| Literature DB >> 32818079 |
Benjamin Y Xu1, Michael Chiang2, Anmol A Pardeshi1, Sasan Moghimi3, Rohit Varma4.
Abstract
Purpose: To develop a deep neural network that detects the scleral spur in anterior segment optical coherence tomography (AS-OCT) images.Entities:
Keywords: angle closure; anterior segment OCT; artificial intelligence; deep learning
Year: 2020 PMID: 32818079 PMCID: PMC7395674 DOI: 10.1167/tvst.9.2.18
Source DB: PubMed Journal: Transl Vis Sci Technol ISSN: 2164-2591 Impact factor: 3.283
Figure 1.(a) Flow diagram of the ResNet-18 CNN architecture used to develop the model. Unmarked AS-OCT images were used as inputs to the CNN. The model predicted the Cartesian coordinates (X- and Y-locations) of the scleral spur. (b) Representative AS-OCT images used in CNN development. FC = fully connected; ReLu = rectified linear units.
Metrics of CNN Model Performance and Intragrader and Intergrader Variability in Open Angle, Angle Closure, and All Eyes (N = 921)
| CNN Prediction Error (µm) | Intragrader Variability (µm) | Intergrader Variability (µm) | |
|---|---|---|---|
| All eyes ( | |||
| X-coordinate | 1.23 ± 64.80 | 0.82 ± 70.84 | −43.56 ± 78.24 |
| Y-coordinate | −0.56 ± 62.10 | 8.07 ± 63.53 | −37.85 ± 73.49 |
| Absolute X-coordinate | 49.27 ± 42.07 | 52.31 ± 47.75 | 67.10 ± 59.27 |
| Absolute Y-coordinate | 47.73 ± 39.70 | 45.88 ± 45.06 | 63.59 ± 52.79 |
| Absolute overall | 73.08 ± 52.06 | 73.92 ± 60.72 | 97.34 ± 73.29 |
| Open angle eyes ( | |||
| X-coordinate | −1.06± 61.58 | 0.84± 67.95 | −44.29± 77.54 |
| Y-coordinate | −4.36± 57.11 | 3.19± 56.09 | −43.06± 66.07 |
| Absolute overall | 69.22± 47.70 | 67.37± 56.82 | 96.10± 70.37 |
| Angle closure eyes ( | |||
| X-coordinate | 8.41 ± 73.65 | 0.78 ± 79.39 | −41.27 ± 80.53 |
| Y-coordinate | 11.37 ± 74.54 µm | 31.60 ± 73.65 | −21.52 ± 91.22 |
| Absolute overall | 85.16 ± 62.41 | 94.44 ± 67.70 | 101.21 ± 81.81 |
Figure 2.Distribution of CNN-predicted and intraobserver errors in scleral spur location definition. (Top) Scatter plots of CNN-predicted errors (left) and intragrader errors (right). Mean error plotted in red. (Middle) Histograms of X-coordinate CNN prediction (left) and intraobserver (right) errors. (Bottom) Histograms of Y-coordinate CNN prediction (left) and intraobserver (right) errors.
Figure 3.Representative images of various absolute distances (50, 100, 150, and 200 µm) between reference grader-marked (cross) and CNN-predicted (circle) scleral spur locations.
Figure 4.Histograms of absolute CNN-predicted (left) and intraobserver (right) errors in scleral spur location.