Iterative metal artifact reduction: evaluation and optimization of technique.

Naveen Subhas1, Andrew N Primak, Nancy A Obuchowski, Amit Gupta, Joshua M Polster, Andreas Krauss, Joseph P Iannotti.   

Abstract

OBJECTIVE: Iterative metal artifact reduction (IMAR) is a sinogram inpainting technique that incorporates high-frequency data from standard weighted filtered back projection (WFBP) reconstructions to reduce metal artifact on computed tomography (CT). This study was designed to compare the image quality of IMAR and WFBP in total shoulder arthroplasties (TSA); determine the optimal amount of WFBP high-frequency data needed for IMAR; and compare image quality of the standard 3D technique with that of a faster 2D technique.
MATERIALS AND METHODS: Eight patients with nine TSA underwent CT with standardized parameters: 140 kVp, 300 mAs, 0.6 mm collimation and slice thickness, and B30 kernel. WFBP, three 3D IMAR algorithms with different amounts of WFBP high-frequency data (IMARlo, lowest; IMARmod, moderate; IMARhi, highest), and one 2D IMAR algorithm were reconstructed. Differences in attenuation near hardware and away from hardware were measured and compared using repeated measures ANOVA. Five readers independently graded image quality; scores were compared using Friedman's test.
RESULTS: Attenuation differences were smaller with all 3D IMAR techniques than with WFBP (p < 0.0063). With increasing high-frequency data, the attenuation difference increased slightly (differences not statistically significant). All readers ranked IMARmod and IMARhi more favorably than WFBP (p < 0.05), with IMARmod ranked highest for most structures. The attenuation difference was slightly higher with 2D than with 3D IMAR, with no significant reader preference for 3D over 2D.
CONCLUSIONS: IMAR significantly decreases metal artifact compared to WFBP both objectively and subjectively in TSA. The incorporation of a moderate amount of WFBP high-frequency data and use of a 2D reconstruction technique optimize image quality and allow for relatively short reconstruction times.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25172218     DOI: 10.1007/s00256-014-1987-2

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  7 in total

1.  Frequency split metal artifact reduction (FSMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelrieß
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

2.  Normalized metal artifact reduction (NMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelriess
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

3.  Reduction of CT artifacts caused by metallic implants.

Authors:  W A Kalender; R Hebel; J Ebersberger
Journal:  Radiology       Date:  1987-08       Impact factor: 11.105

4.  Normalized metal artifact reduction in head and neck computed tomography.

Authors:  Michael M Lell; Esther Meyer; Michael A Kuefner; Matthias S May; Rainer Raupach; Michael Uder; Marc Kachelriess
Journal:  Invest Radiol       Date:  2012-07       Impact factor: 6.016

5.  Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation.

Authors:  Fabian Bamberg; Alexander Dierks; Konstantin Nikolaou; Maximilian F Reiser; Christoph R Becker; Thorsten R C Johnson
Journal:  Eur Radiol       Date:  2011-01-20       Impact factor: 5.315

6.  Metal artefact reduction from dental hardware in carotid CT angiography using iterative reconstructions.

Authors:  Fabian Morsbach; Moritz Wurnig; Daniel M Kunz; Andreas Krauss; Bernhard Schmidt; Spyros S Kollias; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2013-05-19       Impact factor: 5.315

7.  Reduction of metal artifacts from hip prostheses on CT images of the pelvis: value of iterative reconstructions.

Authors:  Fabian Morsbach; Sebastian Bickelhaupt; Guido A Wanner; Andreas Krauss; Bernhard Schmidt; Hatem Alkadhi
Journal:  Radiology       Date:  2013-03-19       Impact factor: 11.105

  7 in total
  20 in total

1.  Reduction of Metal Artifacts and Improvement in Dose Efficiency Using Photon-Counting Detector Computed Tomography and Tin Filtration.

Authors:  Wei Zhou; David J Bartlett; Felix E Diehn; Katrina N Glazebrook; Amy L Kotsenas; Rickey E Carter; Joel G Fletcher; Cynthia H McCollough; Shuai Leng
Journal:  Invest Radiol       Date:  2019-04       Impact factor: 6.016

2.  Society of Skeletal Radiology 2014 annual meeting summary.

Authors:  Michael G Fox; Laura W Bancroft
Journal:  Skeletal Radiol       Date:  2014-12-11       Impact factor: 2.199

3.  Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results.

Authors:  Seema Toso; Meryle Laurent; Elise Dupuis Lozeron; Pauline Brindel; Marirosa Cristallo Lacalamita; Sylviane Hanquinet
Journal:  Pediatr Radiol       Date:  2018-07-28

4.  Clinics in diagnostic imaging (167). Total shoulder arthroplasty glenoid component loosening with secondary rotator cuff failure.

Authors:  Tien Jin Tan; Ahmad Mohammad Aljefri; Marc Bruce Elliott; Savvas Nicolaou
Journal:  Singapore Med J       Date:  2016-04       Impact factor: 1.858

5.  Iterative metal artifact reduction improves dose calculation accuracy : Phantom study with dental implants.

Authors:  Manuel Maerz; Pia Mittermair; Andreas Krauss; Oliver Koelbl; Barbara Dobler
Journal:  Strahlenther Onkol       Date:  2016-03-11       Impact factor: 3.621

6.  Iterative metal artefact reduction (MAR) in postsurgical chest CT: comparison of three iMAR-algorithms.

Authors:  Joel Aissa; Johannes Boos; Lino Morris Sawicki; Niklas Heinzler; Karl Krzymyk; Martin Sedlmair; Patric Kröpil; Gerald Antoch; Christoph Thomas
Journal:  Br J Radiol       Date:  2017-08-22       Impact factor: 3.039

7.  Clinical evaluation of the iterative metal artefact reduction algorithm for post-operative CT examination after maxillofacial surgery.

Authors:  Arsany Hakim; Johannes Slotboom; Olivier Lieger; Fabian Schlittler; Roland Giger; Chantal Michel; Roland Wiest; Franca Wagner
Journal:  Dentomaxillofac Radiol       Date:  2017-03-07       Impact factor: 2.419

8.  Impact of different metal artifact reduction techniques on attenuation correction in 18F-FDG PET/CT examinations.

Authors:  Ole Martin; Joel Aissa; Johannes Boos; Katrin Wingendorf; David Latz; Christian Buchbender; Susanne Gaspers; Christina Antke; Martin Sedlmair; Gerald Antoch; Benedikt M Schaarschmidt
Journal:  Br J Radiol       Date:  2019-11-01       Impact factor: 3.039

9.  Relationship Between Glenoid Component Shift and Osteolysis After Anatomic Total Shoulder Arthroplasty: Three-Dimensional Computed Tomography Analysis.

Authors:  Eric T Ricchetti; Bong-Jae Jun; Yuxuan Jin; Jason C Ho; Thomas E Patterson; Jarrod E Dalton; Kathleen A Derwin; Joseph P Iannotti
Journal:  J Bone Joint Surg Am       Date:  2021-08-04       Impact factor: 6.558

10.  Dosimetric impact of using a commercial metal artifact reduction tool in carbon ion therapy in patients with hip prostheses.

Authors:  Jingfang Zhao; Weiwei Wang; Kambiz Shahnaz; Xianwei Wu; Jingfang Mao; Ping Li; Qing Zhang
Journal:  J Appl Clin Med Phys       Date:  2021-06-23       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.