| Literature DB >> 32789900 |
Yurii Shepelytskyi1,2, Tao Li3, Vira Grynko1,2, Camryn Newman4, Francis T Hane2,3, Mitchell S Albert2,3,5.
Abstract
PURPOSE: To test octafluorocyclobutane (OFCB) as an inhalation contrast agent for fluorine-19 MRI of the lung, and to compare the image quality of OFCB scans with perfluoropropane (PFP) scans THEORY AND METHODS: After normalizing for the number of signal averages, a theoretical comparison between the OFCB signal-to-noise ratio (SNR) and PFP SNR predicted the average SNR advantage of 90% using OFCB during gradient echo imaging. The OFCB relaxometry was conducted using single-voxel spectroscopy and spin-echo imaging. A comparison of OFCB and PFP SNRs was performed in vitro and in vivo. Five healthy Sprague-Dawley rats were imaged during single breath-hold and continuous breathing using a Philips Achieva 3.0T MRI scanner (Philips, Andover, MA). The scan time was constant for both gases. Statistical comparison between PFP and OFCB scans was conducted using a paired t test and by calculating the Bayes factor.Entities:
Keywords: fluorine-19; lung magnetic resonance imaging; octafluorocyclobutane; perfluoropropane
Mesh:
Substances:
Year: 2020 PMID: 32789900 PMCID: PMC7689774 DOI: 10.1002/mrm.28473
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668
Measured T1 and relaxation times and gradient image SNR of the studied gases
| T1 (ms) |
| GRE SNR (experimental values) | GRE SNR (normalized for NSA) | |
|---|---|---|---|---|
| OFCB | 28.5 ± 1.2 | 10.5 ± 1.8 | 45.52 | 45.52 |
| PFP | 18.6 ± 0.4 | 6.26 ± 0.27 | 30.26 | 30.26 |
| OFCB‐O2 | 20.4 ± 0.21 | 8.6 ± 0.5 | 14.52 | 14.52 |
| PFP‐O2 | 14.98 ± 0.61 | 5.4 ± 0.3 | 9.42 | 9.42 |
| OFCB‐O2 (in vivo) | 17.77 ± 1.5 | 3.4 ± 0.4 | 9.72 ± 2.1 (breath‐hold) | 0.61 ± 0.13 (breath hold) |
| 14.48 ± 4.51 (continuous breathing, 70°) | 0.1 ± 0.03 (continuous breathing, 70°) | |||
| 10.23 ± 0.70 (continuous breathing, 90°) | 0.39 ± 0.03 (continuous breathing, 90°) | |||
| PFP‐O2 (in vivo) | 12.8 ± 1.1 | 2.2 ± 0.3 | 7.66 ± 2.0 (breath‐hold) | 0.32 ± 0.08 (breath hold) |
| 12.68 ± 4.09 (continuous breathing, 70°) | 0.06 ± 0.02 (continuous breathing, 70°) | |||
| 8.81 ± 0.46 (continuous breathing, 90°) | 0.21 ± 0.01 (continuous breathing, 90°) |
Abbreviations: GRE, gradient echo; NSA, number of signal averages; O2, oxygen; OFCB, octafluorocyclobutane; PFP, perfluoropropane; SNR, signal‐to‐noise ratio.
FIGURE 1Theoretical dependence of in vivo octafluorocyclobutane–oxygen (OFCB‐O2) to perfluoropropane–oxygen (PFP–O2) signal‐to‐noise ratio (SNR) as a function of pulse repetition times (TRs). The SNR values were normalized on the number of signal averages (NSA) used for the animal scans (NSAOFCB = 16, NSAPFP = 24). The normalized SNR of PFP gas can excite the normalized SNR of OFCB only if TROFCB <7 ms. However, this value is impractical for a 70°‐Ernst angle; therefore, it will never be used for the real scans. The white dot represents the experimental results. It can be seen that the experimental result nicely agrees with theoretical calculations
FIGURE 2A, A representative inversion recovery curve measured for pure octafluorocyclobutane (OFCB). B, The measured spectra of 8 mL of pure perfluoropropane (PFP; blue) and OFCB (red). C, Spectra were obtained from 8 mL of PFP and OFCB breathing mixtures. The horizontal lines represent the maximum value of the corresponding magnetic resonance spectroscopy peak
FIGURE 3In vivo lung ventilation images of a healthy rat acquired in axial projections. The first column shows scans acquired during a single breath‐hold; the second column corresponds to the scans acquired during continuous breathing and using a 70°‐Ernst angle; the third column contains scans obtained during continuous breathing using a 90° flip angle (FA). The signal‐to‐noise ratio (SNR) of the single breath‐hold octafluorocyclobutane (OFCB) scan was 21% higher compared with the corresponding perfluoropropane (PFP) scan. For continuous breathing, the SNR of the OFCB image acquired using the Ernst angle of 70° was 15% stronger. Finally, during continuous breathing scans in the full recovery regime, the OFCB SNR exceeded PFP SNR by 17%. The red arrows indicate the chemical shift artifact associated with second spectral peak of PFP
FIGURE 4Non‐normalized signal‐to‐noise ratio (SNR) box charts of scans conducted using the single breath‐hold protocol, and continuous breathing protocol. The black boxes correspond to octafluorocyclobutane (OFCB) scans, and the blue boxes correspond to perfluoropropane (PFP) scans. The red squares illustrate the mean SNR of the group. Whiskers illustrate the standard deviation from the mean SNR. The SNR values of the OFCB scans were statistically significantly higher compared with the PFP scans. The corresponding P values and Bayes factors are shown in the figure legend. The wider scatter of SNR values obtained using the Ernst angle condition during continuous breathing can be explained by the absence of respiratory gating. The Ernst angle condition strongly depends on the T1 relaxation time of the gas in the lungs, and the absence of gating caused a T1 variation because of the fluorinated gas‐concentration differences during imaging