| Literature DB >> 32411432 |
Masayuki Ouchi1,2, Takuya Shiba3.
Abstract
PURPOSE: To evaluate the clinical outcome of blended vision combined with micromonovision (MMBV) using a +4 diopter (D) addition power (add) diffractive intraocular lens (MIOL) and a +2.75D add MIOL with a myopic target of -0.5 D.Entities:
Year: 2020 PMID: 32411432 PMCID: PMC7204239 DOI: 10.1155/2020/6143832
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Patient characteristics and baseline data.
| Parameter | Control | MMBV | |
|---|---|---|---|
| Mean age (y) | 64.2 ± 9.2 (47–79) | 68.7 ± 7.4 (46–78) |
|
| Sex (men/women) | 13/17 | 12/18 |
|
| UDVA (logMAR) | 0.63 ± 0.59 (−0.08–2.0) | 0.64 ± 0.53 (−0.17–1.70) |
|
| CDVA (logMAR) | 0.15 ± 0.23 (−0.08–2.0) | 0.14 ± 0.23(−0.17–0.70) |
|
| SE (D) | −2.37 ± 7.00 (−14.1–3.25) | −1.93 ± 4.46 (−10.5–3.5) |
|
| CA (D) | 0.71 ± 0.48 (0–1.75) | 0.61 ± 0.52 (0–1.25) |
|
| AL (mm) | 25.38 ± 1.63 (21.69–27.01) | 24.13 ± 1.71 (21.10–28.23) |
|
UDVA, uncorrected distance visual acuity; CDVA, best-corrected distance visual acuity; logMAR: logarithm of the minimum angle of resolution; SE, manifest spherical equivalent; CA, corneal astigmatism; D, diopter. Student's t-test, §Fisher's exact probability test, ∫Mann–Whitney's U test, and †Welch's t-test.
Postoperative refractive data.
|
|
ZMB, ZMB00-implanted eye; ZKB, ZKB00-implanted eye; SE, manifest spherical equivalent; RS, refractive spherical equivalent; CA, corneal astigmatism. Steel's multiple comparison test and ∫Kruskal–Wallis's test.
Postoperative visual acuity.
| Monocular | Control (60 eyes) | MMBV ZMB (30 eyes) | MMBV ZKB (30 eyes) |
|
|---|---|---|---|---|
| UDVA | −0.05 ± 0.13 | −0.01 ± 0.15 | 0.05 ± 0.23 |
|
| CDVA | −0.17 ± 0.28 | −0.10 ± 0.22 | −0.12 ± 0.07 |
|
| UNVA | 0.17 ± 0.18 | 0.19 ± 0.20 | 0.22 ± 0.18 |
|
| CNVA | 0.04 ± 0.11 | 0.04 ± 0.10 | 0.04 ± 0.10 |
|
| Binocular | Control (60 eyes) | MMBV (60 eyes) | ||
| UDVA | −0.10 ± 0.07 | −0.09 ± 0.11 |
| |
| CDVA | −0.12 ± 0.07 | −0.15 ± 0.13 |
| |
| UNVA | 0.02 ± 0.14 | 0.06 ± 0.13 |
| |
| CNVA | 0.02 ± 0.12 | 0.03 ± 0.10 |
|
ZMB, ZMB00-implanted eye; ZKB, ZKB00-implanted eye; UDVA, uncorrected distance visual acuity; CDVA, best-corrected distance visual acuity; UNVA, uncorrected near visual acuity; CNVA, best-corrected near visual acuity. Kruskal–Wallis's test and ∫Mann–Whitney's U test.
Figure 1(a) Uncorrected binocular multidistance VA test. The MMBV group had significantly better intermediate VA at 70 cm than the control group (p=0.02, Mann–Whitney's U test) while no significant difference in near VA at 30 cm and 40 cm was detected. (b) Distance-corrected binocular multidistance VA test. The MMBV group had significantly worse near VA at 30 and 40 cm but better intermediate VA at 70 cm than the Control group (p=0.04, p=0.05, and p=0.02, respectively, Mann–Whitney's U test).
Figure 2Binocular contrast sensitivity. The MMBV group had significantly better contrast sensitivity at 12 cpd than the control group (p=0.05, Mann–Whitney's U test).
Figure 3Binocular low-contrast VA with 12.5% contrast and 6.0% contrast. The MMBV group had significantly better VA at both 12.5% and 6.0% contrast than the control group (p=0.05 and p=0.04, respectively, Welch's t-test).
Postoperative outcomes.
| Number (%) |
| ||
|---|---|---|---|
| Control | MMBV | ||
|
| |||
| Never | 29 (96.7) | 30 (100) | 0.32 |
| Occasional | 1 (3.3) | 0 (0) | |
| Always | 0 (0) | 0 (0) | |
|
| |||
|
| |||
| Never | 25 (83.3) | 23 (76.7) | 0.49 |
| Occasional | 5 (6.7) | 6 (20.0) | |
| Always | 0 (0) | 1 (3.3) | |
|
| |||
|
| |||
| Severe | 0 (0) | 0 (0) | 0.86 |
| Moderate | 6 (20) | 4 (13.3) | |
| Slight | 7 (23.3) | 11 (36.7) | |
| None | 17 (56.7) | 15 (50) | |