Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device.

Sergio Lins de-Azevedo-Vaz1, Karla de Faria Vasconcelos, Frederico Sampaio Neves, Saulo Leonardo Sousa Melo, Paulo Sérgio Flores Campos, Francisco Haiter-Neto.   

Abstract

OBJECTIVE: To assess the accuracy of cone-beam computed tomography (CBCT) in periimplant fenestration and dehiscence detection, and to determine the effects of 2 voxel sizes and scan modes. STUDY
DESIGN: One hundred titanium implants were placed in bovine ribs in which periimplant fenestration and dehiscence were simulated. CBCT images were acquired with the use of 3 protocols of the i-CAT NG unit: A) 0.2 mm voxel size half-scan (180°); B) 0.2 mm voxel size full-scan (360°); and C) 0.12 mm voxel size full scan (360°). Receiver operating characteristic curves and diagnostic values were obtained. The Az values were compared with the use of analysis of variance.
RESULTS: The Az value for dehiscence in protocol A was significantly lower than those of B or C (P < .01). They did not statistically differ for fenestration (P > .05).
CONCLUSIONS: Protocol B yielded the highest values. The voxel sizes did not affect fenestration and dehiscence detection, and for dehiscence full-scan performed better than half-scan.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23217543     DOI: 10.1016/j.oooo.2012.10.003

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  21 in total

1.  Comparison of the influence of FOV sizes and different voxel resolutions for the assessment of periodontal defects.

Authors:  M E Kolsuz; N Bagis; K Orhan; H Avsever; K Ö Demiralp
Journal:  Dentomaxillofac Radiol       Date:  2015-04-22       Impact factor: 2.419

2.  Evaluation of the diagnostic efficacy of two cone beam computed tomography protocols in reliably detecting the location of the inferior alveolar nerve canal.

Authors:  Aditya Tadinada; Sydney Schneider; Sumit Yadav
Journal:  Dentomaxillofac Radiol       Date:  2017-03-13       Impact factor: 2.419

3.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

Review 4.  Use of cone beam computed tomography in periodontology.

Authors:  Buket Acar; Kıvanç Kamburoğlu
Journal:  World J Radiol       Date:  2014-05-28

5.  Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.

Authors:  Rocharles C Fontenele; Eduarda Hl Nascimento; Taruska V Vasconcelos; Marcel Noujeim; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2018-05-10       Impact factor: 2.419

6.  Comparison of cone-beam computed tomography with bitewing radiography for detection of periodontal bone loss and assessment of effects of different voxel resolutions: an in vitro study.

Authors:  Hayriye Cetmili; Melek Tassoker; Sevgi Sener
Journal:  Oral Radiol       Date:  2018-06-08       Impact factor: 1.852

7.  Difference in the artefacts production and the performance of the metal artefact reduction (MAR) tool between the buccal and lingual cortical plates adjacent to zirconium dental implant.

Authors:  Eduarda Helena Leandro Nascimento; Rocharles Cavalcante Fontenele; Gustavo Machado Santaella; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2019-07-10       Impact factor: 2.419

8.  Comparison of CBCT with different voxel sizes and intraoral scanner for detection of periodontal defects: an in vitro study.

Authors:  Murat Icen; Kaan Orhan; Çiğdem Şeker; Gediz Geduk; Fethiye Cakmak Özlü; Murat İnanç Cengiz
Journal:  Dentomaxillofac Radiol       Date:  2020-03-10       Impact factor: 2.419

9.  Comparison between different cone-beam computed tomography devices in the detection of mechanically simulated peri-implant bone defects.

Authors:  Jun Ho Kim; Reinaldo Abdala-Júnior; Luciana Munhoz; Arthur Rodriguez Gonzalez Cortes; Plauto Christopher Aranha Watanabe; Claudio Costa; Emiko Saito Arita
Journal:  Imaging Sci Dent       Date:  2020-06-18

10.  Diagnostic Accuracy of Three Cone Beam Computed Tomography Systems and Periapical Radiography for Detection of Fenestration Around Dental Implants.

Authors:  Amir Eskandarloo; Samira Saati; Mahbubeh Purabdolahi Ardakani; Mohamadreza Jamalpour; Naser Mohammad Gholi Mezerji; Vahid Akheshteh
Journal:  Contemp Clin Dent       Date:  2018 Jul-Sep
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.