| Literature DB >> 33324865 |
Hiraku Iramina1, Mitsuhiro Nakamura, Yuki Miyabe1, Nobutaka Mukumoto1, Tomohiro Ono1, Hideaki Hirashima1, Takashi Mizowaki.
Abstract
OBJECTIVE: To quantify and correct megavoltage (MV) scattered X-rays (MV-scatter) on an image acquired using a linac-mounted kilovoltage (kV) imaging subsystem. METHODS AND MATERIALS: A linac-mounted flat-panel detector (FPD) was used to acquire an image containing MV-scatter by activating the FPD only during MV beam irradiation. 6-, 10-, and 15 MV with a flattening-filter (FF; 6X-FF, 10X-FF, 15X-FF), and 6- and 10 MV without an FF (6X-FFF, 10X-FFF) were used. The maps were acquired by changing one of the irradiation parameters while the others remained fixed. The mean pixel values of the MV-scatter were normalized to the 6X-FF reference condition (MV-scatter value). An MV-scatter database was constructed using these values. An MV-scatter correction experiment with one full arc image acquisition and two square field sizes (FSs) was conducted. Measurement- and estimation-based corrections were performed using the database. The image contrast was calculated at each angle.Entities:
Year: 2020 PMID: 33324865 PMCID: PMC7731796 DOI: 10.1259/bjro.20190048
Source DB: PubMed Journal: BJR Open ISSN: 2513-9878
Reference condition for the MV-scatter map, variable parameters, and scatterers used in this study
| Parameter | Description | Scatterer |
|---|---|---|
| Reference condition for MV-scatter map acquisition | Field size: 10.0 × 10.0 cm2, dose rate: 400 MU/min, gantry and collimator angles: 0°, flat-panel detector (FPD) position from isocenter: 70 cm | Cuboid |
| Field size [cm2] | 2.5 × 2.5, 5.0 × 5.0, 7.5 × 7.5, 10.0 × 10.0, 12.5 × 12.5, 15.0 × 15.0, 17.5 × 17.5, 20.0 × 20.0, 22.5 × 22.5, 25.0 × 25.0, 27.5 × 27.5, 30.0 × 30.0 | Cuboid and Lung |
| Dose rate [MU/min] | 6, 10, and 15 MV beam with flattening-filter (FF): 20, 60, 100, 200, 300, 400, 500, 600; 6 MV beam without FF: 400, 600, 800, 1,000, 1,200, 1,400; 10 MV beam without FF: 400, 800, 1,200, 1,600, 2,000, 2,400 | Cuboid |
| Gantry angle [°] | 0, 15, 30, 45, 60, 75, 90, 105, 120, 135, 150, 165, 180, 195, 210, 225, 240, 255, 270, 285, 300, 315, 330, 345 | Cuboid and Cylindrical |
| Collimator anglea [°] | 0, 15, 30, 45, 60, 75, 90, 105, 120, 135, 150, 165, 175, 185, 195, 210, 225, 240, 255, 270, 285, 300, 315, 330, 345 | Cuboid |
| FPD position from the isocenter [cm] | 40, 50, 60, 70, 80 | Cuboid |
MV, megavoltage.
Due to mechanical limitations, the movement range of the collimator angle ranged from 185°/195° to 195°/185° counterclockwise/clockwise.
Figure 1.Images of phantoms used in this study: (left) 3D-printed thoracic phantom (Lung phantom), (middle) water-equivalent cuboid phantom (Cuboid phantom), and (right) water-equivalent cylindrical phantom (Cylindrical phantom). The length of the rulers in the middle of the top row is 30 cm. 3D, three-dimensional.
Figure 2.Schematic of MV-scatter correction procedures. MV, megavoltage.
Figure 3.(a) Dark-field image, (b) second image, (c) subtracted image, and (d) averaged image (MV-scatter map) under reference condition of 6 MV photon beam with flattening-filter (6X-FF). The window levels and widths for the pixel value were 1500 and 3000 for (a, b), and 350 and 700 for (c, d), respectively. (e) The pixel value profiles are indicated in the white dashed rectangles in (a–c). The MV-scatter maps of the reference condition acquired by 6X-FF, 10X-FF, 15X-FF, 6 MV photon beam without flattening-filter (6X-FFF), and 10X-FFF are presented in the center. The window level and width for the pixel value were 350 and 700, respectively. FFF, flattening-filter-free; MV, megavoltage.
Figure 4.Transverse pixel value profiles of MV-scatter maps acquired with MV FS of (a) 5.0 × 5.0, (c) 10.0 × 10.0, and (e) 30.0 × 30.0 cm2. Longitudinal pixel value profiles of MV-scatter maps acquired with MV FS of (b) 5.0 × 5.0, (d) 10.0 × 10.0, and (f) 30.0 × 30.0 cm2. FS, field size; MV, megavoltage.
Figure 5.FS dependencies for (a) Cuboid and (c) Lung phantoms with various MV photon beam energies. Results obtained with square FSs ranging from 2.5 × 2.5 to 15.0 × 15.0 cm2 for (b) Cuboid phantom and (d) Lung phantom. The intersections of the two dashed lines in (a–d) indicate the MV-scatter value obtained under the reference condition of 6 MV photon beam with flattening-filter (6X-FF). FS, field size; FF, flattening-filter.
Figure 6.Dose rate dependencies for (a) 6, 10, and 15 MV photon beam with flattening-filter (6X-FF, 10X-FF, and 15X-FF) and (b) 6 and 10 MV photon beam without flattening-filter (6X-FFF and 10X-FFF). The intersections of the two dashed lines in (a, b) indicate the MV-scatter value obtained under the reference condition of 6X-FF. FF, flattening-filter; MV, megavoltage.
Figure 7.Gantry angle dependencies of (a) Cuboid phantom and (b) Cylindrical phantom with various MV photon beam energies. The dashed line indicates the MV-scatter value obtained under the reference condition of 6 MV photon beam with flattening-filter (6X-FF). (c) Collimator angle dependencies with various MV photon beam energies. The dashed line indicates the MV-scatter value obtained under the reference condition of 6X-FF.(d) FPD position dependencies with various MV photon beam energies. The intersection of the two dashed lines indicates the MV-scatter value obtained under the reference condition of 6X-FF. FF, flattening-filter; FPD, flat panel detector; MV, megavoltage.
MV-scatter database based on the Cuboid phantom results
| MV energy | Field size | Dose rate | Gantry angle | Collimator angle | FPD position | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| [MV] |
| [cm2] |
| [MU/min] |
| [°] |
| [°] |
| [cm] |
|
| 6X-FF | 1.00 | 6.25 | 0.12 |
| 0.0025 × | 0 | 1.00 | 0 | 1.00 | 40 | 2.83 |
FPD, flat-panel detector; MV, megavoltage.
Figure 8.kV images acquired during MV beam irradiation (MV+kV) with FS of 10.0 × 10.0 cm2, where MV-scatter corrected images (MVScorr) and kV image without MV beam irradiation are shown. The Individual: QUASAR = MV+kV images were corrected by the MV-scatter map acquired using QUASAR at each gantry angle.The Estimation: Ref-QUASAR = MV-scatter maps were estimated from the database and MV-scatter image of QUASAR phantom acquired by the reference condition (FS: 10.0 × 10.0 cm2, dose rate: 400 MU/min, gantry and collimator angles: 0°, and FPD position: 70 cm from isocenter) with 6 MV photon beam with flattening-filter (6X-FF). The Estimation: Ref-Cuboid = MV-scatter maps were estimated from the database and MV-scatter image of the Cuboid phantom acquired by the reference condition of 6X-FF.The window level and width were 400 and 800, respectively. FF, flattening-filter; FPD, flat panel detector; MV, megavoltage.
Figure 9.Boxplots of relative errors of image contrasts to reference image for kV image acquired during MV beam irradiation (MV+kV) and MV-scatter corrected (MVScorr) image for (a) FF and (b) FF-free (FFF) beams with FS of 5.0 × 5.0 cm2, and for (c) FF and (d) FFF beams with FS of 10.0 × 10.0 cm2. The Individual: QUASAR = MV+kV images were corrected by the MV-scatter map acquired using QUASAR at each gantry angle. The Estimation: Ref-QUASAR = MV-scatter maps were estimated from the database and MV-scatter image of the QUASAR phantom acquired by the reference condition (FS: 10.0 × 10.0 cm2, dose rate: 400 MU/min, gantry and collimator angles: 0°, and FPD position: 70 cm from isocenter) with 6 MV photon beam with flattening-filter (6X-FF). The Estimation: Ref-Cuboid = MV-scatter maps were estimated from the database and MV-scatter image of the Cuboid phantom acquired by the reference condition with 6X-FF. FF, flattening-filter; FPD, flat panel detector; FS, field size; MV, megavoltage.