PURPOSE: This study compared 2 cone-beam computed tomography (CBCT) systems in the detection of mechanically simulated peri-implant buccal bone defects in dry human mandibles. MATERIALS AND METHODS: Twenty-four implants were placed in 7 dry human mandibles. Peri-implant bone defects were created in the buccal plates of 16 implants using spherical burs. All mandibles were scanned using 2 CBCT systems with their commonly used acquisition protocols: i-CAT Gendex CB-500 (Imaging Sciences, Hatfield, PA, USA; field of view [FOV], 8 cm×8 cm; voxel size, 0.125 mm; 120 kVp; 5 mA; 23 s) and Orthopantomograph OP300 (Intrumentarium, Tuusula, Finland; FOV, 6 cm×8 cm; voxel size, 0.085 mm; 90 kVp; 6.3 mA; 13 s). Two oral and maxillofacial radiologists assessed the CBCT images for the presence of a defect and measured the depth of the bone defects. Diagnostic performance was compared in terms of the area under the curve (AUC), accuracy, sensitivity, specificity, and intraclass correlation coefficient. RESULTS: High intraobserver and interobserver agreement was found (P<0.05). The OP300 showed slightly better diagnostic performance and higher detection rates than the CB-500 (AUC, 0.56±0.03), with a mean accuracy of 75.0%, sensitivity of 81.2%, and specificity of 62.5%. Higher contrast was observed with the CB-500, whereas the OP300 formed more artifacts. CONCLUSION: Within the limitations of this study, the present results suggest that the choice of CBCT systems with their respective commonly used acquisition protocols does not significantly affect diagnostic performance in detecting and measuring buccal peri-implant bone loss.
PURPOSE: This study compared 2 cone-beam computed tomography (CBCT) systems in the detection of mechanically simulated peri-implant buccal bone defects in dry human mandibles. MATERIALS AND METHODS: Twenty-four implants were placed in 7 dry human mandibles. Peri-implant bone defects were created in the buccal plates of 16 implants using spherical burs. All mandibles were scanned using 2 CBCT systems with their commonly used acquisition protocols: i-CAT Gendex CB-500 (Imaging Sciences, Hatfield, PA, USA; field of view [FOV], 8 cm×8 cm; voxel size, 0.125 mm; 120 kVp; 5 mA; 23 s) and Orthopantomograph OP300 (Intrumentarium, Tuusula, Finland; FOV, 6 cm×8 cm; voxel size, 0.085 mm; 90 kVp; 6.3 mA; 13 s). Two oral and maxillofacial radiologists assessed the CBCT images for the presence of a defect and measured the depth of the bone defects. Diagnostic performance was compared in terms of the area under the curve (AUC), accuracy, sensitivity, specificity, and intraclass correlation coefficient. RESULTS: High intraobserver and interobserver agreement was found (P<0.05). The OP300 showed slightly better diagnostic performance and higher detection rates than the CB-500 (AUC, 0.56±0.03), with a mean accuracy of 75.0%, sensitivity of 81.2%, and specificity of 62.5%. Higher contrast was observed with the CB-500, whereas the OP300 formed more artifacts. CONCLUSION: Within the limitations of this study, the present results suggest that the choice of CBCT systems with their respective commonly used acquisition protocols does not significantly affect diagnostic performance in detecting and measuring buccal peri-implant bone loss.
Peri-implantitis is characterized by an inflammatory process around an implant, which ranges from soft tissue inflammation to progressive bone loss.1 Image overlapping of adjacent anatomical structures on intraoral radiographs may prevent the detection of buccal and lingual peri-implant bone defects.23 In contrast, cone-beam computed tomography (CBCT) provides cross-sectional images of buccal and lingual plates.45678910 CBCT images have been described as more effective for the detection of peri-implant bone defects than conventional radiographic methods.459Despite the usefulness of CBCT in the detection of periimplant bone defects,4571011 beam-hardening artifacts from dental implants may affect its diagnostic performance.1213 In this context, 3 previous studies reported that CBCT was less accurate for assessing peri-implant bone loss than intraoral radiography.141516 Metal artifact reduction algorithms were tested to improve the detection of peri-implant defects on CBCT images, and also showed unclear results.9Several authors have tested whether different acquisition parameters on CBCT devices reduce artifact formation and improve the peri-implant defect detection rate.45711 Although the effects of field-of-view (FOV) size and number of frames have been well documented,47 the influence of other acquisition parameters such as kilovoltage peak (kVp) values in several acquisition protocols commonly used for dental implants remains unclear. To our knowledge, only 1 recent study has investigated the effectiveness of CBCT images obtained with different kVp settings in the detection of peri-implant defects. The aforementioned authors suggested that higher kVp values led to a higher detection rate of bone defects.11Thus, the aim of this study was to compare 2 different CBCT devices with their commonly used acquisition protocols for dental implants in the detection of mechanically simulated peri-implant buccal bone defects in dry human mandibles.
Materials and Methods
The present study was approved by the Ethics Committee of the School of Dentistry of the University of São Paulo, under number CAAE 67441817.7.0000.0075.An oral surgeon placed 24 cylindrical titanium dental implants to the alveolar bone crest level in 7 human dry edentulous mandibles. In order to avoid the influence of beam-hardening artifacts, a maximum of 4 implants with a minimum inter-implant distance of 20 mm were placed per mandible. All implants were placed in posterior mandibular sites (i.e., either first lower premolar or second lower molar sites). After implant placement, buccal bone defects were created with a spherical bur (3017HL; KG Sorensen, São Paulo, Brazil). Peri-circumferential defects with a diameter of 2.0 mm were created at 16 random implant sites, whereas the other 8 implant sites remained with no defects. All the defect depths were measured by a periodontal probe (North Carolina #15, Hu-Friedy, Chicago, IL, USA), and confirmed with a digital caliper. The depths were 0.5 mm, 1.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, 4.5 mm, 7.0 mm, and 15.0 mm, simulating differences in the severity of peri-implantitis, as shown in Figure 1.
Fig. 1
A. Implants placed to the alveolar bone crest level in the mandible. B. After implant placement, buccal bone defects are created with a spherical bur. C. The defect depths are measured by a periodontal probe.
CBCT imaging
Red wax was applied to the mandibles to simulate softtissue attenuation, as described previously.5 Two CBCT devices were used: i-CAT Gendex CB-500 (Imaging Sciences, Hatfield, PA, USA; FOV, 8 cm×8 cm; voxel size, 0.125 mm; 120 kVp; 5 mA; 23 s) and Orthopantomograph OP300 (Intrumentarium, Tuusula, Finland, FOV, 6 cm×8 cm; voxel size, 0.085 mm; 90 kVp; 6.3 mA; 13 s). The aforementioned acquisition protocols are among the most commonly used in implant dentistry for these 2 respective CBCT devices.The CBCT images were obtained by an oral and maxillofacial radiologist as DICOM files, and then exported as TIFF files using imaging software (Adobe Photoshop CS 8.0, Adobe Systems Inc., San Jose, CA, USA).
Data collection and image analysis
Two oral and maxillofacial radiologists participated in this study as observers. A total of 48 TIFF images of cross-sectional CBCT slices were selected and analyzed. The diagnostic performance of the 2 devices was also compared. Before image analysis, both observers were calibrated using representative pilot images. The differences between bone defects and artifacts were explained to both observers using presentation software (Microsoft PowerPoint for Mac; Microsoft Corporation, Redmond, WA, USA). The images used in this training exercise were discarded before initiating the image assessment of the present study.Both observers were asked to make qualitative assessments by rating the presence or absence of a bone defect according to a 5-point scale; 1: definitely absent, 2: probably absent, 3: uncertain, 4: probably present, 5: definitely present, as established in previous studies.457 For images that received a score of up to 4 points on the scale, the observers only analyzed the defects qualitatively. However, when the observers scored an image as 5 (definitely present), a quantitative measurement of the defect was made using software, as depicted in Figure 2.
Fig. 2
Quantitative defect depths are measured on software.
Image analyses were performed by the observers independently and at separate times, and the time allocated for interpretation was not limited. Both observers were blinded to the information on the sample images. Each implant was analyzed separately. All sequences of the observations were randomized and viewed randomly by both observers.
Statistical analyses
The kappa test was performed to assess intraobserver reliability and interobserver reproducibility for the detection of bone defects. Furthermore, intraclass correlation coefficients (ICCs) were calculated to determine the agreement of the linear measurements of bone defect depth performed by both observers. All CBCT-based linear measurements were compared with the measurements made using direct probes in the mandibles. The following ICC criteria were used to evaluate interobserver and intraobserver agreement: slight agreement, 0.00 to 0.20; fair agreement, 0.21 to 0.40; moderate agreement, 0.41 to 0.60; good agreement, 0.61 to 0.80; and excellent agreement, 0.81 to 1.00.17 The area under the curve (AUC), accuracy, sensitivity, and specificity were calculated using statistical software (ver. 24.0, IBM SPSS Statistics for Mac; IBM Corp., Armonk, NY, USA) at a level of significance of 5%.
Results
In total, 48 TIFF images of cross-sectional CBCT slices were selected and assessed. The results of the kappa test, which was used to compare interobserver and intraobserver agreement for the detection of bone defects, are available in Table 1 and showed considerable similarity. Both qualitative and quantitative measurements showed strong intraobserver and interobserver agreement, as described in Table 2. The lowest value was 0.70 and the highest value was 0.95, with statistical significance (P<0.05).
Table 1
Intraobserver and interobserver agreement (kappa) for the qualitative assessment of peri-implant defects
*P<0.05
Table 2
Intra- and interobserver agreement (intraclass correlation coefficients) for the quantitative assessment of peri-implant defects
*P<0.05
The AUC values for the observers are shown in Table 3. The OP300 device showed a slightly higher detection rate than the CB-500 (AUC 0.56±0.03), although the difference was not statistically significant. Table 4 shows the values for accuracy, sensitivity, and specificity. The highest accuracy was obtained by observer 1 (75.0%), and the lowest by observer 2 (70.8%). The sensitivity of both observers was the same (81.2%). For the OP300 device, the accuracy, specificity, and sensitivity of observers 1 and 2 were equal. A comparison of AUC values using the unpaired t-test with the Welch correction is shown in Table 5. There was no significant difference between both CBCT devices regarding the detection of defects (P>0.05).
Table 3
Area under the curve values from a receiver operating characteristic curve analysis
*P<0.05
Table 4
Diagnostic values with confidence intervals for the detection of peri-implant defects
*P value according to the Fischer's test
Table 5
Comparison of area under the curve values using the unpaired t-test with the Welch correction
Discussion
In the present study, dental implants were placed in dry human mandibles to analyze the presence of peri-implant buccal bone defects, in accordance with previous methodologies.518 Of note, controversial results have been reported in studies performed on animal bones.47111415 Some of the aforementioned studies assessed bone defects created by chemical means,411 to simulate peri-implant areas of bone resorption that are commonly irregular and have undefined edges. Nevertheless, the present diagnostic study was conducted on well-delimited defects created by burs with satisfactory homogeneity, enabling standardization of diagnostic performance tests, as described in previous studies.5715Regarding bone defect detection, the agreement values of a previous study19 ranged from 0.01 to 0.57, whereas our results ranged from 0.61 to 0.92. The sensitivity, specificity and AUC results shown in the present study also corroborate findings from a previous meta-analysis.19 The aforementioned authors found the following sensitivity and specificity rates for the diagnosis of CBCT defects: sensitivity, 28%–97%; specificity, 25%–97%; and AUC, 0.60–1.00. Their results are partially in accordance with those found in the present study (sensitivity, 81.2%; specificity, 50%–62.5%), whereas the AUC values (0.51–0.75) were slightly lower, as shown in Table 4.Schulze et al.12 summarized the theoretical constraints related to the formation of artifacts induced by titanium implants, based on simulations and geometrical considerations. The authors demonstrated the effects of varying kVp values, and reported that beam hardening was more prominent when there was lower energy related to kVp. Pauwels et al.20 investigated the effects of a variety of kVp and current settings on image quality, as well as those of different radiation doses administered by a CBCT device (3D Accuitomo 170; J. Morita Mfg. Corp., Kyoto, Japan) comparable to those studied herein (CB-500 and OP300), and reported optimal image quality at 90 kVp. CBCT images obtained at 75 kVp had more noise, thereby reducing defect detection. Similarly, differences in image resolution resulting from different voxel sizes and file resolutions have been described to influence linear measurements and diagnostic performance based on CBCT images. However, the difference in diagnostic performance found herein was not statistically significant, which contrasts with previous results in the literature.721Unlike the results observed by Pauwels et al.,20 corroborating the findings by Schulze et al.,12 the OP300 device presented better specificity than the CB-500 although the OP300 acquisition protocol (with lower kVp settings than in the CB-500 protocol) led to the formation of more artifacts. In general, a higher number of artifacts indicates lower image quality. However, although there were more artifacts in the OP300 images than in the CB-500 images, the OP300 diagnostic values were slightly higher than those of the CB-500 (Tables 1 to 4).Considering the values of interobserver and interobserver agreement, the results for sensitivity, specificity, and AUC could have been better if each observer had been free to manipulate the CBCT images in all 3 cross-sectional slices, instead of having to use only the sagittal image. This limitation was also reported in the study conducted by Dave et al.15 A recent study using 2 CBCT devices with different kVp values (90 kVp and 75 kVp)11 found that devices with a higher kVp setting yielded a higher rate of detection of peri-implant defects. In this study, it was not possible to obtain the same results with the higher kVp device (CB-500, 120 kVp); however, the results corroborate those found for the device with the same kVp that was evaluated in the previous study (OP300, 90 kVp).Researchers have observed that a higher kVp improves the quality of CBCT images in regard to the contrast-to-noise ratio.20 To achieve acceptable image quality with CBCT using a minimal radiation dose, current reduction is preferred over kVp reduction. In another study, the same authors15 showed that noise increased in protocols with the lowest current settings, thereby diminishing the quality of the CBCT image. However, the influence of current could not be evaluated in the present study, since the current acquisition protocols were too similar.In terms merely of the quality of the images obtained using the 2 devices, the images obtained using the CB-500 protocol had higher contrast than those obtained using the OP300 protocol (Fig. 3). This can be attributed to the higher kVp value for CB-500 than OP300. In images with higher contrast, different tissues or structures, such as those indicated by different shades of gray, are less sharply defined and therefore more difficult to distinguish; this may have influenced the results of the diagnostic tests. The attenuation processes of the X-ray beams in contact with structures generate a composite image of different shades of gray, and artifacts from metallic materials add to the non-homogeneity of the gray values.
Fig. 3
Differences in image quality. A. The CB-500 images have higher contrast than the OP300 images. B. In the image obtained using the OP300 protocol, artifact formation is pronounced.
Within the limitations of this study, the present results suggest that the choice of CBCT devices with their respective commonly used acquisition protocols does not significantly affect diagnostic performance in detecting and measuring peri-implant bone loss.
Authors: R Schulze; U Heil; D Gross; D D Bruellmann; E Dranischnikow; U Schwanecke; E Schoemer Journal: Dentomaxillofac Radiol Date: 2011-07 Impact factor: 2.419