Evaluation of flexural strength of Zirconia using three different connector designs: An in vitro study.

Mohammed Samiuddin Ahmed1, Kareti Mahendranadh Reddy1, Y Mahadev Shastry1, S Venkat Aditya1, P Jayakrishna Babu1.   

Abstract

AIM: The aim of this study is to evaluate the flexural strength of zirconia using three different connector designs under vertical and oblique loads. SETTING AND
DESIGN: Invitro - analytical study.
MATERIALS AND METHODS: For simulating zirconia fixed partial prosthesis, a specimen with three octagonal cylinders connected with each other was designed. Each face of the octagon was 3.75 mm ± 0.1 mm, and the total width was 9 mm ± 0.1 mm with a standard connector area of 10 mm2 at cross-section. Three different connector designs, i.e., round, oval, and triangular were milled. Universal testing machine was used to test flexural strength with vertical and oblique forces. STATISTICAL ANALYSIS USED: Intergroup comparison of flexural strength was made using Descriptive statistics (1) one-way ANOVA, Bonferroni's post hoc test (2) Kruskal-Wallis test. The confidence interval was set at 95%, P < 0.05 was considered statistically significant for both the tests.
RESULTS: The highest flexural strength was observed in the triangle connector with vertical forces and lowest with oblique forces.
CONCLUSIONS: Triangle connector design proved to be better than round and oval connectors on the application of vertical loads. Round connector design proved to be better than triangle and oval connector on application of oblique loads. Copyright:
© 2020 The Journal of Indian Prosthodontic Society.

Entities:  

Keywords:  Connector; fixed partial denture; fixed prosthesis; zirconia

Year:  2020        PMID: 33223698      PMCID: PMC7654207          DOI: 10.4103/jips.jips_68_20

Source DB:  PubMed          Journal:  J Indian Prosthodont Soc        ISSN: 0972-4052


INTRODUCTION

The zirconia-based fixed prosthetic restorations demand esthetically pleasing restorations with high strength. For achieving such results technicians design smaller connector size which allows them to give separation of units and naturally appearing embrasures. However, it compromises its overall strength and becomes more prone to fractures.[1] Thus, determining the ideal shape of the connector can be clinically useful. During mastication, the average force on the posteriors was reported to range between 300 and 880 N. Under horizontal and oblique load, it was found to be 275N. It has been reported that the connector design has an influence on the strength of zirconia prosthetic restorations.[2] Providing a proper cross-sectional dimension and shape of the rigid connectors could become challenging due to the specific, natural shape of the abutment teeth.[3] For Zirconia-based restorations, studies have evaluated the different sizes of the connectors. They recommended that the minimum size of the connector to fabricate a clinically acceptable zirconia restoration is 9 mm2 at cross-section.[4] The shape of the connector design needs further studies.

MATERIALS AND METHODS

The study was approved by Institutional review board Ref No. 616/SSCDS/IRB -E/2017.

Designing of specimens in CAD software

For simulating zirconia 3-unit fixed partial prosthesis, an octagonal specimen with three cylinders connected using different connector configuration were designed at RR Dental Labs Pvt Ltd by Mr Ramana Reddy(CDT), Telangana, India. The octagonal shape facilitates to apply oblique loads at 45°. Each octagonal face was 3.75 mm ± 0.1 mm, and the width of each cylinder was set at 9 mm ± 0.1 mm. The length of the cylinder was 26 mm ± 0.1 mm. Each connector designed had a standard area of 10 mm2. The separation between each cylinder was 2 mm due to milling limitations. Three different connector designs that were used in clinical scenarios were chosen, i.e., round, oval, and triangular [Figures 1-3] and prepared for milling. Ten samples were tested for vertical loads and 10 for 45° oblique loads for each of the connector designs totaling 60. The total number of samples has been thus divided into the following groups:
Figure 1

Schematic diagram of the round connector specimen

Figure 3

Schematic diagram of the triangle connector specimen

Schematic diagram of the round connector specimen Schematic diagram of the oval connector specimen Schematic diagram of the triangle connector specimen Group 1: Round connector for vertical force evaluation-(RV) Group 2: Oval connector for vertical force evaluation–(OV) Group 3: Triangular connector for vertical force evaluation-(TV) Group 4: Round connector for oblique force evaluation-(RO) Group 5: Oval connector for oblique force evaluation-(OO) Group 6: Triangular connector for oblique force evaluation-(TO).

Milling of zirconia specimens

The connector designs were milled out of Zirconia blanks (Shine T) of dimensions 98 mm diameter and 14-mm thickness using 5-axis CAD/CAM milling machine (IMES icore 250i). After milling, the specimens were detached from the mounting frame. The supports were grinded off carefully with a low-speed hand-piece using fine grit diamond bur. All the specimens were sintered according to the manufacturer's instructions (1500°C for 6 h) in a furnace (MIHM-VOGT). The specimens were verified for dimensional accuracy with an electronic caliper to an accuracy limit of 0.1 mm.

Testing on universal testing machine

Specimens were subjected to 3-point bend test using universal testing machine. In triangle shape connector, the base of the triangle was oriented upward for vertical load and turned clock-wise to test 45° oblique load. For oval shape connector, it was oriented such that longer dimension of oval was placed vertically for vertical loads and turned clock-wise to test 45° oblique load. No such orientation was required for testing round connector. The specimens were loaded by means of a mandrel of 6 mm width at a crosshead speed of 1 mm/min placed at the center of the octagonal cylinder.

RESULTS

Statistical analysis was performed using IBM SPSS version 25.0. Mean flexural strength between groups were analyzed using one-way ANOVA Bonferroni's post hoc test [Tables 1 and 2] and Kruskal–Wallis ANOVA with post-hoc analysis using Mann–Whitney tests [Tables 3 and 4]. The confidence interval was set at 95%. P < 0.05 was considered statistically significant. All the groups were found to be statistically significant. The highest strength was found in the triangle connector with vertical loads.
Table 1

Statistical analysis of vertical loads, one-way ANOVA, Bonferroni’s post hoc test

nMeanSDSE95% CI for meanMinimum-maximumPPost hoc analysis

Lower boundUpper bound
Round vertical101438.00235.9474.611269.221606.781180-1960<0.011, 2>3
Triangle veritcal101478.80215.2668.071324.811632.791053-1800
Oval vertical101095.60133.4442.201000.151191.05928-1320
Total301337.47260.2447.511240.291434.64928-1960

SD: Standard deviation, SE: Standard error, CI: Confidence interval

Table 2

Statistical analysis of oblique loads, one-way ANOVA, Bonferroni’s post hoc test

nMeanSDSE95% CI for meanMinimum-maximumSignificantPost hoc analysis

Lower boundUpper bound
Round oblique101393.20265.74484.0361203.101583.301018-1858<0.011>2, 3
Triangle oblique10931.00158.08949.992817.911044.09705-1138
Oval oblique101119.10243.32576.946945.041293.16828-1468
Total301147.77292.07053.3241038.711256.83705-1858

SD: Standard deviation, SE: Standard error, CI: Confidence interval

Table 3

Kruskal-Wallis ANOVA, Mann-Whitney post hoc tests for vertical loads

Vertical loadnMinimumMaximumMeanSDPPost hoc analysis
Round10118019601438235.940.0081, 2>3
Triangle (2)10105318001478215.26
Oval (3)1092813201095133.44

SD: Standard deviation

Table 4

Kruskal-Wallis ANOVA, Mann-Whitney post hoc tests for oblique loads

Oblique loadnMinimumMaximumMeanSDPPost hoc analysis
Round10101818581348.10265.740.0331>2, 3
Triangle107051138991.00158.08
Oval1082814681119.10243.32

SD: Standard deviation

Statistical analysis of vertical loads, one-way ANOVA, Bonferroni’s post hoc test SD: Standard deviation, SE: Standard error, CI: Confidence interval Statistical analysis of oblique loads, one-way ANOVA, Bonferroni’s post hoc test SD: Standard deviation, SE: Standard error, CI: Confidence interval Kruskal-Wallis ANOVA, Mann-Whitney post hoc tests for vertical loads SD: Standard deviation Kruskal-Wallis ANOVA, Mann-Whitney post hoc tests for oblique loads SD: Standard deviation

DISCUSSION

We designed octagonal cylinders connected by three different designs of connectors to simulate a 3 unit fixed partial denture (FPD). If we would have designed only connector shaped specimen of triangle, oval and round shape, it would have just represented forces on the connector directly and not forces directed toward the connector through a pontic. Although the use of an anatomic FPD shape would be more clinically relevant, a standardized geometrical shape was needed to calculate the flexural strength. Thus, an octagonal-shape was designed, incorporating two connectors in-between. While designing the specimens, we kept width between each octagonal cylinder as 2 mm due to milling limitations [Figures 4 and 5]. This 2 mm width definitely decreases the strength of the connector, but it is kept standard for all the samples for standardizing. Furthermore, our study is designed to compare different connector designs and not measuring the obsolute connector strength [Figure 6].
Figure 4

Diagramatic representation of the octagonal cylinder

Figure 5

Arrows showing the connector in milled zirconia specimen

Figure 6

Diagramatic representation of triangle shaped cylinder

Diagramatic representation of the octagonal cylinder Arrows showing the connector in milled zirconia specimen Diagramatic representation of triangle shaped cylinder The connector is definitely the weak point of the entire restorations and its size should be adjusted in height and width in order to allow long-term survival of the restoration without the danger of unexpected failure. In fact, in several studies it was shown that the failure of the restoration is almost always due to a fracture that begins at the connector area.[5] The size, shape, and position of connectors all influence the success of the prosthesis.[6] Schmitter et al. stated that 9 mm2 area at cross-section was the ideal connector dimension for zirconia fixed partial prosthesis frameworks. The connector they studied was of 9 mm2 and was found to be optimum for the strength of the prosthesis and soft tissue around the abutment teeth, which could improve both esthetics and periodontal health.[4] We chose connector area of 10 mm2 for our study as it depicts an average connector size in the premolar region. Pantea et al. who compared oval and round shaped zirconia connectors and stated that the behavior of the zirconia-based fixed prosthetic restoration is influenced to a large extent by achieving an optimal connector dimension and crown length. They compared connector of 5 mm2 and 9 mm2 and tested for flexural strength and found 9 mm2 connector of elliptical shape to be significantly stronger. They suggested that the elliptical connector might be stronger due to the wider area of stress distribution.[7] According to Clausen et al.[8] zirconia prosthesis can be used for posterior restorations. They stated that the ceramic had enough fracture strength to withstand mean masticatory force. In the posteriors, the mean maximum posterior masticatory forces varied from 300 to 880N. The resultant force exerted due to vertical load on the cuspal inclines of natural teeth were calculated using the formula (N = mg.cos θ). The average cuspal inclination of 37° was taken. In addition, we also calculated resultant force acting on the patient with bruxism habits. For posteriors, it was 276N and for bruxism habituates, it was 963N. The results of our in vitro study highlight the fact that the strength of zirconia-based fixed prosthetic restorations is influenced by the proper selection of the rigid connector design for the studied samples and the triangle-shaped connector ensured the best strength. All the connector designs were found capable to withstand vertical and lateral forces exerted during mastication. We suggest that appropriate design should be selected depending on the clinical situation. From the above, it is evident that the round-shaped connector subjected to oblique forces and the triangular-shaped connector subjected to vertical forces withstood the force better. Round connector withstood oblique forces better due to equal area of force distribution. Triangle connector withstood vertical forces better due to the flat base, which provided better distribution of forces. An important aspect to be taken into consideration when comparing all these results is the fact that most of the available scientific literature[91011121314151617181920] on zirconia strength uses geometric plane samples that do not reflect the actual configuration of a fixed prosthesis, which has curved lines or uneven material thickness, thus leading to an approach different from the ones applicable in clinical situations.

CONCLUSIONS

Within the limitations of the study, we conclude that: The highest flexural strength was observed in specimens with triangle connectors when force was applied vertically Round connector design was proved to be better than triangle and oval connector on application on oblique loads All the connector designs withstood both vertical and horizontal forces generated during normal mastication. Design of the connector is to be decided by the clinician/technician depending upon the clinical scenario. “One size fits all” cannot be applied in designing the shape and size of the connector.

Limitations

Milling of standardized samples in zirconia is challenging due to the difference in properties of different zirconia blanks Milling calibration changes due to wearing of bur will cause inaccuracies, which might give us false results.

Financial support and sponsorship

RR dental labs.

Conflicts of interest

There are no conflicts of interest.
  17 in total

1.  Effect of connector design on the fracture resistance of all-ceramic fixed partial dentures.

Authors:  Won-Suck Oh; Kenneth J Anusavice
Journal:  J Prosthet Dent       Date:  2002-05       Impact factor: 3.426

2.  Influence of connector design on fracture probability of ceramic fixed-partial dentures.

Authors:  W Oh; N Götzen; K J Anusavice
Journal:  J Dent Res       Date:  2002-09       Impact factor: 6.116

3.  All-ceramic fixed partial dentures designed according to the DC-Zirkon technique. A 2-year clinical study.

Authors:  P Vult von Steyern; P Carlson; K Nilner
Journal:  J Oral Rehabil       Date:  2005-03       Impact factor: 3.837

4.  Flexural strength of a layered zirconia and porcelain dental all-ceramic system.

Authors:  S N White; V G Miklus; E A McLaren; L A Lang; A A Caputo
Journal:  J Prosthet Dent       Date:  2005-08       Impact factor: 3.426

5.  In vitro lifetime of dental ceramics under cyclic loading in water.

Authors:  A R Studart; F Filser; P Kocher; L J Gauckler
Journal:  Biomaterials       Date:  2007-01-26       Impact factor: 12.479

6.  Flexural strength and fracture toughness of dental core ceramics.

Authors:  Handan Yilmaz; Cemal Aydin; Basak E Gul
Journal:  J Prosthet Dent       Date:  2007-08       Impact factor: 3.426

7.  Strength of zirconia fixed partial dentures: review of the literature.

Authors:  M Gargari; F Gloria; A Cappello; L Ottria
Journal:  Oral Implantol (Rome)       Date:  2011-01-23

8.  Dynamic fatigue and fracture resistance of non-retentive all-ceramic full-coverage molar restorations. Influence of ceramic material and preparation design.

Authors:  Jan-Ole Clausen; Milia Abou Tara; Matthias Kern
Journal:  Dent Mater       Date:  2010-02-23       Impact factor: 5.304

9.  Fracture resistance of lithium disilicate-, alumina-, and zirconia-based three-unit fixed partial dentures: a laboratory study.

Authors:  J Tinschert; G Natt; W Mautsch; M Augthun; H Spiekermann
Journal:  Int J Prosthodont       Date:  2001 May-Jun       Impact factor: 1.681

10.  Fracture strength of four-unit Y-TZP FPD cores designed with varying connector diameter. An in-vitro study.

Authors:  C Larsson; L Holm; N Lövgren; Y Kokubo; P Vult von Steyern
Journal:  J Oral Rehabil       Date:  2007-09       Impact factor: 3.837

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.