M Sadeghi1, S Sina1, R Faghihi2.
Abstract
LiF, Mg and Ti cubical TLD chips (known as TLD-100) are widely used for dosimetry purposes. The repeatability of TL dosimetry is investigated by exposing them to doses of (81, 162 and 40.5 mGy) with 662keV photons of Cs-137. A group of 40 cubical TLD chips was randomly selected from a batch and the values of Element Correction Coefficient (ECC) were obtained 4 times by irradiating them to doses of 81 mGy (two times), 162mGy and 40.5mGy. Results of this study indicate that the average reproducibility of ECC calculation for 40 TLDs is 1.5%, while these values for all chips do not exceed 5%.Entities:
Keywords: ECC; LiF; Mg, and Ti; Reproducibility; TLD-100
Year: 2015 PMID: 26688801 PMCID: PMC4681467
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Time-temperature profile (TTP) used for TLD100 read out
|
|
| |
|---|---|---|
|
LiF: Mg, Ti |
| |
| Temp (◦C) | 50 | |
| Time (sec) | 0 | |
|
| ||
| Max Temp (◦C) | 300 | |
| Time (sec) | 13.33 | |
| Rate (◦C/sec) | 25 | |
|
| ||
| Temp (◦C) | 0 | |
| Time (sec) | 0 | |
Figure 1The coefficient of variation (CV%) for the response of each TLD
Percentage difference between ECC values
|
|
|
| |
|---|---|---|---|
| minimum | 0.05 | 0.02 | 0.05 |
| maximum | 4.38 | 4.75 | 3.09 |
| average | 1.63 | 1.65 | 1.17 |
Figure 2Reproducibility of the ECC
Figure 3Coefficient of variation (CV%) for the ECC of each TLD