Effect of connector design on the fracture resistance of all-ceramic fixed partial dentures.

Won-Suck Oh1, Kenneth J Anusavice.   

Abstract

STATEMENT OF PROBLEM: Fracture of all-ceramic fixed partial dentures (FPDs) tends to occur in the connector area.
PURPOSE: The objective of this study was to test the hypothesis that the radii of curvature at the connector affects the fracture resistance of 3-unit FPDs.
MATERIAL AND METHODS: With the use of a standardized silicone mold, 40 three-unit FPD wax patterns were fabricated with the same dimensions and divided into 4 groups of 10 specimens per group. Each pattern was modified at the connector areas of the occlusal embrasure (OE) and the gingival embrasure (GE); 2 wax carvers with radii of curvature at their tips of 0.90 mm and 0.25 mm were used. The dimensions of the connectors were standardized with an electronic caliper to 4 +/- 0.12 mm in height and 5 +/- 0.13 mm in width. Connector designs were as follows: Design I: OE and GE 0.90 mm; Design II: OE 0.90 mm and GE 0.25 mm; Design III: OE 0.25 mm and GE 0.90 mm; and Design IV (control): OE and GE 0.25 mm. An experimental hot-pressed core ceramic was used to make the FPD frameworks, which were consequently cemented on epoxy dies with dual-polymerizing composite (Variolink II) and loaded to fracture in a universal testing machine at a crosshead speed of 0.5 mm/min. The failure load data were analyzed with analysis of variance (ANOVA; P=.05) and Duncan's test (alpha=.01).
RESULTS: The mean failure loads and standard deviations were as follows: 943 +/- 151 N for Design I; 746 +/- 106 N for Design II; 944 +/- 144 N for Design III; and 673 +/- 55 N for Design IV. ANOVA revealed a significant difference (P< or = .0001) between the mean failure loads of different connector designs. The mean loads to failure for Designs I and III were significantly higher than those for Designs II and IV (Duncan's test).
CONCLUSION: Within the limitations of this study and for the experimental ceramic tested, as the radius at the gingival embrasure increased from 0.25 to 0.90 mm, the mean failure load increased by 140%. The radius of curvature at the occlusal embrasure had only a minor effect on the fracture susceptibility of 3-unit FPDs.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 12070517     DOI: 10.1067/mpr.2002.123850

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  27 in total

1.  Finite element analysis to compare stress distribution of connector of lithia disilicate-reinforced glass-ceramic and zirconia-based fixed partial denture.

Authors:  Jie Lin; Akikazu Shinya; Harunori Gomi; Akiyoshi Shinya
Journal:  Odontology       Date:  2011-06-21       Impact factor: 2.634

2.  Strength of zirconia fixed partial dentures: review of the literature.

Authors:  M Gargari; F Gloria; A Cappello; L Ottria
Journal:  Oral Implantol (Rome)       Date:  2011-01-23

3.  Stress distribution on a three-unit implant-supported zirconia framework. A 3D finite element analysis and fatigue test.

Authors:  G Sannino; A Pozzi; R Schiavetti; A Barlattani
Journal:  Oral Implantol (Rome)       Date:  2012-07-17

4.  Fatigue of zirconia and dental bridge geometry: Design implications.

Authors:  G D Quinn; A R Studart; C Hebert; J R VerHoef; D Arola
Journal:  Dent Mater       Date:  2010-09-01       Impact factor: 5.304

5.  Five-year prospective clinical study of posterior three-unit zirconia-based fixed dental prostheses.

Authors:  Roberto Sorrentino; Giorgio De Simone; Stefano Tetè; Simona Russo; Fernando Zarone
Journal:  Clin Oral Investig       Date:  2011-06-11       Impact factor: 3.573

6.  Fracture resistance of three-unit zirconia fixed partial denture with modified framework.

Authors:  Arthur Partiyan; Essam Osman; Mohammad M Rayyan; Moustafa Aboushelib; Ahmed Ibrahim; Ryo Jimbo
Journal:  Odontology       Date:  2016-04-15       Impact factor: 2.634

7.  Fatigue limit of monolithic Y-TZP three-unit-fixed dental prostheses: Effect of grinding at the gingival zone of the connector.

Authors:  Marina Amaral; Regina F Villefort; Renata Marques Melo; Gabriel K R Pereira; Yu Zhang; Luiz Felipe Valandro; Marco Antonio Bottino
Journal:  J Mech Behav Biomed Mater       Date:  2017-03-07

8.  Reliability and failure modes of implant-supported zirconium-oxide fixed dental prostheses related to veneering techniques.

Authors:  Marta Baldassarri; Yu Zhang; Van P Thompson; Elizabeth D Rekow; Christian F J Stappert
Journal:  J Dent       Date:  2011-04-28       Impact factor: 4.379

9.  Fractographic analysis of a dental zirconia framework: a case study on design issues.

Authors:  Ulrich Lohbauer; Gudrun Amberger; George D Quinn; Susanne S Scherrer
Journal:  J Mech Behav Biomed Mater       Date:  2010-07-16

10.  Randomized clinical trial of implant-supported ceramic-ceramic and metal-ceramic fixed dental prostheses: preliminary results.

Authors:  Josephine F Esquivel-Upshaw; Arthur E Clark; Jonathan J Shuster; Kenneth J Anusavice
Journal:  J Prosthodont       Date:  2013-06-12       Impact factor: 2.752

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.