Nicholas J Durr1, Shivang R Dave2, Daryl Lim2, Sanil Joseph3, Thulasiraj D Ravilla3, Eduardo Lage2,4.
Abstract
OBJECTIVE: To assess the quality of eyeglass prescriptions provided by an affordable wavefront autorefractor operated by a minimally trained technician in a low-resource setting. METHODS AND ANALYSIS: 708 participants were recruited from consecutive patients registered for routine eye examinations at Aravind Eye Hospital in Madurai, India, or an affiliated rural satellite vision centre. Visual acuity (VA) and patient preference were compared between trial lenses set to two eyeglass prescriptions from (1) a novel wavefront autorefractor and (2) subjective refraction by an experienced refractionist.Entities:
Keywords: diagnostic tests/investigation; optics and refraction; public health; vision
Year: 2019 PMID: 31276029 PMCID: PMC6579572 DOI: 10.1136/bmjophth-2018-000225
Source DB: PubMed Journal: BMJ Open Ophthalmol ISSN: 2397-3269
Figure 1Testing procedure for the wavefront autorefractor. Participants looked through the open-view wavefront autorefractor at a distant back-lit visual acuity chart, while three 10 s videos of wavefront images were recorded by the device. The autorefractor was flipped over to measure the opposite eye. After repeating three times, the system displayed the autorefractor eyeglass prescription.
Participant demographics and average refractive error measurements
| Gender | n (%) | Age±SD (range) (years) | Subjective refraction | Objective refraction | ||||
|
|
|
|
|
|
| |||
| Female | 413 | 34±12 | −0.48 | −0.02 | 0.00 | −0.55 | 0.00 | 0.04 |
| Male | 296 | 37±14 | −0.59 | −0.03 | 0.00 | −0.71 | −0.03 | 0.04 |
5%, 5th percentile; 95%, 95th percentile; D, dioptres; J, vertical Jackson cross cylinder; J, oblique Jackson cross cylinder; M, spherical equivalent.
Figure 2Correlation and Bland-Altman plots of power vectors measured by autorefractor versus subjective refraction. Correlation (left) and Bland-Altman (right) plots comparing agreement of prescriptions measured by subjective refraction and the prototype autorefractor for (A) spherical equivalent, M, (B) vertical Jackson cross cylinder, J, and (C) oblique Jackson cross cylinder, J. AR, autorefractor prescription; SR, subjective refraction prescription; D, dioptre.
Trends in absolute difference of subjective refraction versus autorefraction of right eye measurements
| All right eyes | Female | Male | Hospital | Vision centre | |||||||||||
| Avg (D) | m (D/year) | b (D) | Avg (D) | m (D/year) | b (D) | Avg (D) | m (D/year) | b (D) | Avg (D) | m (D/year) | b (D) | Avg (D) | m (D/year) | b (D) | |
| | | 0.482 | 0.001 | 0.43 | 0.448 | 0.001 | 0.43 | 0.529 | 0.002 | 0.47 | 0.492 | 0.001 | 0.46 | 0.455 | 0.002 | 0.39 |
| | | 0.110 | 0.002 | 0.04 | 0.111 | 0.003 | 0.02 | 0.109 | 0.001 | 0.06 | 0.116 | 0.001 | 0.07 | 0.096 | 0.004 | −0.03 |
| | | 0.063 | 0.001 | 0.05 | 0.061 | 0.001 | 0.03 | 0.067 | 0.000 | 0.06 | 0.077 | 0.000 | 0.07 | 0.028 | 0.000 | 0.02 |
AR, autorefractor prescription; Avg, average; D, dioptre; J, vertical Jackson cross cylinder; J, oblique Jackson cross cylinder; M, spherical equivalent; SR, subjective refraction prescription; b, intercept of linear fit; m, slope of linear fit.
Figure 3Box plot of visual acuity before and after correction. Visual acuity of the right eyes without correction (VA), with trial lenses set to the autorefractor-determined prescription (VA) and with trial lenses set to the subjective refraction-determined prescription (VA). There was a statistically significant difference (p<0.01) between average visual acuity measurements among all combinations within each age group (indicated by *). logMAR, logarithm of the minimum angle of resolution.
Participant preference of trial lens prescriptions with masked origin
| Age group | Participants, n (%) | P value | |||
| All | No preference | Preferred SR | Preferred AR | ||
| 15–40 | 438 (61.9) | 96 (21.9) | 174 (39.7) | 168 (38.4) | 0.52 |
| 41–70 | 270 (38.1) | 82 (30.4) | 123 (45.6) | 65 (24.1) | <0.01 |
| All | 708 (100.0) | 178 (25.1) | 297 (41.9) | 233 (32.9) | <0.01 |
AR, autorefractor prescription; SR, subjective refraction prescription.