Literature DB >> 23822434

In vivo measurements for high dose rate brachytherapy with optically stimulated luminescent dosimeters.

Renu Sharma1, Paul A Jursinic.   

Abstract

PURPOSE: To show the feasibility of clinical implementation of OSLDs for high dose-rate (HDR) in vivo dosimetry for gynecological and breast patients. To discuss how the OSLDs were characterized for an Ir-192 source, taking into account low gamma energy and high dose gradients. To describe differences caused by the dose calculation formalism of treatment planning systems.
METHODS: OSLD irradiations were made using the GammaMedplus iX Ir-192 HDR, Varian Medical Systems, Milpitas, CA. BrachyVision versions 8.9 and 10.0, Varian Medical Systems, Milpitas, CA, were used for calculations. Version 8.9 used the TG-43 algorithm and version 10.0 used the Acuros algorithm. The OSLDs (InLight Nanodots) were characterized for Ir-192. Various phantoms were created to assess calculated and measured doses and the angular dependence and self-absorption of the Nanodots. Following successful phantom measurements, patient measurements for gynecological patients and breast cancer patients were made and compared to calculated doses.
RESULTS: The OSLD sensitivity to Ir-192 compared to 6 MV is between 1.10 and 1.25, is unique to each detector, and changes with accumulated dose. The measured doses were compared to those predicted by the treatment planning system and found to be in agreement for the gynecological patients to within measurement uncertainty. The range of differences between the measured and Acuros calculated doses was -10%-14%. For the breast patients, there was a discrepancy of -4.4% to +6.5% between the measured and calculated doses at the skin surface when the Acuros algorithm was used. These differences were within experimental uncertainty due to (random) error in the location of the detector with respect to the treatment catheter.
CONCLUSIONS: OSLDs can be successfully used for HDR in vivo dosimetry. However, for the measurements to be meaningful one must account for the angular dependence, volume-averaging, and the greater sensitivity to Ir-192 gamma rays than to 6 MV x-rays if 6 MV x-rays were used for OSLD calibration. The limitations of the treatment planning algorithm must be understood, especially for surface dose measurements. Use of in vivo dosimetry for HDR brachytherapy treatments is feasible and has the potential to detect and prevent gross errors. In vivo HDR brachytherapy should be included as part of the QA for a HDR brachytherapy program.

Entities:  

Mesh:

Year:  2013        PMID: 23822434     DOI: 10.1118/1.4811143

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Development of a novel and low-cost anthropomorphic pelvis phantom for 3D dosimetry in radiotherapy.

Authors:  Somayyeh Babaloui; Shakardokht Jafari; Wojciech Polak; Mahdi Ghorbani; Michael Wj Hubbard; Annika Lohstroh; Alireza Shirazi; Ramin Jaberi
Journal:  J Contemp Brachytherapy       Date:  2020-10-30

2.  In vivo dosimetry and acute toxicity in breast cancer patients undergoing intraoperative radiotherapy as boost.

Authors:  Jason Joon Bock Lee; Jinhyun Choi; Sung Gwe Ahn; Joon Jeong; Ik Jae Lee; Kwangwoo Park; Kangpyo Kim; Jun Won Kim
Journal:  Radiat Oncol J       Date:  2017-06-30

Review 3.  In vivo dosimetry in brachytherapy: Requirements and future directions for research, development, and clinical practice.

Authors:  Gabriel P Fonseca; Jacob G Johansen; Ryan L Smith; Luc Beaulieu; Sam Beddar; Gustavo Kertzscher; Frank Verhaegen; Kari Tanderup
Journal:  Phys Imaging Radiat Oncol       Date:  2020-09-28

4.  3D in vivo dosimetry of HDR gynecological brachytherapy using micro silica bead TLDs.

Authors:  Ramin Jaberi; Somayyeh Babaloui; Zahra Siavashpour; Maryam Moshtaghi; Alireza Shirazi; Musa Joya; Mohammad Hadi Gholami; Shakardokht Jafari
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.