Quantitative assessment of beam perturbations caused by silicon diodes used for in vivo dosimetry.

A Sen1, E I Parsai, S W McNeeley, K M Ayyangar.   

Abstract

PURPOSE: To measure the effect of silicon diode detectors used for in vivo dosimetry on beam characteristics and determine whether this effect is clinically significant. METHODS AND MATERIALS: Commercially available photon and electron diodes were placed on the central axis of photon and electron beams. The beam characteristics were measured for 6- and 10-MV photon and 6-20-MeV electron energies from a Varian Clinac 1800 medical linear accelerator. Water was used for the medium, and measurements were made for various clinically common field sizes and depths.
RESULTS: Beam attenuations along the central axis were 10 and 7.5% for 6- and 10-MV photons, respectively. Electron beam dose reductions were between 13 and 25% for 20-6-MeV electrons. Photon beam flatness varied up to 7% at different depths, but the symmetry was not affected much. Electron beam flatness and symmetry were significantly changed to as much as 18 and 6%, respectively.
CONCLUSION: Use of diode detectors on central axis of photon and electron beams for in vivo dosimetry causes significant attenuation and alteration of the beam characteristics. The percentage of the volume affected is significant (e.g., 23% of the volume in a 4 x 4 field gets 10% less dose for a 6-MV photon beam), especially if these diodes are used for in vivo dosimetry on the central axis every day for every treatment, as is done in some clinics. Other beam parameters such as penumbra and skin dose are also affected. It is therefore recommended that the diodes be used only as needed.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8823277     DOI: 10.1016/s0360-3016(96)00279-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  3 in total

1.  Development of a novel and low-cost anthropomorphic pelvis phantom for 3D dosimetry in radiotherapy.

Authors:  Somayyeh Babaloui; Shakardokht Jafari; Wojciech Polak; Mahdi Ghorbani; Michael Wj Hubbard; Annika Lohstroh; Alireza Shirazi; Ramin Jaberi
Journal:  J Contemp Brachytherapy       Date:  2020-10-30

2.  In vivo dosimetry using a single diode for megavoltage photon beam radiotherapy: implementation and response characterization.

Authors:  V C Colussi; A S Beddar; T J Kinsella; C H Sibata
Journal:  J Appl Clin Med Phys       Date:  2001       Impact factor: 2.102

3.  Dose Shadowing and Prosthesis Involvement for Megavoltage Photon In vivo Diode Dosimetry.

Authors:  Nicholas Ade; Dete Van Eeden; F C P Du Plessis
Journal:  J Med Phys       Date:  2019-12-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.