Literature DB >> 10673663

Effective dose from radiation absorbed during a panoramic examination with a new generation machine.

R A Danforth1, D E Clark.   

Abstract

OBJECTIVES: The purpose of this investigation was to measure the tissue-absorbed dose and to calculate the whole-body effective dose (E) for a new generation panoramic machine (Planmeca PM 2002 CC Proline), operating in the panoramic examination mode. Comparisons could then be made with historical panoramic and intraoral radiographic dose measurement values and effective dose estimates. STUDY
DESIGN: Thermoluminescent dosimeters were embedded in a humanoid, tissue-equivalent phantom at anatomically significant sites, representing key tissues. Absorbed dose measurements were obtained after every 5 panoramic exposures of a 25-exposure total. The measured average tissue-absorbed doses from a single panoramic exposure were used in the calculation of the whole-body E.
RESULTS: The whole-body E for the PM 2002 CC Proline panoramic examination is 3.85 microSv. This is below the panoramic average of 6.7 microSv. The PM 2002 CC Proline panoramic examination delivers approximately 5% to 12% of the E of a complete mouth intraoral radiographic examination.
CONCLUSIONS: The effective dose from the PM 2002 CC Proline examination is at the low end of the range for other panoramic machines and is far below either a D-speed or E-speed film intraoral radiographic examination.

Entities:  

Mesh:

Year:  2000        PMID: 10673663     DOI: 10.1067/moe.2000.103526

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  18 in total

1.  Characterization of optically stimulated luminescence dosemeters to measure organ doses in diagnostic radiology.

Authors:  A Endo; T Katoh; I Kobayashi; R Joshi; J Sur; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2011-11-24       Impact factor: 2.419

2.  Shielding effect of thyroid collar for digital panoramic radiography.

Authors:  G-S Han; J-G Cheng; G Li; X-C Ma
Journal:  Dentomaxillofac Radiol       Date:  2013-09-04       Impact factor: 2.419

3.  Detection accuracy of proximal caries by phosphor plate and cone-beam computerized tomography images scanned with different resolutions.

Authors:  Jun-Ge Cheng; Zhi-Ling Zhang; Xiao-Yan Wang; Zu-Yan Zhang; Xu-Chen Ma; Gang Li
Journal:  Clin Oral Investig       Date:  2011-07-30       Impact factor: 3.573

Review 4.  Management of non-syndromic dens evaginatus affecting permanent maxillary central incisors: a systematic review.

Authors:  Violaine Smail-Faugeron; Julie Picou Rollin; Michèle Muller Bolla; Frederic Courson
Journal:  BMJ Case Rep       Date:  2016-10-05

5.  Patient radiation dose and protection from cone-beam computed tomography.

Authors:  Gang Li
Journal:  Imaging Sci Dent       Date:  2013-06-14

6.  Sequential test selection by quantifying of the reduction in diagnostic uncertainty for the diagnosis of proximal caries.

Authors:  Umut Arslan; Ahmet Ergun Karaağaoğlu; Nursel Akkaya; Leyla Berna Cağırankaya; Ayşe Özden Kansu; Hilmi Kansu
Journal:  Balkan Med J       Date:  2013-06-01       Impact factor: 2.021

Review 7.  Management of non-syndromic double tooth affecting permanent maxillary central incisors: a systematic review.

Authors:  Violaine Smail-Faugeron; Jeanne Terradot; Michèle Muller Bolla; Frederic Courson
Journal:  BMJ Case Rep       Date:  2016-06-15

8.  Thyroid Radiation Dose to Patients from Diagnostic Radiology Procedures over Eight Decades: 1930-2010.

Authors:  Lienard A Chang; Donald L Miller; Choonsik Lee; Dunstana R Melo; Daphnée Villoing; Vladimir Drozdovitch; Isabelle Thierry-Chef; Sarah J Winters; Michael Labrake; Charles F Myers; Hyeyeun Lim; Cari M Kitahara; Martha S Linet; Steven L Simon
Journal:  Health Phys       Date:  2017-12       Impact factor: 1.316

9.  Effective dose from direct and indirect digital panoramic units.

Authors:  Gun-Sun Lee; Jin-Soo Kim; Yo-Seob Seo; Jae-Duk Kim
Journal:  Imaging Sci Dent       Date:  2013-06-14

10.  An in-office, cost effective technique for measuring width of bone using intra-oral periapical radiographs in occlusal projection.

Authors:  Shrikar R Desai; I Karthikeyan; Rika Singh
Journal:  J Indian Soc Periodontol       Date:  2013-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.