| Literature DB >> 33299436 |
Somayyeh Babaloui1,2, Shakardokht Jafari2,3, Wojciech Polak3, Mahdi Ghorbani4, Michael Wj Hubbard2, Annika Lohstroh2,5, Alireza Shirazi1, Ramin Jaberi2,6.
Abstract
PURPOSE: The aim of this study was to construct a low-cost, anthropomorphic, and 3D-printed pelvis phantom and evaluate the feasibility of its use to perform 3D dosimetry with commercially available bead thermoluminescent dosimeters (TLDs).Entities:
Keywords: 3D dosimetry; anthropomorphic phantom; radiotherapy; silica bead TLDs
Year: 2020 PMID: 33299436 PMCID: PMC7701919 DOI: 10.5114/jcb.2020.100380
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1A) Syringe with a 2.5 mm spacing arrangement of silica bead TLDs, B) The syringe located on the two legs into the water tank, C) The positioning of the water tank for CT scanning
Fig. 2A) 3D model view of the organs, B) 3D-printed organs with placement of thermoluminescent dosimeters (TLDs) at their surface, C) The anthropomorphic pelvis phantom in the water tank on the CT coach, D) The phantom in the water tank connected to high-dose-rate (HDR) unit
3D-printing parameters to construct the pelvis phantom organs
| Setting | Value |
|---|---|
| Bed temperature | 40°C |
| Extruder temperature | 160°C |
| Layer height | 0.2 mm |
| First layer height | 0.2 mm |
| Vertical shells | N/A |
| Horizontal shells | N/A |
| Speed of print movement (perimeter) | 80.0 mm/s |
| Speed of print movement (external perimeter) | 50% |
| Speed of print movement (infill) | 80% |
| First layer print movement speed | 80.0 mm/s |
| Fill density | 90% |
| Fill pattern | Solid |
| Fill angle | 0 |
| Nozzle diameter | 0.40 mm |
| Filament diameter | 1.5 mm |
Fig. 3High-dose-rate (HDR) plan on the CT image of the phantom (coronal and axial view); each thermoluminescent dosimeter (TLD) is clearly visible
Result of the organs’ dose obtained at different silica bead resolution on the syringe according to DVH (the doses are in cGy)
| Organs (mean dose) | No bead | Hresa(2.5 mm) | Mresb(5 mm) | Lresc (7.5 mm) |
|---|---|---|---|---|
| Rectum | 3449.59 | 3445.89 | 3448.44 | 3449.36 |
| Rectal wall | 3639.23 | 3634.43 | 3638.47 | 3638.79 |
| Bladder | 3578.50 | 3576.96 | 3578.53 | 3578.50 |
| CTV | 4522.45 | 4519.65 | 4522.21 | 4522.44 |
| PTV | 4491.17 | 4488.50 | 4490.79 | 4491.14 |
high resolution, bmedium resolution, clow resolution
Result of percentage dose differences of the organs at different silica bead resolution on the syringe according to DVH
| (Hresavs. no bead) | (Mresbvs. no bead) | (Lresc vs. no bead) | |
|---|---|---|---|
| Rectum | 0.11 | 0.03 | 0.00 |
| Rectal wall | 0.13 | 0.02 | 0.01 |
| Bladder | 0.04 | 0.00 | 0.00 |
| CTV | 0.06 | 0.01 | 0.00 |
| PTV | 0.06 | 0.01 | 0.00 |
high resolution, bmedium resolution, clow resolution
Result of the comparing planes with and without beads at different source to plane distance (SPDs) and different bead resolutions on the syringe with Verisoft software (3%, 3 mm criteria)
| γindex (mean) | % difference (mean) | |
|---|---|---|
| SPD = 102 cm | ||
| High resolution (2.5 mm) | 0.071 | 0.246 |
| Medium resolution (5.0 mm) | 0.008 | 0.026 |
| Low resolution (7.5 mm) | 0.001 | 0.004 |
| SPD = 100 cm | ||
| High resolution (2.5 mm) | 0.080 | 0.299 |
| Medium resolution (5.0 mm) | 0.014 | 0.053 |
| Low resolution (7.5 mm) | 0.003 | 0.009 |
| SPD = 99 cm | ||
| High resolution (2.5 mm) | 0.064 | 0.952 |
| Medium resolution (5.0 mm) | 0.007 | 0.676 |
| Low resolution (7.5 mm) | 0.003 | 0.659 |
Result of the comparing planes with and without beads at different source to plane distance (SPDs) and different bead resolutions n the syringe with Verisoft software (2%, 2 mm criteria)
| γindex (mean) | % difference (mean) | ||
|---|---|---|---|
| SPD = 102 cm | |||
| High resolution (2.5 mm) | 0.107 | 0.246 | |
| Medium resolution (5.0 mm) | 0.011 | 0.026 | |
| Low resolution (7.5 mm) | 0.002 | 0.004 | |
| SPD = 100 cm | |||
| High resolution (2.5 mm) | 0.120 | 0.299 | |
| Medium resolution (5.0 mm) | 0.021 | 0.053 | |
| Low resolution (7.5 mm) | 0.004 | 0.009 | |
| SPD = 99 cm | |||
| High resolution (2.5 mm) | 0.096 | 0.952 | |
| Medium resolution (5.0 mm) | 0.010 | 0.676 | |
| Low resolution (7.5 mm) | 0.004 | 0.659 | |
Fig. 4Dose surface histograms obtained by thermoluminescent dosimeters (TLDs) readout and point doses calculation of treatment planning system (TPS)
Print statistics for each individual internal organ in the anthropomorphic pelvis phantom
| Organ | Print time (h) | Mass (g) | Print cost (GBP) |
|---|---|---|---|
| Bladder | 5.56 | 42.3 | 40 |
| Uterus | 5.42 | 85.2 | 40 |
| Rectum and sigmoid | 7.07 | 27.1 | 40 |
| Total | 18.05 | 154.6 | 120 |