Ian Paddick1, Diana Grishchuk1, Alexis Dimitriadis1.
Abstract
PURPOSE: To compare planning indices achieved using manual and inverse planning approaches for Gamma Knife radiosurgery of arterio-venous malformations (AVMs). METHODS AND MATERIALS: For a series of consecutive AVM patients, treatment plans were manually created by expert planners using Leksell GammaPlan (LGP). Patients were re-planned using a new commercially released inverse planning system, IntuitivePlan. Plan quality metrics were calculated for both groups of plans and compared.Entities:
Keywords: Gamma Knife; inverse planning; radiosurgery; treatment planning
Mesh:
Year: 2020 PMID: 32755072 PMCID: PMC7497913 DOI: 10.1002/acm2.12973
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Patient demographics.
| Parameter | Mean | Range |
|---|---|---|
| Patient age (Years) | 37.4 | 11–68 |
| Gender m/f | 8/12 | |
| Target volume (cc) | 2.567 | 0.3–7.5 |
| Prescription dose (Gy) | 22.2 | 18–25 |
| Proportion of targets in proximity to OAR | 10% |
Fig. 1Transaxial T1 weighted magnetic resonance imaging slices with the manual (left) and inverse (right) plans for case number 6. The blue circles denote the isocentre size and position. The 18 Gy (prescription dose) and 12 Gy isodose lines are shown in yellow and green respectively. The improved concavity of the prescription isodose (red arrow) on the medial edge of target is shown on the inverse plan.
Mean values of manual and IntuitivePlan plan parameters. P‐values in bold if <0.05.
| Parameters | Manual plan | Intuitive Plan |
|
|---|---|---|---|
| Planning time (min) | 11.58 | 3.79 |
|
| Beam on time (min) | 49.52 | 57.97 |
|
| Planning plus beam on time (min) | 61.10 | 61.76 | 0.433 |
| Number of isocentres | 10.8 | 40.4 |
|
| Prescription isodose (%) | 45.35 | 50.85 |
|
| Coverage (%) | 96.3 | 96.8 |
|
| Selectivity | 0.856 | 0.884 | 0.071 |
| Paddick conformity index | 0.824 | 0.855 |
|
| Gradient index | 2.76 | 2.85 | 0.096 |
| Efficiency index (%) | 48.1 | 47.0 | 0.242 |
| Normal brain V12 (cc) | 5.79 | 5.81 | 0.497 |
| Mean brain (Skull) dose (Gy) | 0.35 | 0.35 | 0.500 |
Fig. 2Box plots showing planning parameters for the 20 cases: (a) Planning time; (b) Beam on time; (c) Planning and beam on time; (d) Number of isocentres; (e) Target coverage; (f) Selectivity; (g) Paddick Conformity Index; (h) Gradient Index; (i) Efficiency Index; (j) 12 Gy volume.