Joo Han Oh1, Sanghyeon Lee2, Sung-Min Rhee1, Hyeon Jang Jeong3, Jae Chul Yoo2.
Abstract
BACKGROUND: Glenoid baseplate location is important to good clinical outcomes of reverse total shoulder arthroplasty (RTSA). The glenoid vault is the determining factor for glenoid baseplate location, but, to date, there are no reports on the effect of central cage location within the glenoid vault on RTSA outcomes when using the Exactech® Equinoxe® Reverse System. The purpose of this study was to determine the appropriate cage location in relation to the glenoid vault and monitor for vault and/or cortex penetration by the cage.Entities:
Keywords: Anatomy; Arthropathies; Arthroplasty; Rotator cuff injuries; Shoulder
Year: 2019 PMID: 33330190 PMCID: PMC7713875 DOI: 10.5397/cise.2019.22.1.24
Source DB: PubMed Journal: Clin Shoulder Elb ISSN: 1226-9344
Fig. 1.Illustration of glenoid baseplate (Equinoxe® Reverse system; Exactech®, Gainesville, FL, USA) (permission from Exactech).
Fig. 2.Evaluation of glenoid vault depth in preoperative computed tomography scan.
Fig. 3.Estimation of central cage center (permission from Exactech).
A: inferior glenoid rim–center distance, B: inferior glenoid rim–cage distance, C: center–cage center distance=‘A’+3.75 mm (half radius of central peg)-‘B’.
Fig. 4.Group A. Out of the vault, superior position. (A) Preoperative X-ray and computed tomography (CT) scan. (B) Postoperative X-ray and CT scan.
Fig. 5.Group B. In the vault, superior position. (A) Preoperative X-ray and computed tomography (CT) scan. (B) Postoperative X-ray and CT scan.
Fig. 6.Group C. In the vault, appropriate position. (A) Preoperative X-ray and computed tomography (CT) scan. (B) Postoperative X-ray and CT scan.
Cage Position within the Glenoid Vault
| Variable | Group A[ | Group B[ | Group C[ |
|---|---|---|---|
| SNUBH (n=9) | 3 (33.3) | 2 (22.2) | 4 (44.4) |
| SMC (n=13) | 4 (30.8) | 4 (30.8) | 5 (38.5) |
Values are presented as number (%).
SNUBH: Seoul National University Bundang Hospital, SMC: Samsung Medical Center.
Cage out of the vault, superior position;
Cage in the vault, superior position;
Cage in the vault, appropriate position.
Center–Prosthesis Distance
| Variable | Mean center–prosthesis distance (mm) | Appropriate fixation group distance (mm) | Inappropriate fixation group distance (mm) |
|---|---|---|---|
| SNUBH | 5.0 | 4.37 | 6.30 |
| SMC | 5.21 | 3.15 | 7.27 |
SNUBH: Seoul National University Bundang Hospital, SMC: Samsung Medical Center.