Cara Simone Kosack1, Wim de Kieviet2, Kubra Bayrak2, Anastacija Milovic2, Anne Laure Page3.
Abstract
Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor Xpress Creatinine (Nova Biomedical Cooperation, Waltham, MA, USA) point of care analyzer was evaluated for its diagnostic performance in indicating drug therapy change. Creatinine was measured in parallel using the Nova StatSensor Xpress Creatinine analyzer and the Vitros 5,1FS (Ortho Clinical Diagnostics, Inc, Rochester, USA), which served as reference standard. The precision (i.e., repeatability and reproducibility) and accuracy of the StatSensor Xpress Creatinine analyzer were calculated using a panel of specimens with normal, low pathological and high pathological values. Two different Nova StatSensor Xpress Creatinine analyzers were used for the assessment of accuracy using repeated measurements. The coefficient of variation of the StatSensor Xpress Creatinine analyzers ranged from 2.3 to 5.9% for repeatability and from 4.2 to 9.0% for between-run reproducibility. The concordance correlation agreement was good except for high values (>600 µmol/L). The Bland-Altman analysis in high pathological specimens suggests that the Nova StatSensor Xpress Creatinine test tends to underestimate high creatinine values (i.e., >600 µmol/L). The Nova StatSensor Xpress Creatinine analyzers showed acceptable to good results in terms of repeatability, inter-device reproducibility and between-run reproducibility over time using quality control reagents. The analyzer was found sufficiently accurate for detecting pathological values in patients (age >10 year) and can be used with a moderate risk of misclassification.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25886375 PMCID: PMC4401790 DOI: 10.1371/journal.pone.0122433
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Mean measurements in μmol/L, standard deviation and coefficient of variation of 8 repeated tests per specimen and per Nova StatSensor Xpress Creatinine analyzer.
| StatSensor 1 | StatSensor 2 | StatSensor 3 | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| mean | SD | CV % | mean | SD | CV % | mean | SD | CV % | mean CV % | ||
|
| 5.6 | 7.6 | 4.7 | 5.9 | |||||||
| Sample 1 | 62.6 | 3.96 | 6.3 | 64.1 | 4.97 | 7.8 | 60.1 | 3.04 | 5.0 | 6.4 | |
| Sample 2 | 58.8 | 2.25 | 3.8 | 59.3 | 3.80 | 6.4 | 57.9 | 2.03 | 3.5 | 4.6 | |
| Sample 3 | 55.0 | 3.62 | 6.6 | 56.4 | 4.87 | 8.6 | 57.1 | 3.14 | 5.5 | 6.9 | |
|
| 3.8 | 4.4 | 3.6 | 3.9 | |||||||
| Sample 1 | 271.1 | 6.40 | 2.4 | 275.3 | 7.32 | 2.7 | 270.8 | 8.50 | 3.1 | 2.7 | |
| Sample 2 | 169.9 | 6.70 | 3.9 | 172.8 | 7.44 | 4.3 | 170.4 | 7.82 | 4.6 | 4.3 | |
| Sample 3 | 302.5 | 15.61 | 5.2 | 296.8 | 18.61 | 6.3 | 289.9 | 8.60 | 3.0 | 4.8 | |
|
| 2.5 | 2.1 | 2.3 | 2.3 | |||||||
| Sample 1 | 695.3 | 21.50 | 3.1 | 696.1 | 21.31 | 3.1 | 702.6 | 15.30 | 2.2 | 2.8 | |
| Sample 2 | 637.6 | 12.42 | 2.0 | 640.3 | 7.53 | 1.2 | 633.6 | 13.53 | 2.1 | 1.8 | |
| Sample 3 | 372.0 | 8.58 | 2.3 | 374.3 | 7.89 | 2.1 | 369.4 | 9.66 | 2.6 | 2.3 | |
|
|
|
|
|
| |||||||
Mean measurements, standard deviation and coefficient of variation of 11 daily testing of quality control reagent per reagent and per Nova StatSensor Xpress Creatinine analyzer.
| StatSensor 1 | StatSensor 2 | StatSensor 3 | |||||||
|---|---|---|---|---|---|---|---|---|---|
| mean | SD | CV % | mean | SD | CV % | mean | SD | CV % | |
|
| 97.4 | 5.64 | 5.8 | 96.7 | 4.84 | 5.0 | 92.2 | 5.93 | 6.4 |
|
| 210.3 | 12.9 | 6.1 | 211.5 | 9.89 | 4.7 | 200.9 | 18.1 | 9.0 |
|
| 597.6 | 26.7 | 4.5 | 611.6 | 25.7 | 4.2 | 583.0 | 37.7 | 6.5 |
Results of the measurements of the 5 linearity reagents per Nova StatSensor Xpress Creatinine analyzer and corresponding correlation coefficient.
| StatSensor 1 | StatSensor 2 | StatSensor 3 | ||
|---|---|---|---|---|
|
| ||||
|
| 88 | 99 | 99 | |
|
| 229 | 214 | 179 | |
|
| 378 | 415 | 399 | |
|
| 588 | 606 | 623 | |
|
| 895 | 908 | 804 | |
|
| 0.997 | 0.997 | 0.993 | |
Fig 1Scatter plot of Nova StatSensor Xpress Creatinine analyzer vs. Vitros 5,1FS (Ortho Clinical Diagnostics, Rochester, USA) measurements.
Fig 2Bland Altman analysis of the accuracy of Nova StatSensor Xpress Creatinine analyzer vs. Vitros 5,1FS (Ortho Clinical Diagnostics, Rochester, USA) using duplicate measurements.
Summary of the Bland Altman analysis per type of specimen (normal range, low pathological, high pathological) and globally.
| (in μmol/L) | Bias (95% CI) | Lower 95% LoA | Upper 95% LoA |
|---|---|---|---|
|
| 3 (-2.6;8.6) | -21.1 | 27 |
|
| -4.3 (-16.6;8.0) | -56.7 | 48.1 |
|
| -30.9 (-67.4;5.7) | -187 | 125.3 |
|
| 4.5 (-9.9;19.0) | -46.7 | 56.7 |
|
| -10.7 (-23.5;2.0) | -109.6 | 88.2 |
|
| 0.8 (-5.2;6.7) | -43.2 | 44.8 |
Cross-tabulation of Nova StatSensor Xpress Creatinine analyzers and reference method for classification in normal or pathological values (≥115 μmol/L).
| Nova StatSensor Xpress | ||
|---|---|---|
| Pathological | Normal | |
|
| 38 | 0 |
|
| 2 | 20 |
Comparison of creatinine clearance results determined by the Nova StatSensor Xpress Creatinine analyzers or reference method modeled from the results obtained in the study applied to a hypothetical 65-kg, 30 year old male (N = 60).
| Nova StatSensor Xpress Creatinine | |||||
|---|---|---|---|---|---|
| 10–30 mL/min | 30–50 mL/min | ≥50 mL/min | Total | ||
|
| <10 mL/min | 1 | 0 | 0 | 1 |
| 10–30 mL/min | 21 | 2 | 0 | 23 | |
| 30–50 mL/min | 0 | 10 | 2 | 12 | |
| ≥50 mL/min | 0 | 0 | 24 | 24 | |
| Total | 22 | 12 | 26 | 60 | |
Comparison of creatinine clearance results determined by Nova StatSensor Xpress Creatinine analyzers or reference method modeled from the results obtained in the study applied to a hypothetical 50-kg, 20 year old female (N = 60).
| Nova StatSensor Xpress Creatinine | ||||||
|---|---|---|---|---|---|---|
| <10 mL/min | 10–30 mL/min | 30–50 mL/min | ≥50 mL/min | Total | ||
|
| <10 mL/min | 3 | 2 | 0 | 0 | 5 |
| 10–30 mL/min | 0 | 25 | 0 | 0 | 25 | |
| 30–50 mL/min | 0 | 2 | 6 | 0 | 8 | |
| ≥50 mL/min | 0 | 0 | 2 | 20 | 22 | |
| Total | 3 | 29 | 8 | 20 | 60 | |