Jae Yong Han1, Sangchul Yoon1,2, Nicolas Scott Brown3, Sueng Han Han1, Jinu Han4.
Abstract
PURPOSE: To compare refractive error measured by hand-held wavefront aberrometers with postcycloplegic autorefraction (AR) and cycloplegic refraction (CR).Entities:
Keywords: Corneal wavefront aberration; Mydriatics; Refractive errors; Refractometry
Mesh:
Year: 2020 PMID: 32495531 PMCID: PMC7269738 DOI: 10.3341/kjo.2019.0132
Source DB: PubMed Journal: Korean J Ophthalmol ISSN: 1011-8942
Fig. 1The measurement of portable wavefront aberrometer. (A) Measurement with the portable OVITZ P10 wavefront aberrometer. (B) Once the device had been positioned properly, the examiner was able to see the array of Hartmann-Shack images on the screen. The array of Hartmann-Shack spots should be clearly focused, and contained within the pupil.
Mean values of power vectors measured by three different measurement techniques
Values are presented as vectors (diopters); Values in parentheses indicate 1 standard error of mean.
*Calculated from repeated measure analysis of variance.
Fig. 2Bland-Altman plot showing the differences between wavefront aberrometer and postcycloplegic autorefraction (AR) (left), wavefront aberrometer and cycloplegic refraction (CR) (right) in 51 patients at each power vectors (M, J, J). Agreement between the wavefront aberrometer and postcycloplegic AR, (A) the mean M difference was −0.115 diopters (D), and the limits of agreement (LOA) was −1.578 to 1.348. (B) For J, the mean difference was 0.239 D, and the LOA was −0.371 to 0.850. (C) For J, the mean difference was −0.015 D, and the LOA was −0.768 to 0.738. Between the wavefront and CR, (D) the mean M difference was −0.220 D, and the LOA was −1.790 to 1.350. (E) The mean difference was 0.300 D and the LOA was −0.526 to 1.127 for J. (F) The mean difference for J was −0.079 D and the LOA was −0.662 to 0.504.
Comparison of axes of the cylindrical components between wavefront aberrometer, postcycloplegic AR, and CR in subjects with cylinder 0.75 diopters or more with AR (n = 26)
Values are presented as number (%).
AR = autorefraction; CR = cycloplegic refraction.