Evaluation and clinical implementation of in vivo dosimetry for kV radiotherapy using radiochromic film and micro-silica bead thermoluminescent detectors.

Antony L Palmer1, Shakardokht M Jafari2, Ioanna Mone3, Sarah Muscat4.   

Abstract

PURPOSE: kV radiotherapy treatment calculations are based on flat, homogenous, full-scatter reference conditions. However, clinical treatments often include surface irregularities and inhomogeneities, causing uncertainty. Therefore, confirmation of actual delivered doses in vivo is valuable. The current study evaluates, and implements, radiochromic film and micro silica bead TLD for in vivo kV dosimetry.
METHODS: The kV energy and dose response of EBT3 film and silica bead TLD was established and uncertainty budgets determined. In vivo dosimetry measurements were made for a consecutive series of 30 patients using the two dosimetry systems.
RESULTS: Energy dependent calibration factors were required for both dosimetry systems. The standard uncertainty estimate for in vivo measurement with film was 1.7% and for beads was 1.5%. The mean measured dose was -2.1% for film and -2.6% for beads compared to prescription. Deviations up to -9% were found in cases of large surface irregularity, or with underlying air cavities or bone. Dose shielding by beads could be clinically relevant at low kV energies and superficial depths.
CONCLUSIONS: Both film and beads may be used to provide in vivo verification of delivered doses in kV radiotherapy, particularly for complex situations that are not well represented by standard reference condition calculations.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Film; In vivo; Silica bead; kV

Mesh:

Year:  2017        PMID: 29173920     DOI: 10.1016/j.ejmp.2017.08.009

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Development of a novel and low-cost anthropomorphic pelvis phantom for 3D dosimetry in radiotherapy.

Authors:  Somayyeh Babaloui; Shakardokht Jafari; Wojciech Polak; Mahdi Ghorbani; Michael Wj Hubbard; Annika Lohstroh; Alireza Shirazi; Ramin Jaberi
Journal:  J Contemp Brachytherapy       Date:  2020-10-30

2.  Off-axis dose distribution with stand-in and stand-off configurations for superficial radiotherapy treatments.

Authors:  Keith Davey; Margaret Moore; Sinéad Cleary; Christoph Kleefeld; Mark J Foley
Journal:  J Appl Clin Med Phys       Date:  2019-10-12       Impact factor: 2.102

3.  3D in vivo dosimetry of HDR gynecological brachytherapy using micro silica bead TLDs.

Authors:  Ramin Jaberi; Somayyeh Babaloui; Zahra Siavashpour; Maryam Moshtaghi; Alireza Shirazi; Musa Joya; Mohammad Hadi Gholami; Shakardokht Jafari
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.