Jarmo Teuho1,2, Leon Riehakainen3, Aake Honkaniemi4, Olli Moisio3, Chunlei Han4, Marko Tirri3,5, Shihao Liu6, Tove J Grönroos3,7, Jie Liu8, Lin Wan9, Xiao Liang8, Yiqing Ling8, Yuexuan Hua8, Anne Roivainen3,4,10, Juhani Knuuti3,4, Qingguo Xie8,11,12, Mika Teräs5,13, Nicola D'Ascenzo8,11, Riku Klén3.
Abstract
BACKGROUND: We investigated the image quality of 11C, 68Ga, 18F and 89Zr, which have different positron fractions, physical half-lifes and positron ranges. Three small animal positron emission tomography/computed tomography (PET/CT) systems were used in the evaluation, including the Siemens Inveon, RAYCAN X5 and Molecubes β-cube. The evaluation was performed on a single scanner level using the national electrical manufacturers association (NEMA) image quality phantom and analysis protocol. Acquisitions were performed with the standard NEMA protocol for 18F and using a radionuclide-specific acquisition time for 11C, 68Ga and 89Zr. Images were assessed using percent recovery coefficient (%RC), percentage standard deviation (%STD), image uniformity (%SD), spill-over ratio (SOR) and evaluation of image quantification.Entities:
Keywords: Image quality; NEMA; PET/CT; Positron emission tomography; Preclinical imaging; Small-animal
Year: 2020 PMID: 33301074 PMCID: PMC7728905 DOI: 10.1186/s13550-020-00724-z
Source DB: PubMed Journal: EJNMMI Res ISSN: 2191-219X Impact factor: 3.138
Technical characteristics of the preclinical PET/CT systems involved in the evaluation [15–17]
| System | Detector crystal | Detector size (mm) | Detector rings (N) | Ring diameter (cm) | Transaxial FOV (mm) | Axial FOV (mm) | Energy resolution (%) | Timing resolution (ns) | NEMA resolution (mm) | NEMA sensitivity (%) |
|---|---|---|---|---|---|---|---|---|---|---|
| RAYCAN | LYSO | 1.89 × 1.89 × 13 | 26 | 19.2 | 130 | 53 | 15 | 5 | 1.9 | 1.7 |
| Inveon | LSO | 1.51 × 1.51 × 10 | 80 | 16.1 | 100 | 127 | 15 | 3.43 | 1.8 | 9.3 |
| Molecubes | LYSO | 25.4 × 25.4 × 8* | 5 | 7.6 | 72 | 130 | 12 | 2.4 | 1.1 | 12.4 |
The resolution values are reported based on the filtered backprojection (FBP) reconstruction for all systems
*Monolithic detector
Physical properties of different radionuclides used for PET image quality evaluation [8]
| Radionuclide | Physical half-life | Branching ratio | Max β+ energy (MeV) | Maximum β+ range in water (mm) | Prompt gamma branching ratio and energy |
|---|---|---|---|---|---|
| 11C | 20.4 min | 0.998 | 0.960 | 4.2 | – |
| 68Ga | 67.8 min | 0.877 | 1.899 | 9.2 | 0.03, 1.08 meV |
| 18F | 109.8 min | 0.969 | 0.634 | 2.4 | – |
| 89Zr | 3.27 d | 0.227 | 0.902 | 3.8 | 0.99, 909.2 keV |
Number of events recorded on each PET/CT system and the final acquisition times used in the evaluation of different nuclei
| PET/CT system | The number of events (prompts + delays) for each isotope | |||
|---|---|---|---|---|
| [18F]FDG | [11C]acetate | [68Ga]chloride | [89Zr]oxalate | |
| RAYCAN | 37.96 M | 41.07 M | 38.37 M | 39.59 M |
| Inveon | 185.30 M | 190.60 M | 192.32 M | 178.04 M |
| Molecubes | 382.43 M | 407.83 M | 394.38 M | 480.08 M |
| Acquisition times | 20 min | 31 min 41 s | 25 min 32 s | 72 min |
Doses and activity concentrations at scan start times with acquisition times of the radiotracers on different PET/CT systems
| PET/CT system | Dose (MBq)/activity concentration (kBq/mL) at scan start time for each radionuclide | |||
|---|---|---|---|---|
| [11C]acetate | [68Ga]chloride | [18F]FDG | [89Zr]oxalate | |
| RAYCAN | 3.69/167.73 | 3.72/169.55 | 3.70/168.18 | 3.68/167.27 |
| Inveon | 3.40*/154.55 | 3.73/169.55 | 3.76/170.91 | 3.68/167.27 |
| Molecubes | 3.69/167.73 | 3.72/169.09 | 3.70/168.18 | 3.68/167.27 |
*Activity outside the specified limits of NEMA (3.70 MBq ± 5%)
Image reconstruction parameters and data corrections implemented on the PET/CT systems
| System | Reconstruction algorithm | Matrix size | Pixel size | Iterations/ Subsets | Post-filter | β-value/prior penalty | Data corrections | PSF applied |
|---|---|---|---|---|---|---|---|---|
| RAYCAN | 3D-OSEM-PSF | 280 × 280 × 100 | 0.5 × 0.5 × 0.5 | 2/12 | Low | – | AC | Yes |
| Inveon | SP-MAP | 128 × 128 × 159 | 0.8 × 0.8 × 0.8 | 18 /16 | – | 0.00427838 | AC, SC, RC | – |
| Molecubes | 3D-OSEM | 192 × 192 × 384 | 0.4 × 0.4 × 0.4 | 30/7 | – | – | AC, SC, RC | – |
AC attenuation correction, SC Scatter correction, RC randoms correction
Fig. 1Image quality from one plane in the rod region of the NEMA phantom with different preclinical PET/CT systems. The systems shown are a RAYCAN, b Inveon and c Molecubes. Slight differences in the noise level between various radionuclides in the images within a system can be seen. The 18F and 89Zr have similar whereas appearance, whereas the positron range is the major contributing factor to the blurred appearance of the images with 68Ga. All images have been scaled between 0 and 1.25 times the mean activity of the uniform compartment
Fig. 2Recovery coefficients (%RC) with different radionuclides with the a RAYCAN, b Inveon and c Molecubes systems. The maximal recovery is achieved with 18F. 11C and 89Zr are close to the values with 18F. 68Ga shows the lowest recovery on all systems and rod sizes
Fig. 3Percentage standard deviation with different rod sizes and radionuclides with the a RAYCAN, b Inveon and c Molecubes systems. Most variability in the is seen with 68Ga with rod size of 1 mm on the Molecubes system and with rod size of 1 mm and 2 mm on the Inveon system, reflecting system resolution. The RAYCAN system has more variability between radionuclides, where 68Ga shows the highest variability on nearly all rod sizes
The mean activity (kBq/mL) and the percentage standard deviation (%SD) measured from the uniform compartment of the phantom
| PET/CT system | 11C | 11C | 68Ga | 68Ga | 18F | 18F | 89Zr | 89Zr |
|---|---|---|---|---|---|---|---|---|
| RAYCAN | 190.13 | 5.81 | 168.90 | 6.91 | 182.57 | 6.03 | 196.61 | 6.24 |
| Inveon | 171.57 | 5.99 | 170.16 | 7.05 | 185.50 | 4.85 | 205.30 | 5.69 |
| Molecubes | 186.67 | 8.46 | 143.94 | 7.16 | 174.46 | 7.39 | 154.20 | 7.34 |
Percentage difference (%Δ) of the mean activity in the uniform compartment of the phantom compared to the reference activity measured at scan start time
| PET/CT system | 11C | 68Ga | 18F | 89Zr |
|---|---|---|---|---|
| RAYCAN | 13.36 | − 0.38 | 8.56 | 17.54 |
| Inveon | 11.01 | 0.36 | 8.54 | 22.73 |
| Molecubes | 11.29 | − 14.87 | 3.73 | − 7.82 |
Spill-over ratios (SOR) with the percentage standard deviation () in parenthesis measured from the phantom water and air compartment
| SOR ( | 11C | 68Ga | 18F | 89Zr |
|---|---|---|---|---|
| RAYCAN* | ||||
| Water | 0.27 (6.83) | 0.32 (8.75) | 0.26 (8.28) | 0.25 (8.74) |
| Air | 0.16 (8.54) | 0.25 (10.42) | 0.15 (8.97) | 0.16 (7.64) |
| Inveon | ||||
| Water | 0.16 (10.10) | 0.09 (34.01) | 0.05 (27.75) | < 0.01 (240.08) |
| Air | 0.01 (46.04) | 0.06 (55.52) | 0.03 (29.03) | 0.01 (106.38) |
| Molecubes | ||||
| Water | 0.17 (19.50) | 0.12 (13.78) | 0.07 (15.85) | 0.07 (16.71) |
| Air | 0.13 (13.04) | 0.09 (13.78) | 0.07 (12.14) | 0.06 (13.61) |
*System has no data corrections for randoms and scatter
Comparison of %RC and %SD values reported from previous multi-radionuclide studies
| System | Radionuclide | Activity (MBq) | Acquisition time (min) | Algorithm | Iterations/updates | %RC | %RC | %RC | %RC | %RC | %SD | References |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ALBIRA II | 18F | 8 | 15 | MLEM | 25 | N/A | 0.31 | 0.60 | 0.76 | 0.79 | 6.4 | [ |
| 68Ga | 8 | 15 | MLEM | 25 | N/A | 0.10 | 0.30 | 0.42 | 0.56 | 6.7 | ||
| ARGUS | 18F | 3.7 | 20 | FORE-2D OSEM | 48 | 0.28 | 0.8a | 0.9a | 1.0a | 0.92 | 15.1 | [ |
| 68Ga | 3.7 | 20.88 | FORE-2D OSEM | 48 | 0.17 | 0.3a | 0.4a | 0.5a | 0.72 | 17.0 | ||
| Mediso nanoScan | 18F | 3.7 | 30 | Tera-Tomo 3D | 52 | 0.18 | 0.46 | 0.68 | 0.79 | 0.87 | 5.0a | [ |
| 68Ga | 4.19 | 30 | Tera-Tomo 3D | 52 | 0.09 | 0.20 | 0.36 | 0.48 | 0.60 | 5.0a | ||
| Inveon | 18F | 3.7 | 20 | 3D-OSEM MAP | 2/18 | 0.2a | 0.4a | 0.7a | 0.8a | 1.0a | 2a | [ |
| 68Ga | 3.7 | 22.88 | 3D-OSEM MAP | 2/18 | 0.1a | 0.2a | 0.4a | 0.7a | 0.6a | 3a | ||
| 89Zr | 3.7 | 80.4 | 3D-OSEM MAP | 2/18 | 0.2a | 0.4a | 0.6a | 0.8a | 0.9a | 2a | ||
| Molecules | 18F | 4–8 | 20 | 3D-OSEM | 100 | 0.3 | 0.79 | 0.85 | 1.01 | 0.97 | N/A | [ |
| 11C | 4–8 | 30 | 3D-OSEM | 100 | 0.52 | 0.56 | 0.86 | 0.93 | 1.02 | N/A | ||
| 68Ga | 4–8 | 20 | 3D-OSEM | 100 | 0.11 | 0.35 | 0.46 | 0.56 | 0.57 | N/A |
MLEM maximum-likelihood expectation–maximization, FORE-2D OSEM Fourier rebinning followed by 2D ordered-subsets expectation-maximization, 3D-OSEM MAP 3D ordered-subsets expectation-maximization in combination with maximum a posteriori, 3D-OSEM 3D ordered-subsets expectation-maximization
aValues have been derived from the reference Figures and are approximate, N/A = the value is not reported in the reference