Literature DB >> 24225753

Performance and abnormal cell flagging comparisons of three automated blood cell counters: Cell-Dyn Sapphire, DxH-800, and XN-2000.

Julie Hotton1, Jan Broothaers, Caroline Swaelens, Brigitte Cantinieaux.   

Abstract

OBJECTIVES: To compare two hematological analyzers-the DxH-800 (DxH; Beckman-Coulter, Miami, FL) and XN-2000 (XN; Sysmex, Kobe, Japan)-with the Cell-Dyn Sapphire (SAPH; Abbott, Santa Clara, CA).
METHODS: We analyzed 4,375 samples. Slide reviews were made in the presence of blast, abnormal lymphocyte, and immature granulocyte (IG) flags or nucleated RBC (NRBC) count.
RESULTS: The analyzers exhibited excellent correlations for CBC and neutrophils but displayed a limit correlation for lymphocytes. The XN did not miss circulating blasts (0.5%-95% in microscopy). For NRBCs, the XN demonstrated a sensitivity of 90%; DxH, 74%; and SAPH, 29%. Only the XN demonstrated a correlation with microscopy, permitting a WBC six-part differential until 15% of NRBCs. The XN and DxH gave useful IG counts with a cutoff less than 5% and a WBC level more than 2,500/mm(3). For abnormal lymphocytes detection, only XN demonstrated sensitivity of more than 95%, but its specificity of 54% requires adaptation.
CONCLUSION: The XN increases the sensitivity of abnormal cell detection compared with the other counters, permitting a seven-part differential between predefined levels, decreasing the slide review from 20% to 9%.

Entities:  

Keywords:  Abnormal lymphocytes; Blast; Cell-Dyn Sapphire; DxH-800; Flag; HemaCAM; IG; NRBC; XN-2000

Mesh:

Year:  2013        PMID: 24225753     DOI: 10.1309/AJCPE5R4SOQBUULZ

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  10 in total

1.  Validation rules for blood smear revision after automated hematological testing using Mindray CAL-8000.

Authors:  Sabrina Buoro; Tommaso Mecca; Michela Seghezzi; Barbara Manenti; Giovanna Azzarà; Cosimo Ottomano; Giuseppe Lippi
Journal:  J Clin Lab Anal       Date:  2016-10-06       Impact factor: 2.352

2.  Performance evaluation of the white cell precursor channel on the Sysmex XN hematology analyzer in HIV specimens.

Authors:  Elise Schapkaitz; Phoyisile Khoza
Journal:  J Clin Lab Anal       Date:  2017-09-26       Impact factor: 2.352

3.  A comparison of Sysmex-XN 2000 and Yumizen H2500 automated hematology analyzers.

Authors:  Milena Małecka; Olga Ciepiela
Journal:  Pract Lab Med       Date:  2020-10-29

4.  Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers.

Authors:  Shubham Rastogi; Sukesh Chandran Nair; Pandiyan Murugan; Asady Sukanya Sukumar; Joy J Mammen; Saravanan Mullai
Journal:  Pract Lab Med       Date:  2022-04-18

5.  A Comparison of Mindray BC-6800, Sysmex XN-2000, and Beckman Coulter LH750 Automated Hematology Analyzers: A Pediatric Study.

Authors:  Olga Ciepiela; Iwona Kotuła; Szymon Kierat; Sandra Sieczkowska; Anna Podsiadłowska; Anna Jenczelewska; Karolina Księżarczyk; Urszula Demkow
Journal:  J Clin Lab Anal       Date:  2016-05-17       Impact factor: 2.352

6.  Detection of Apoptotic Lymphocytes Through Sysmex XN-1000 As a Diagnostic Marker for Mononucleosis Syndrome.

Authors:  Silvia Sale; Addolorata Emanuela Carone; Maurizio Fumi; Ylenia Pancione; Vincenzo Rocco
Journal:  J Clin Lab Anal       Date:  2016-04-19       Impact factor: 2.352

7.  Comments re article on comparison of performance and abnormal cell flagging of two automated hematology analyzers: Sysmex XN 3000 and Beckman Coulter DxH 800.

Authors:  Elena Sukhacheva
Journal:  Int J Lab Hematol       Date:  2019-11-26       Impact factor: 2.877

8.  Analytical comparison between two hematological analyzer systems: Mindray BC-5180 vs Sysmex XN-1000.

Authors:  Jing Wang; Shengmei Zhao; Zhenguo Su; Xiangdong Liu
Journal:  J Clin Lab Anal       Date:  2019-06-20       Impact factor: 2.352

9.  Mature and Immature/Activated Cells Fractionation: Time for a Paradigm Shift in Differential Leucocyte Count Reporting?

Authors:  Rana Zeeshan Haider; Najeed Ahmed Khan; Eloisa Urrechaga; Tahir Sultan Shamsi
Journal:  Diagnostics (Basel)       Date:  2021-05-21

10.  Silicone Oil Emulsification after Vitrectomy for Rhegmatogenous Retinal Detachment.

Authors:  Jian Yu; Yuan Zong; Chunhui Jiang; Haohao Zhu; Guohua Deng; Gezhi Xu
Journal:  J Ophthalmol       Date:  2020-02-24       Impact factor: 1.909

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.