Andrea Pradal1, Ludovica Nucci2, Nicola Derton1, Maria Elena De Felice2, Gianluca Turco1, Vincenzo Grassia2, Luca Contardo1.
Abstract
The aim of the present study was to evaluate the primary stability of a two-miniscrew system inserted into a synthetic bone and to compare the system with the traditional one. Forty-five bi-layered polyurethane blocks were used to simulate maxillary cancellous and cortical bone densities. Samples were randomly assigned to three groups-one-miniscrew system (Group A, N = 23), two-miniscrew system (Group B, N = 22) and archwire-only (Group C, N = 10). A total of 67 new miniscrews were subdivided into Group A (23 singles) and Group B (22 couples). 30 mm of 19″ × 25″ archwires were tied to the miniscrew. The load was applied perpendicularly to the archwire. Maximum Load Value (MLV), Yield Load (YL) and Loosening Load (LL) were recorded for each group. The YL of Group B and C had a mean value respectively of 4.189 ± 0.390 N and 3.652 ± 0.064 N. The MLV of Group A, B and C had a mean value respectively of 1.871 ± 0.318N, of 4.843 ± 0.515 N and 4.150 ± 0.086 N. The LL of Group A and B had a mean value respectively of 1.871 ± 0.318 N and of 2.294 ± 0.333 N. A two- temporary anchorage device (TAD) system is on average stiffer than a one-TAD system under orthodontic loading.Entities:
Keywords: TADs; bone anchorage; load value; miniscrews; primary stability
Year: 2020 PMID: 33167442 PMCID: PMC7711976 DOI: 10.3390/jfb11040080
Source DB: PubMed Journal: J Funct Biomater ISSN: 2079-4983
Mann-Whitney Test for Insertion Torque Values.
| One-Miniscrew | Two-Miniscrews | ||||
|---|---|---|---|---|---|
| Mean Value | SD | Mean Value | SD |
| |
| Insertion Torque (Ncm) | 3.870 | 0.388 | 4.014 | 0.399 | 0.089 |
Figure 1Mechanical test of one-miniscrew.
Figure 2Test design and clinical simulation.
Figure 3Boxplot distribution Insertion Torque Values.
Main Value for Yield Load, Maximum Load Value, Loosening Load for Group A, B and C.
| One-Miniscrew (23 Singles) | Two-Miniscrews (22 Couples) | Archwire Only | |||||
|---|---|---|---|---|---|---|---|
| Mean Value | SD | Mean Value | SD | Mean Value | SD |
| |
| Yield (N) | NA | NA | 4.189 N | ±0.390 N | 3.652 N | ±0.064 N | 0.000 |
| Maximum Load Value (N) | 1.871 N | ±0.318 N | 4.843 N | ±0.515 N | 4.150 N | ±0.086 N | 0.000 |
| Loosening Load (N) | 1.871 N | ±0.318N | 2.294 N | ±0.333 N | NA | NA | 0.000 |
Figure 4Boxplot Yield Points Group B and Group C.
Figure 5Boxplot Maximum Load Values of Group A, B, C.
Figure 6Boxplot Loosening Load of Group A and Group B.
Figure 7Load-Displacement Graph.