with findings from recent small studies using G6 in medical-surgical patients on hospital floor units (2,10,11), our MARD (12.58%) falls within the range of these reports (9.4–12.8%) (2,10,11). Our accuracy results are even more encouraging given our eligibility criteria including critically ill patients with potentially interfering treatments, compared with prior studies. Nevertheless, concerns remain regarding complete replacement of POC glucose testing with CGM (1,12–14), given that our data reveal several instances of potential failures to detect hypo- or hyperglycemia (zone D) or mistreatment (zone E). Thus, while promising, caution must be exercised in utilizing CGM as the sole modality for glucose testing. Limitations of our study include its small sample size and observational nature. While inclusion of critically ill patients is a strength and our accuracy results are in line with previous studies excluding sick patients, further studies are required to evaluate CGM safety and accuracy in a larger critically ill population. Differentiation of results by potentially interfering treatments is needed. In sum, our results provide proof of concept that rtCGM can be used safely and effectively during critical illness, while offering reliable accuracy and decreasing POC glucose testing frequency. Overall, CGM use in the ICU setting should be further explored as a treatment aid for inpatient diabetes management of the critically ill patient.
  14 in total

1.  Statistical tools to analyze continuous glucose monitor data.

Authors:  William Clarke; Boris Kovatchev
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

Review 2.  Diabetes Technology in the Inpatient Setting for Management of Hyperglycemia.

Authors:  Georgia M Davis; Rodolfo J Galindo; Alexandra L Migdal; Guillermo E Umpierrez
Journal:  Endocrinol Metab Clin North Am       Date:  2020-03       Impact factor: 4.741

3.  Usefulness and Safety of Remote Continuous Glucose Monitoring for a Severe COVID-19 Patient with Diabetes.

Authors:  Emi Ushigome; Masahiro Yamazaki; Masahide Hamaguchi; Takamasa Ito; Shin Matsubara; Yasuhiro Tsuchido; Yu Kasamatsu; Masaki Nakanishi; Naohisa Fujita; Michiaki Fukui
Journal:  Diabetes Technol Ther       Date:  2020-07-15       Impact factor: 6.118

4.  A Pilot Study of the Feasibility and Accuracy of Inpatient Continuous Glucose Monitoring.

Authors:  Bala G Nair; E Patchen Dellinger; David R Flum; G Alec Rooke; Irl B Hirsch
Journal:  Diabetes Care       Date:  2020-05-11       Impact factor: 19.112

5.  Performance of a Factory-Calibrated Real-Time Continuous Glucose Monitoring System Utilizing an Automated Sensor Applicator.

Authors:  Viral N Shah; Lori M Laffel; R Paul Wadwa; Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2018-06       Impact factor: 6.118

6.  The performance of flash glucose monitoring in critically ill patients with diabetes.

Authors:  Paolo Ancona; Glenn M Eastwood; Luca Lucchetta; Elif I Ekinci; Rinaldo Bellomo; Johan Mårtensson
Journal:  Crit Care Resusc       Date:  2017-06       Impact factor: 2.159

Review 7.  Measures of Accuracy for Continuous Glucose Monitoring and Blood Glucose Monitoring Devices.

Authors:  Guido Freckmann; Stefan Pleus; Mike Grady; Steven Setford; Brian Levy
Journal:  J Diabetes Sci Technol       Date:  2018-11-19

8.  Individualizing Inpatient Diabetes Management During the Coronavirus Disease 2019 Pandemic.

Authors:  Francisco J Pasquel; Guillermo E Umpierrez
Journal:  J Diabetes Sci Technol       Date:  2020-05-05

9.  Remote Glucose Monitoring of Hospitalized, Quarantined Patients With Diabetes and COVID-19.

Authors:  Gilat Shehav-Zaltzman; Gad Segal; Noa Konvalina; Amir Tirosh
Journal:  Diabetes Care       Date:  2020-05-14       Impact factor: 19.112

10.  Feasibility of Inpatient Continuous Glucose Monitoring During the COVID-19 Pandemic: Early Experience.

Authors:  Sirimon Reutrakul; Matthew Genco; Harley Salinas; Robert M Sargis; Carlie Paul; Yuval Eisenberg; Jiali Fang; Rachel N Caskey; Sarah Henkle; Sam Fatoorehchi; Amanda Osta; Pavan Srivastava; Alexia Johnson; Sarah E Messmer; Michelle Barnes; Sarida Pratuangtham; Brian T Layden
Journal:  Diabetes Care       Date:  2020-08-07       Impact factor: 19.112

View more
  17 in total

Review 1.  Blood Sugar Targets in Surgical Intensive Care—Management and Special Considerations in Patients With Diabetes

Authors:  Johannes Roth; Oliver Sommerfeld; Andreas L Birkenfeld; Christoph Sponholz; Ulrich A Müller; Christian von Loeffelholz
Journal:  Dtsch Arztebl Int       Date:  2021-09-17       Impact factor: 5.594

2.  Evaluating perioperative glycemic status after different types of pancreatic surgeries via continuous glucose monitoring system: a pilot study.

Authors:  Yishen Mao; Xingfei Zhao; Lihui Zhou; Bin Lu; Chen Jin; Deliang Fu; Lie Yao; Ji Li
Journal:  Gland Surg       Date:  2021-10

3.  Hospital Diabetes Meeting 2022.

Authors:  Jingtong Huang; Andrea M Yeung; Kevin T Nguyen; Nicole Y Xu; Jean-Charles Preiser; Robert J Rushakoff; Jane Jeffrie Seley; Guillermo E Umpierrez; Amisha Wallia; Andjela T Drincic; Roma Gianchandani; M Cecilia Lansang; Umesh Masharani; Nestoras Mathioudakis; Francisco J Pasquel; Signe Schmidt; Viral N Shah; Elias K Spanakis; Andreas Stuhr; Gerlies M Treiber; David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2022-07-29

4.  Subcutaneous continuous glucose monitoring in critically ill patients during insulin therapy: a meta-analysis.

Authors:  Yan Yao; Yi-He Zhao; Wen-He Zheng; Hui-Bin Huang
Journal:  Am J Transl Res       Date:  2022-07-15       Impact factor: 3.940

Review 5.  The patient-centered diabetes management during the COVID-19 pandemic.

Authors:  Noriko Kodani; Mitsuru Ohsugi
Journal:  Glob Health Med       Date:  2022-08-31

6.  A Fatal Case of Takotsubo Cardiomyopathy Secondary to Refractory Hypoglycemia in Severe Starvation: An Autopsy Case Report.

Authors:  Jin Kirigaya; Noriaki Iwahashi; Reiko Tanaka; Yoshiaki Inayama; Ichiro Takeuchi
Journal:  Cureus       Date:  2022-03-18

7.  Remote Continuous Glucose Monitoring With a Computerized Insulin Infusion Protocol for Critically Ill Patients in a COVID-19 Medical ICU: Proof of Concept.

Authors:  Georgia M Davis; Eileen Faulds; Tara Walker; Debbie Vigliotti; Marina Rabinovich; Joi Hester; Limin Peng; Barbara McLean; Patricia Hannon; Norma Poindexter; Petrena Saunders; Citlalli Perez-Guzman; Seema S Tekwani; Greg S Martin; Guillermo Umpierrez; Shivani Agarwal; Kathleen Dungan; Francisco J Pasquel
Journal:  Diabetes Care       Date:  2021-02-09       Impact factor: 19.112

8.  Characteristics of and Mortality Associated With Diabetic Ketoacidosis Among US Patients Hospitalized With or Without COVID-19.

Authors:  Francisco J Pasquel; Jordan Messler; Robert Booth; Beata Kubacka; April Mumpower; Guillermo Umpierrez; Joseph Aloi
Journal:  JAMA Netw Open       Date:  2021-03-01

Review 9.  Continuous glucose monitoring in the hospital: an update in the era of COVID-19.

Authors:  Chikara Gothong; Lakshmi G Singh; Medha Satyarengga; Elias K Spanakis
Journal:  Curr Opin Endocrinol Diabetes Obes       Date:  2022-02-01       Impact factor: 3.243

Review 10.  Diabetes and COVID-19; A Bidirectional Interplay.

Authors:  Paraskevi Kazakou; Vaia Lambadiari; Ignatios Ikonomidis; Aikaterini Kountouri; Georgios Panagopoulos; Stavros Athanasopoulos; Eleni Korompoki; Ioannis Kalomenidis; Meletios A Dimopoulos; Asimina Mitrakou
Journal:  Front Endocrinol (Lausanne)       Date:  2022-02-17       Impact factor: 5.555

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.