Literature DB >> 32953443

Effect of macro-design in the primary stability of short and extra-short implants using resonance frequency analysis. An ex vivo study.

Ramón Silva1, Pablo Villalón1, Felipe Cáceres1.   

Abstract

OBJECTIVE: This study aimed to determine the effect of macro design in the primary stability of short and extra-short implants using resonance frequency analysis (RFA).
MATERIAL AND METHODS: On an ex-vivo model using pig's ribs, we inserted 80 short and extra-short dental implants (20 implants per brand): Biohorizons®(B) 4.6 × 6mm; Intralock®(I) 4.75 × 6.5 mm; Straumann®(S) 4.1 × 4mm; and Tixos®(T) 5 × 5mm. Primary implant stability was measured using an RFA device. We compared mean ISQ values through ANOVA test.
RESULTS: Mean ISQ values: B = 73.36 (±3.39); I = 75.13 (±3.88); S = 65.38 (±8.38); T = 72.13 (±11). B and I showed higher ISQ than S (p-value < 0.001). Short (I) showed higher ISQ than extra-short (B,S,T) implants (p-value = 0.001). Tapered (B,I) had higher ISQ than parallel (S,T) implants (p-value < 0.001). There was a moderate positive correlation between ISQ and length (r = 0.52), and a weak correlation with diameter (r = 0.33). DISCUSSION: The final result is a combination of implant design, length, and diameter. Tapered design (B and I) and larger implants (I) showed better primary stability in terms of ISQ values. This information could be beneficial at implant selection in a severely reabsorbed low-quality bone, privileging length (as long as it is safe), and conical walls design.
© 2020 Craniofacial Research Foundation. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Extra short implant; Implant stability; Short implant

Year:  2020        PMID: 32953443      PMCID: PMC7486604          DOI: 10.1016/j.jobcr.2020.08.020

Source DB:  PubMed          Journal:  J Oral Biol Craniofac Res        ISSN: 2212-4268


  32 in total

1.  Role of clinician's experience and implant design on implant stability. An ex vivo study in artificial soft bones.

Authors:  Georgios E Romanos; Abdulaziz Basha-Hijazi; Bhumija Gupta; Yan-Fang Ren; Hans Malmstrom
Journal:  Clin Implant Dent Relat Res       Date:  2012-06-21       Impact factor: 3.932

Review 2.  Impact of implant length and diameter on survival rates.

Authors:  Franck Renouard; David Nisand
Journal:  Clin Oral Implants Res       Date:  2006-10       Impact factor: 5.977

Review 3.  Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation.

Authors:  Fawad Javed; Hameeda Bashir Ahmed; Roberto Crespi; Georgios E Romanos
Journal:  Interv Med Appl Sci       Date:  2013-12-20

4.  Ultrashort sintered porous-surfaced dental implants used to replace posterior teeth.

Authors:  Douglas Deporter; Bunnai Ogiso; Dong-Seok Sohn; Kevin Ruljancich; Michael Pharoah
Journal:  J Periodontol       Date:  2008-07       Impact factor: 6.993

Review 5.  Assessment of implant stability as a prognostic determinant.

Authors:  N Meredith
Journal:  Int J Prosthodont       Date:  1998 Sep-Oct       Impact factor: 1.681

Review 6.  Implant surface morphology and primary stability: is there a connection?

Authors:  Fawad Javed; Khalid Almas; Roberto Crespi; George E Romanos
Journal:  Implant Dent       Date:  2011-02       Impact factor: 2.454

Review 7.  Can Osseointegration Be Achieved Without Primary Stability?

Authors:  Mohanad Al-Sabbagh; Walied Eldomiaty; Yasser Khabbaz
Journal:  Dent Clin North Am       Date:  2019-04-15

8.  Correlation Between Resonance Frequency Analysis and Bone Quality Assessments at Dental Implant Recipient Sites.

Authors:  Min-Wen Fu; Earl Fu; Fu-Gong Lin; Wei-Jeng Chang; Yao-Dung Hsieh; E-Chin Shen
Journal:  Int J Oral Maxillofac Implants       Date:  2017 Jan/Feb       Impact factor: 2.804

9.  Comparative in vivo study of alloy titanium implants with two different surfaces: biomechanical and SEM analysis.

Authors:  Francisley Ávila Souza; Thayane Silveira Mata Furtado; Ulisses Ribeiro Campos Dayube; Willian Moraes Melo; Renato Sussumu Nishioka; Pier Paolo Poli; Carlo Maiorana; Paulo Sérgio Perri de Carvalho
Journal:  Clin Oral Investig       Date:  2019-04-10       Impact factor: 3.573

10.  Influence of Implant Design and Under-Preparation of the Implant Site on Implant Primary Stability. An In Vitro Study.

Authors:  Mariano Herrero-Climent; Bernardo Ferreira Lemos; Federico Herrero-Climent; Carlos Falcao; Helder Oliveira; Manuela Herrera; Francisco Javier Gil; Blanca Ríos-Carrasco; José-Vicente Ríos-Santos
Journal:  Int J Environ Res Public Health       Date:  2020-06-20       Impact factor: 3.390

View more
  2 in total

1.  Influence of Inter-Radicular Septal Bone Quantity in Primary Stability of Immediate Molar Implants with Different Length and Diameter Placed in Mandibular Region. A Cone-Beam Computed Tomography-Based Simulated Implant Study.

Authors:  Arshad Jamal Sayed; Safia Shoeb Shaikh; Shoeb Yakub Shaikh; Mohammed Abid Hussain; Sabahat Ullah Khan Tareen; Vaibhav Awinashe
Journal:  J Pharm Bioallied Sci       Date:  2021-06-05

2.  Primary Implant Stability Analysis of Different Dental Implant Connections and Designs-An In Vitro Comparative Study.

Authors:  Perry Raz; Haya Meir; Shifra Levartovsky; Alon Sebaoun; Ilan Beitlitum
Journal:  Materials (Basel)       Date:  2022-04-23       Impact factor: 3.748

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.