Literature DB >> 14640434

Relationship between cycloplegic and wavefront-derived refraction.

Luis E Fernández de Castro1, Helga P Sandoval, Oday Al Sarraf, David T Vroman, Kerry D Solomon.   

Abstract

PURPOSE: To compare cycloplegic refractions to preoperative wavefront-derived refractions from three Hartmann-Shack-based wavefront devices.
METHODS: Ninety-nine eyes of 55 subjects were evaluated preoperatively using the LADARWave, WaveScan, and Zywave aberrometers. With cycloplegic refraction as the standard, the spherical equivalent refraction, sphere, and cylinder were compared to the wavefront-derived refractions using the three wavefront devices.
RESULTS: The LADARWave unit showed the highest correlation between wavefront-derived sphere and mean spherical equivalent refraction values relative to cycloplegic refraction, followed by Zywave and WaveScan. The WaveScan unit exhibited higher variability than the LADARWave or Zywave. Although Alcon Laboratories (LADARWave) and Bausch & Lomb Surgical (Zywave) recommend dilation, VISX specifically recommends against dilation in their WaveScan unit. All devices showed high cylinder error vectors when the cycloplegic cylinder was greater than 1.00 D.
CONCLUSION: Although wavefront-derived refractions appeared to correlate well to cycloplegic refractions, further research is required to reduce variability and improve the calculation of cylinder terms.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14640434     DOI: 10.3928/1081-597X-20031101-12

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Validity of autorefractor based screening method for irregular astigmatism compared to the corneal topography- a cross sectional study.

Authors:  Alicia Galindo-Ferreiro; Julita De Miguel-Gutierrez; Manuel González-Sagrado; Alberto Galvez-Ruiz; Rajiv Khandekar; Silvana Schellini; Julio Galindo-Alonso
Journal:  Int J Ophthalmol       Date:  2017-09-18       Impact factor: 1.779

2.  An evaluation of the Bausch & Lomb Zywave aberrometer.

Authors:  Michael J Dobos; Michael D Twa; Mark A Bullimore
Journal:  Clin Exp Optom       Date:  2009-05       Impact factor: 2.742

3.  Accuracy of the Hand-held Wavefront Aberrometer in Measurement of Refractive Error.

Authors:  Jae Yong Han; Sangchul Yoon; Nicolas Scott Brown; Sueng Han Han; Jinu Han
Journal:  Korean J Ophthalmol       Date:  2020-06
  3 in total