Leena Lehti1,2, Marcus Söderberg3,4, Helena Mellander1,5, Johan Wassélius1,5.
Abstract
PURPOSE: Onyx® embolization causes severe artifacts on subsequent CT-examinations, thereby seriously limiting the diagnostic quality.The purpose of this work was to compare the diagnostic quality of the tailored metal artifact reducing algorithms iMAR to standard reconstructions of CTA in patients treated with Onyx® embolization.Entities:
Keywords: Computed tomography; DECT, Dual Energy Computed Tomography; DLP, Dose Length Product; ED, Effective Dose; FOV, Field of View; Image quality; Iterative reconstruction algorithms; Metal artifact reduction; Onyx® embolization; PACS, Picture Archiving and Communication System; SD, Standard Deviation; VMI, Virtual Monoenergetic Images; iMAR, iterative Metal Artifact Reduction
Year: 2020 PMID: 32944593 PMCID: PMC7481136 DOI: 10.1016/j.ejro.2020.100255
Source DB: PubMed Journal: Eur J Radiol Open ISSN: 2352-0477
CT parameters for the arterial phase.
| Variables | |
|---|---|
| Reference effective mAs | 210/81 |
| Tube voltage kV | 80/Sn140 |
| Pitch | 0.55 |
| Rotation time (s) | 0.33 |
| Detector configuration | 128 × 0.6 |
| Nominal beam width (mm) | 38.4 |
| Convolution kernel | I30f |
| Matrix | 512 × 512 |
| Reconstructed slice thickness (mm) | 3 |
Patient characteristics (median and range).
| Variables | |
|---|---|
| Patients (n) | 12 |
| Age (years) | 74 (44−87) |
| Weight (kg) | 86 (60−132) |
| Height (cm) | 177 (160−191) |
| Body mass index (kg/m2) | 27 (21−41) |
| Plasma creatinine (µmol/L) | 87 (49−110) |
| Onyx® volume (ml) | 12 (1,5−19,5) |
Median and range for the attenuation (HU) and noise (SD) in the CT reconstructions.
| Diaphragm aorta | Artery close to Onyx | Psoas at the level of diaphragm | Psoas at the level of Onyx | |||||
|---|---|---|---|---|---|---|---|---|
| Reconstruction | HU | SD | HU | SD | HU | SD | HU | SD |
| Standard | 365 (217−452) | 25 (23−29) | 641 (410−1133) | 86 (40−124) | 46(35−57) | 19 (15−23) | 59 (17−156) | 35 (23−135) |
| Neuro coils | 365 (217−451) | 27 (24−37) | 280 (82−901) | 105 (28−209) | 46(35−57) | 19 (16−24) | 32 (14−52) | 22 (17−104) |
| Dental fillings | 364 (217−452) | 27 (23−30) | 401 (214−625) | 107 (34−291) | 47(36−56) | 19 (16−24) | 35 (11−63) | 27 (19−129) |
| Spine implants | 365 (217−452) | 26 (24−30) | 560 (335−1038) | 64 (28−182) | 46(35−57) | 19 (17−24) | 45 (15−78) | 33 (21−99) |
| Shoulder implants | 364 (217−451) | 26 (23−30) | 354 (101−980) | 114 (31−171) | 46(35−57) | 19 (17−23) | 35 (9−51) | 24 (18−48) |
| Pacemaker | 365 (217−452) | 26 (23−30) | 314 (124−863) | 95 (28−186) | 46(35−57) | 19 (17−23) | 35 (11−56) | 25 (19−51) |
| Thoracic coils | 364 (217−452) | 27 (23−30) | 397 (175−803) | 100 (39−152) | 46(36−57) | 20 (17−24) | 34 (9−52) | 26 (20−51) |
| Hip implants | 364 (217−452) | 26 (23−30) | 270 (195−794) | 95 (37−177) | 46(35−57) | 19 (17−24) | 33 (15−59) | 25 (19−49) |
| Extremity implants | 364 (217−452) | 26 (23−30) | 277 (226−869) | 90 (33−213) | 46(36−57) | 19 (17−24) | 33 (15−57) | 25 (19−50) |
Fig. 1Subjective image quality. Violin plots describing the subjective ratings for all 12 cases for overall image quality (panel A), severity of metal artifacts (panel B), ability to assess vessel patency (panel C), ability to assess adjacent tissues (panel D) and the ability to assess the integrity of stent structures (panel E). The dotted lines indicate the shift between non-diagnostic and diagnostic quality. Panel F shows average ratings for both readers for all 5 tasks for all tested reconstructions.
Fig. 3Examples of artifact reduction of the two iMAR algorithms favored by both radiologists in the blind image quality assessment (Hip implants, middle column and Extremity implants, right column), compared to the reconstruction without iMAR (Standard, left column). Please note the improvement regarding vessel patency, stent integrity and the reduction of streak artifacts in the aneurysmal sac and surrounding tissues.
Fig. 2An example of all tested reconstructions at the same level of the abdominal aorta following Onyx® embolization of a type 2 endoleak after. Standard reconstruction (without iMAR), and all examined iMAR reconstructions. Please note the partial erasing of stent structures in the Neuro coil reconstruction.