| Literature DB >> 33240520 |
Melissa Nataatmadja1,2,3, Angela W S Fung4, Beryl Jacobson5, Jack Ferera6, Eva Bernstein6, Paul Komenda7,8, Andre Mattman4, David Seccombe5, Adeera Levin3,6,9.
Abstract
BACKGROUND: The StatSensor is a point-of-care device which measures creatinine in capillary whole blood. Previous studies reported an underestimation of the creatinine measurements at high creatinine concentrations and were performed in the prestandardization era for creatinine.Entities:
Keywords: CKD screening; CKD staging; chronic kidney disease; fingerprick test; point-of-care testing
Year: 2020 PMID: 33240520 PMCID: PMC7672720 DOI: 10.1177/2054358120970716
Source DB: PubMed Journal: Can J Kidney Health Dis ISSN: 2054-3581
Study Participant Demographics.
| Number of patients | 60 |
|---|---|
| Age, years, mean ± SD (range) | 61.9 ± 15.0 (27-94) |
| Sex | |
| Male, No. (%) | 33 (55.0) |
| Female, No. (%) | 27 (45.0) |
| Weight, kg, mean ± SD (range) | 79.3 ± 18.2 (45.2-124.8) |
| Body mass index, kg/m2, mean ± SD (range) | 28.0 ± 5.9 (16.8-51.3) |
| Kidney | |
| Native, No. (%) | 40 (66.7) |
| Transplant, No. (%) | 20 (33.3) |
| Ethnicity | |
| Nonblack, No. (%) | 57 (95.0) |
| Black, No. (%) | 3 (5.0) |
| Plasma creatinine, µmol/L, mean ± SD (range) | 137 ± 59 (50-357) |
| eGFR (plasma), mL/min/1.73 m2, mean ± SD (range) | 50 ± 20 (15-112) |
| eGFR, No. (%) | |
| ≥90 mL/min/1.73 m2 | 2 (3.3) |
| 60-89 mL/min/1.73 m2 | 15 (25.0) |
| 30-59 mL/min/1.73 m2 | 35 (58.3) |
| 15-29 mL/min/1.73 m2 | 8 (13.3) |
| <15 mL/min/1.73 m2 | 0 (0) |
Note. eGFR = estimated glomerular filtration rate.
Creatinine and eGFR in Serum, Plasma, and Capillary Whole Blood (Direct Fingerstick and SAFE-T-FILL).
| Vendor calibration | Postanalytical bias correction | |
|---|---|---|
| Serum creatinine, µmol/L | 142 ± 61 | 138 ± 59 |
| Plasma creatinine, µmol/L | 141 ± 61 | 137 ± 59 |
| eGFR[ | NC | 50 ± 20 |
| Capillary (fingerstick) creatinine, µmol/L | 153 ± 73 | 137 ± 62 |
| eGFR[ | 45 ± 18 | 50 ± 20 |
| Capillary (SAFE-T-FILL) creatinine, µmol/L | 163 ± 90 | 136 ± 65 |
| eGFR[ | 43 ± 19 | 50 ± 19 |
Note. eGFR = estimated glomerular filtration rate; NC = not calculated; CKD-EPI = Chronic Kidney Disease Epidemiology Collaboration.
eGFR was calculated from CKD-EPI equation.[13]
Figure 1.Comparison of direct fingerstick StatSensor creatinine and Roche plasma enzymatic creatinine (ID-GC-MS aligned). (A) Deming Regression Plot. Shaded area represents 95% CI. StatSensor Cr = 1.129 (95% CI = 0.931-1.293), plasma enzymatic Cr = −6.188 (95% CI = −26.498 to 18.25), R2 = 0.883, n = 60. Solid line, linear best fit determined by 1/x-weighted Deming regression; dashed line, line of identity y = x and (B) Bland-Altman difference plot. Shaded area represents 95% CI. Solid line, mean bias = 12.7 (95% CI = −38.4 to 64.3); dashed line, no difference.
Note. ID-GC-MS = isotope dilution gas chromatography mass spectrometry; CI = confidence interval.
Figure 2.Comparison of SAFE-T-FILL StatSensor creatinine and Roche plasma enzymatic creatinine. (A) Deming Regression Plot. Shaded area represents 95% CI. StatSensor Cr = 1.385 (95% CI = 1.121-1.679), plasma enzymatic Cr = −25.483 (95% CI = −62.941 to 6.149), R2 = 0.847, n = 55. Solid line, linear best fit determined by 1/x-weighted Deming regression; dashed line, line of identity y = x and (B) Bland-Altman difference plot. Shaded area represents 95% CI. Solid line, mean bias = 29.6 µmol/L (95% CI = −26.0 to 128.8); dashed line, no difference.
Note. CI = confidence interval.
Figure 3.Clinical concordance of eGFR calculated by the CKD-EPI equation between capillary whole blood StatSensor creatinine in singlet compared with plasma enzymatic creatinine. (A) Direct fingerstick and (B) SAFE-T-FILL capillary collection device.
Note. Uncorrected creatinine value (filled circle), bias-corrected creatinine value (unfilled circle), eGFR calculated using StatSensor creatinine value compared with eGFR calculated using reference creatinine value results in same Kidney Disease: Improving Global Outcomes CKD stage (green), discrepancy by 1 stage (yellow), discrepancy by 2 stages (red). CI = confidence interval; eGFR = estimated glomerular filtration rate; CKD-EPI = Chronic Kidney Disease Epidemiology Collaboration.
Clinical Performance for Nova StatSensor Using an eGFR Decision Threshold of 60 mL/min/1.73 m2.
| Calibration | Sensitivity, % | Specificity, % | Positive predictive value, % | Negative predictive value, % | |
|---|---|---|---|---|---|
| StatSensor Direct Fingerstick | Vendor calibration | 86.1 | 82.4 | 92.5 | 70.0 |
| Postanalytical bias correction | 88.4 | 76.5 | 90.5 | 72.2 | |
| StatSensor SAFE-T-FILL | Vendor calibration | 89.7 | 50.0 | 70.9 | 66.7 |
| Postanalytical bias correction | 86.8 | 68.8 | 86.8 | 68.8 |
Note. Positive test when plasma creatinine determined eGFR < 60 mL/min/1.73 m2. Negative test when plasma creatinine determined eGFR ≥ 60 mL/min/1.73 m2. eGFR = estimated glomerular filtration rate; CI = confidence interval.
Effect of Recalibration on Interpretation of Patient Creatinine and eGFR Results.
| Patient | Calibration | Plasma enzymatic | StatSensor direct fingerstick | StatSensor SAFE-T-FILL | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Mean Cr, µmol/L | Mean eGFR, mL/min/1.73 m2
| CKD stage | Mean Cr, µmol/L | Mean eGFR, mL/min/1.73 m2
| CKD stage | Mean Cr, µmol/L | Mean eGFR, mL/min/1.73 m2
| CKD stage | ||
| Patient A | Vendor calibration | 193 | 30 | III b | 152 | 40 | III b | 201 | 29 | IV |
| Postanalytical bias correction | — | — | — | 135 | 46 | III a | 163 | 37 | III b | |
| Patient B | Vendor calibration | 278 | 20 | IV | 380 | 14 | V | 423 | 12 | V |
| Postanalytical bias correction | — | — | — | 330 | 16 | IV | 323 | 16 | IV | |
| Patient C | Vendor calibration | 71 | 72 | II | 120 | 38 | III b | 119 | 39 | III b |
| Postanalytical bias correction | — | — | — | 108 | 43 | III b | 103 | 46 | III a | |
| Patient D | Vendor calibration | 131 | 38 | III b | 124 | 41 | III b | 118 | 45 | III a |
| Postanalytical bias correction | — | — | — | 112 | 46 | III a | 102 | 51 | III a | |
Note. eGFR = estimated glomerular filtration rate; CI = confidence interval; CKD = chronic kidney disease.