Burak Cilhoroz1,2, David Giles3, Amanda Zaleski1,4, Beth Taylor1,4, Bo Fernhall5, Linda Pescatello1,6.
Abstract
Heart rate variability (HRV) measurements via ambulatory monitors have become common. We examined the validity of recording R-R intervals using the Polar V800™ compared to 12-lead electrocardiograms (ECG) among middle-aged (44.7±10.1years); overweight to obese (29.8±4.3 kg.m-2) adults (n = 25) with hypertension (132.3±12.2/ 84.3±10.2 mmHg). After resting for 5-min in the supine position, R-R intervals were simultaneously recorded using the Polar V800™ and the 12-lead ECG. Artifacts present in uncorrected (UN) R-R intervals were corrected with the Kubios HRV Premium (ver. 3.2.) automatic (AC) and threshold-based (TBC) correction, and manual correction (MC) methods. Intra-class correlation coefficients (ICC), Bland-Altman limits of agreement (LoA), and effect sizes (ES) were calculated. We detected 71 errors with the Polar V800™ for an error rate of 0.85%. The bias (LoAs), ES, and ICC between UN and ECG R-R intervals were 0.69ms (-215.80 to +214.42ms), 0.004, and 0.79, respectively. Correction of artifacts improved the agreeability between the Polar V800™ and ECG HRV measures. The biases (LoAs) between the AC, TBC, and MC and ECG R-R intervals were 3.79ms (-130.32 to +137.90ms), 1.16ms (-92.67 to +94.98ms), and 0.37ms (-41.20 to +41.94ms), respectively. The ESs of AC, TBC, and MC were 0.024, 0.008, and 0.002, and ICCs were 0.91, 0.95, and 1.00, respectively. R-R intervals measured using the Polar V800™ compared to 12-lead ECG were comparable in adults with hypertension, especially after the artifacts corrected by MC. However, TBC correction also yielded acceptable results.Entities:
Mesh:
Year: 2020 PMID: 33031480 PMCID: PMC7544136 DOI: 10.1371/journal.pone.0240220
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Types, description, and correction methods of artifacts.
| Type | Description | Correction |
|---|---|---|
| T1 | A positive or negative single interval difference greater than 20 ms | Artifact noted but not corrected |
| T2 | A long interval followed by a short interval while the two R-R intervals either side were less than 20 ms | Long and short intervals averaged |
| T3 | A short interval followed by a long interval while the two R-R intervals on either side were less than 20 ms | Short and long intervals averaged |
| T4 | When the HR monitor missed R-R intervals equivalent to two or three ECG R-R intervals | Artificially long R-R interval divided by the number of missed R-waves (beats) |
| T5 | When the HR monitor detected extra short R-R intervals equivalent to one ECG R-R interval | Two extra short R-R intervals added together to approach corresponding ECG R-R interval |
| T6a | When the HR monitor entirely missed the R-R interval, while there was no difference in the HR monitor time stamp | Artifact noted but not corrected |
| T6b | When the HR monitor entirely missed the R-R interval and there was a larger than expected gap in the HR monitor time stamp | The known interval subtracted from the difference in time stamp |
aArtifact is undetectable without simultaneous ECG recording.
bR-R interval of 0 ms was placed to maintain time synchrony between both devices.
Subjects characteristics of the total sample and by sex (mean ± SD).
| Variable | Total Sample (n = 25) | Men (n = 17) | Women (n = 8) |
|---|---|---|---|
| Age (year) | 44.7±10.1 | 42.5±10.3 | 49.4±8.4 |
| Height (cm) | 172.3±11.0 | 176.9±9.6 | 162.6±5.4 |
| Weight (kg) | 93.7±30.6 | 90.9±16.2 | 86.4±17.2 |
| Body Mass Index (kg.m-2) | 29.8±4.3 | 28.2±3.9 | 32.0±5.3 |
| Waist Circumference (cm) | 99.9±12.1 | 95.9±10.7 | 105.1±14.0 |
| Heart Rate (bpm) | 73.6±11.0 | 71.8±11.9 | 78.4±7.2 |
| Systolic Blood Pressure (mmHg) | 132.3±12.2 | 129.6±11.8 | 140.9±7.9 |
| Diastolic Blood Pressure (mmHg) | 84.3±10.2 | 84.4±11.3 | 86.5±7.5 |
| Maximum Oxygen Consumption (VO2peak) | 33.5±9.8 | 38.0±8.2 | 24.1±5.5 |
aSignificant differences at P < 0.05.
Fig 1Bland-Altman plots for the ECG and the UN Polar V800™ HR monitor R-R intervals (a) Bland-Altman plots for the ECG and the AC Polar V800™ HR monitor R-R intervals (b) Bland-Altman plots for the ECG and the TBC Polar V800™ HR monitor R-R intervals (c) Bland-Altman plots for the ECG and the MC Polar V800™ HR monitor R-R intervals (d) ECG: electrocardiogram; HR: heart rate; UN: uncorrected R-R intervals; AC: R-R intervals corrected by automatic correction method of the Kubios HRV Premium (ver 3.2); TBC: R-R intervals corrected by threshold-based correction method of the Kubios HRV Premium (ver 3.2); MC: R-R intervals corrected manually.
Comparison of HRV measures calculated from UN Polar V800™ and ECG R-R intervals (mean ± SD).
| HRV Measure | ECG (mean±SD) | Polar UN (mean±SD) | Bias (LoA) | ICC (95% CI) | Effect Size |
|---|---|---|---|---|---|
| SDNN (ms) | 55.5±26.7 | 90.5±62.2 | -34.95 (-143.58 to 73.69) | 0.27 (-0.07–0.57) | 0.730 |
| RMSSD (ms) | 38.3±29.4 | 92.8±96.3 | -57.51 (-223.81 to 114.78) | 0.21 (-0.11–0.52) | 0.765 |
| pNN50 (%) | 13.6±16.2 | 15.1±15.9 | -1.50 (-6.14 to 3.21) | 0.98 (0.95–0.99) | 0.091 |
| LF (ms2) | 1215±1352.1 | 6924.2±16947.2 | -5709.13 (-38536.30 to 27118.05) | 0.53 (-0.97–0.57) | 0.475 |
| HF (ms2) | 740.8±1262.8 | 4861.0±6863.9 | -4120.18 (-17136.40 to 8896.033) | 0.07 (-0.21–0.39) | 0.835 |
| LF (nu) | 67.3±18.7 | 59.7±20.3 | 7.53 (-30.13 to 45.20) | 0.49 (0.14–0.73) | 0.387 |
| HF (nu) | 32.7±18.6 | 40.2±20.2 | -7.53 (-45.20 to 30.13) | 0.49 (0.14–0.73) | 0.387 |
| LF/HF Ratio | 3.6±4.0 | 3.0±4.2 | 0.67 (-3.37 to 4.71) | 0.87 (0.72–0.94) | 0.163 |
| SampEn | 1.5±0.2 | 1.2±0.5 | 0.33± (-0.58 to 1.24) | 0.39 (-0.18 to 0.71) | 0.836 |
ECG: electrocardiogram; UN: uncorrected R-R intervals; ICC: intra-correlation coefficient; LoA: limits of agreement; SDNN: standard deviation of normal-to-normal N-N intervals; RMSSD: root mean square of the successive difference of intervals; pNN50%: the percentage of successive normal cardiac inter-beat intervals greater than 50 msec; LF: low frequency; HF: high frequency; SampEn: Sample Entropy; SD: standard deviation.
Comparison of HRV measures calculated from Kubios Premium (ver. 3.2) AC Polar V800™ and ECG R-R intervals (mean ± SD).
| HRV Measure | ECG (mean±SD) | Polar AC (mean±SD) | Bias (LoA) | ICC (95% CI) | Effect Size |
|---|---|---|---|---|---|
| SDNN (ms) | 55.5±26.7 | 57.1±28.1 | -1.60 (-23.27 to 20.06) | 0.92 (0.83–0.96) | 0.058 |
| RMSSD (ms) | 38.3±29.4 | 36.2±25.4 | 2.13 (-25.57 to 29.84) | 0.87 (0.73–0.94) | 0.078 |
| pNN50 (%) | 13.6±16.2 | 13.1±16.1 | 0.51 (-7.28 to 8.29) | 0.97 (0.93–0.99) | 0.031 |
| LF (ms2) | 1215±1352.1 | 1214.2±1360.9 | 0.80 (-369.45 to 371.06) | 0.99 (0.98–1.00) | 0.001 |
| HF (ms2) | 740.8±1262.8 | 749.1±1239.5 | -8.27 (-1115.84 to 1099.293) | 0.90 (0.79–0.96) | 0.007 |
| LF (nu) | 67.3±18.7 | 66.7±19.6 | 0.57 (-24.29 to 25.44) | 0.79 (0.57–0.90) | 0.030 |
| HF (nu) | 32.7±18.6 | 33.3±19.6 | -0.57 (-25.44 to 24.29) | 0.79 (0.73–0.95) | 0.030 |
| LF/HF Ratio | 3.6±4.0 | 3.5±3.9 | 0.12 (-1.76 to 2.00) | 0.97 (0.94–0.99) | 0.031 |
| SampEn | 1.5±0.2 | 1.5±0.3 | 0.02 (-0.21 to 0.21) | 0.81 (0.57–0.91) | 0.039 |
AC: R-R intervals corrected by automatic correction method of Kubios HRV Premium (ver. 3.2).
Comparison of HRV measures calculated from Kubios Premium (ver. 3.2) TBC Polar V800™ and ECG R-R intervals (mean ± SD).
| HRV Measure | ECG (mean±SD) | Polar TBC (mean±SD) | Bias (LoA) | ICC (95% CI) | Effect Size |
|---|---|---|---|---|---|
| SDNN (ms) | 55.5±26.7 | 56.1±27.6 | -0.55 (-10.59 to 9.49) | 0.98 (0.96–0.99) | 0.020 |
| RMSSD (ms) | 38.3±29.4 | 37.8±27.5 | 0.53 (-9.57 to 10.63) | 0.98 (0.96–0.99) | 0.019 |
| pNN50 (%) | 13.6±16.2 | 13.8±16.2 | -0.16 (-1.33 to 1.02) | 0.99 (0.98–1.00) | 0.010 |
| LF (ms2) | 1215±1352.1 | 1209.2±1352.5 | 5.86 (-275.86 to 287.59) | 1.00 (0.99–1.00) | 0.004 |
| HF (ms2) | 740.8±1262.8 | 703.7±1201.6 | 37.07 (-460.59 to 534.74) | 0.98 (0.95–0.99) | 0.030 |
| LF (nu) | 67.3±18.7 | 67.3±18.0 | -0.02 (-4.04 to 4.01) | 0.99 (0.99–1.00) | 0.001 |
| HF (nu) | 32.7±18.6 | 32.7±18.0 | 0.02 (-4.01 to 4.04) | 0.99 (0.99–1.00) | 0.001 |
| LF/HF Ratio | 3.6±4.0 | 3.5±3.9 | 0.12 (-0.79 to 1.02) | 0.99 (0.98–1.00) | 0.029 |
| SampEn | 1.5±0.2 | 1.5±0.3 | 0.02 (-0.35 to 0.40) | 0.93 (0.85–0.97) | 0.039 |
TBC: R-R intervals corrected by threshold-based correction method of Kubios HRV Premium (ver. 3.2).
Comparison of HRV measures calculated from MC R-R intervals between the ECG and polar (mean ± SD).
| HRV Measure | ECG (mean±SD) | Polar MC (mean±SD) | Bias (LoA) | ICC (95% CI) | Effect Size |
|---|---|---|---|---|---|
| SDNN (ms) | 55.5±26.7 | 55.3±26.7 | 0.21 (-0.70 to 1.12) | 1.00 (1.00–1.00) | 0.009 |
| RMSSD (ms) | 38.3±29.4 | 38.1±29.1 | 0.18 (-1.82 to 2.18) | 1.00 (0.99–1.00) | 0.006 |
| pNN50 (%) | 13.6±16.2 | 13.7±16.1 | -0.11 (-1.52 to 1.30) | 1.00 (0.99–1.00) | 0.007 |
| LF (ms2) | 1215±1352.1 | 1208.2±1351 | 6.85 (-65.77 to 79.48) | 1.00 (1.00–1.00) | 0.005 |
| HF (ms2) | 740.8±1262.8 | 730.9±1267.1 | 9.94 (-44.20 to 64.07) | 1.00 (1.00–1.00) | 0.008 |
| LF (nu) | 67.3±18.7 | 67.6±18.6 | -0.35 (-1.68 to 0.98) | 1.00 (1.00–1.00) | 0.019 |
| HF (nu) | 32.7±18.6 | 32.3±18.6 | 0.35 (-0.98 to 1.68) | 1.00 (1.00–1.00) | 0.019 |
| LF/HF Ratio | 3.6±4.0 | 3.7±4.0 | -0.04 (-0.23 to 0.14) | 1.00 (1.00–1.00) | 0.011 |
| SampEn | 1.5±0.2 | 1.5±0.2 | 0.01 (-0.11 to 0.12) | 0.97 (0.93–0.99) | 0.001 |
MC: R-R intervals corrected manually.