| Literature DB >> 33134215 |
Ahmadi A1, Negarestani A2, Sina S3, Sarshough S4.
Abstract
BACKGROUND: Thermoluminescence dosimetry(TLD) has been known as one of the most effective methods for dose estimation in diagnostic radiology. Orthopantomogram (OPG) imaging is used by many dentists, oral and maxillofacial surgeons as an effective tool for choosing an appropriate treatment plan.Entities:
Keywords: Diagnostic Radiology; Panoramic; Radiography; Radiology; Rando Phantom; TLD-100; Thermoluminescent Dosimetry
Year: 2020 PMID: 33134215 PMCID: PMC7557459 DOI: 10.31661/jbpe.v0i0.878
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure 1Imaging geometry
Figure 2Thermoluminescence dosimetry(TLD) -100 calibration curve
Figure 3Comparison of absorbed dose of different organs in Orthopantomogram (OPG) imaging.
Entrance dose measured in each step
| Organ | Surface dose (µGy) | |
|---|---|---|
| Entrance skin dose(µGy) | ||
| First time | Second time | |
| Right cheek | 79.55 | |
| Left cheek | 47.24 | |
| Temporal surface-Right | 11.15 | |
| Temporal surface-Left | 4.80 | |
| Head | 13.29 | |
| Right back of the neck | 360.58 | |
| Left back of the neck | 334.28 | |
| Neck | 377.59 | |
| Thyroid surface | 52.68 | |
| Eye-Right | 29.25 | 30.00 |
| Eye-Left | 29.79 | 24.98 |
Thermoluminescence dosimetry error analysis of dosimeters
| Uncertainty Component | Type A (%) | Type B (%) |
|---|---|---|
| Iteration of measurements with TLDs | 4 | --- |
| Dose Calibration | --- | 5 |
| Correction of energy dependency | --- | --- |
| Correction for nonlinearity and fading | --- | --- |
| Combination of errors | σA-total=4 | σB-total=5 |
| Total error | ||
Comparison of measured doses in different investigations with the results of this study
| This study 68kv, 8mA, 13.5s | Gavala [ | Eftekhari moghadam [ | |
|---|---|---|---|
| Country | Iran | Athens | Iran |
| Year | 2017 | 2009 | 2015 |
| parotid | 3397.48 | 320 | 1825 |
| thyroid | 46.75 | 60 | 1415 |
| submandibular | 1572.50 | 416 | 1450 |
| Entrance skin Dose (µGy) | 360.58 Right back of neck; 334.28 Left back of neck | 170 back of neck | |
| 90 skin (occipital area) | |||
| 79.55 for right cheek | 20 for skin (zygomatic area) | ||
| 47.24 for left cheek | |||