Milena Małecka1,2, Olga Ciepiela1,2.
Abstract
A complete blood count is a highly automated laboratory test. The use of highly advanced measurement methods increases the accuracy and sensitivity of the determination of individual hematological parameters, especially in the case of white blood cells differentiation. Therefore, it is necessary to make comparative analyses, which involve performance of the analyzers used in daily work. The aim of the study was to indicate whether the results obtained using two compared analyzers show significant differences.Entities:
Keywords: Automated hematology analyzer; Comparison; Complete blood count; Horiba Yumizen H2500; Sysmex XN-2000
Year: 2020 PMID: 33204791 PMCID: PMC7649520 DOI: 10.1016/j.plabm.2020.e00186
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
The number of samples analyzed for Sysmex XN-2000 and YUMIZEN H2500.
| Parameter | Number of samples |
|---|---|
| WBC [109/L] | 241 |
| Neutrophils [109/L ] | 223 |
| Lymphocytes [109/L ] | 223 |
| Monocytes [109/L ] | 223 |
| Eosinophils [109/L ] | 223 |
| Immature granulocytes [109/L ] | 223 |
| RBC [1012/L], HGB [g/L], HCT [%], MCV [fL], | 241 |
| RDW-CV [%] | 221 |
| PLT [109/L ] | 241 |
| MPV [fL], PCT [%], | 204 |
Detection limit of Yumizen H2500 and Sysmex XN-2000.
| Parameter [Units] | Detection limit of Yumizen H2500 | Detection limit of Sysmex XN-2000 |
|---|---|---|
| WBC [109/L] | 0.000–300 | 0.000–440 |
| RBC [1012/L] | 0.0–8.8 | 0.00–8.60 |
| HGB [g/L] | 0.0–24.5 | 0.0–26.0 |
| HCT [%] | 0.0–68.8 | 0.0–75.0 |
| PLT [109/L ] | 0.0–5000 | 0.0–5000 |
A passing-bablok regression analysis for sysmex XN-2000 and YUMIZEN H2500 comparison.
| Sysmex XN-2000 and YUMIZEN H2500 | ||
|---|---|---|
| Parameter | Correlation coefficient (r) | Significance level (P value) |
| 0.991 | P = 0.29 | |
| 0.947 | P = 0.42 | |
| 0.887 | P = 0.33 | |
| 0.851 | P = 0.16 | |
| 0.826 | P < 0.01 | |
| 0.796 | P < 0.01 | |
| 0.987 | P = 0.68 | |
| 0.994 | P = 0.46 | |
| 0.962 | P = 0.57 | |
| 0.860 | P = 0.29 | |
| 0.960 | P = 0.17 | |
| 0.392 | P = 0.01 | |
| 0.782 | P = 0.85 | |
| 0.993 | P = 0.48 | |
| 0.929 | P = 0.32 | |
| 0.990 | P = 0.28 | |
| 0.931 | P = 0.04 | |
Fig. 1A Passing- Bablok regression and Bland-Altman Analysis for SYSMEX XN-2000 and YUMIZEN H2500 comparison. Graphs for hematological parameters showed statistically significant differences between the analyzers compared.