Thuy Duong Do1, Stephan Rheinheimer1, Hans-Ulrich Kauczor1,2, Wolfram Stiller1,2, Tim Weber1, Stephan Skornitzke3.
Abstract
OBJECTIVES: To quantitatively and qualitatively evaluate image quality in dual-layer CT (DLCT) compared to single-layer CT (SLCT) in the thorax, abdomen, and pelvis in a reduced-dose setting.Entities:
Keywords: Abdomen; Radiation exposure; Thorax; Tomography, X-ray computed
Mesh:
Year: 2020 PMID: 32394278 PMCID: PMC7476988 DOI: 10.1007/s00330-020-06894-7
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1Flowchart illustrating study inclusion and exclusion criteria. In total, 31 patients could be included into the study with repeated CT scans consisting of 51 SLCTlow, 39 DLCThigh, and 34 DLCTlow
CT acquisition parameters for SLCTlow, DLCTlow, and DLCThigh
| SLCTlow | DLCTlow | DLCThigh | |
|---|---|---|---|
| Tube potential | 100 kVp | 120 kVp | 120 kVp |
| Automated tube current modulation (dose right index) | 14 | 13 | 17 |
| Predicted CTDIvol | 7.8 mGy | 7.5 mGy | 11.5 mGy |
| Predicted average mAs | 174 mAs | 74 mAs | 116 mAs |
| Minimum mAs | 65 mAs | 65 mAs | 65 mAs |
| Maximum mAs | 300 mAs | 300 mAs | 350 mAs |
Fig. 2Image examples of regions of interest (ROIs) used for the quantitative evaluation in the thorax at the height of the tracheal bifurcation (a), the upper abdomen at the height of the portal vein (b), and the lower abdomen at the level of the lumbar spine L4 (c). Rectangle for measurements of the lateral and anterior-posterior diameter of the torso
Fig. 3Image examples from the same patient from conventional reduced-dose SLCTlow (column 1), DLCTlow (column 2), and DLCThigh (column 3) from thorax (a), upper abdomen (b), and pelvis (c). DLCThigh (column 2) showed best quantitative and qualitative analysis whereas SLCTlow (column 1) and DLCTlow (column 3) showed comparable results
Evaluation of patient radiation exposure: means of scan length, CTDIvol, DLP, and SSDE of SLCTlow, DLCThigh, and DLCTlow
| Scan length (mm) | CTDIvol (mGy) | DLP (mGy·cm) | SSDE Thorax (mGy) | SSDE upper abdomen (mGy) | SSDE lower abdomen (mGy) | |
|---|---|---|---|---|---|---|
| SLCTlow | 640.71 ± 39.52 | 5.74 ± 1.26 | 408.58 ± 107.58 | 5.00 ± 1.20 | 8.41 ± 1.79 | 7.28 ± 1.72 |
| DLCTlow | 626.91 ± 38.77 | 6.57 ± 0.72 | 444.68 ± 57.63 | 7.28 ± 0.66 | 9.19 ± 1.04 | 8.30 ± 0.66 |
| DLCThigh | 642.04 ± 37.18 | 9.34 ± 2.13 | 647.08 ± 155.12 | 7.92 ± 0.91 | 14.11 ± 2.62 | 11.52 ± 1.96 |
Evaluation of automatic exposure control: means and standard deviations of patients’ lateral and anterior-posterior (AP) diameter, applied tube current-time product at the measurement location, and results of statistical testing. Statistical significance of results is indicated by italics
| Anatomic location | SLCTlow (*) | DLCTlow (**) | DLCThigh (***) | |||
|---|---|---|---|---|---|---|
| * vs. ** | * vs. *** | ** vs. *** | ||||
| Thorax | ||||||
| Lateral (mm) | 389.9 ± 31.3 | 379.6 ± 31.7 | 384.3 ± 31.6 | 0.53 | ||
| AP (mm) | 231.3 ± 31.3 | 229.6 ± 28.5 | 237.5 ± 31.0 | 0.70 | 0.74 | 0.51 |
| mAs | 98.8 ± 29.1 | 66.1 ± 3.0 | 74.3 ± 12.6 | 0.11 | ||
| Upper abdomen | ||||||
| Lateral (mm) | 326.5 ± 23.6 | 320.4 ± 25.9 | 326.3 ± 27.1 | 0.86 | 0.052 | |
| AP (mm) | 247.2 ± 29.8 | 242.1 ± 32.0 | 252.7 ± 35.9 | 0.56 | 0.20 | 0.087 |
| mAs | 151.9 ± 41.5 | 77.4 ± 14.4 | 123.9 ± 35.4 | |||
| Lower abdomen | ||||||
| Lateral (mm) | 348.5 ± 36.9 | 339.3 ± 36.0 | 342.7 ± 39.5 | 0.88 | ||
| AP (mm) | 236.6 ± 31.3 | 233.7 ± 34.4 | 240.9 ± 35.4 | 0.68 | 0.46 | 0.77 |
| mAs | 174.1 ± 58.5 | 80.7 ± 25.4 | 116.9 ± 34.4 | |||
Fig. 4Example of tube current-time product curves for the three evaluated acquisition protocols in one patient. Note the changes in tube current in the different anatomical regions caused by the automated exposure control, as well as the changes during the rotation of the x-ray tube around the patient table
Results of the quantitative image quality evaluation: image noise and CNR. Statistical significance of results is indicated by italics
| Quantitative parameters | SLCTlow (*) | DLCTlow (**) | DLCThigh (***) | |||
|---|---|---|---|---|---|---|
| Anatomic location | * vs. ** | * vs. *** | ** vs. *** | |||
| Thorax | ||||||
| Noise (fat) (HU) | 5.4 ± 2.0 | 5.2 ± 2.6 | 5.0 ± 2.2 | 0.29 | 0.08 | 0.53 |
| CNR (lung) | − 157.8 ± 52.7 | − 165.4 ± 45.6 | − 173.9 ± 51.5 | 0.20 | 0.33 | |
| Upper abdomen | ||||||
| Noise (fat) (HU) | 5.9 ± 1.6 | 6.1 ± 2.2 | 5.6 ± 1.6 | 0.32 | 0.61 | 0.16 |
| CNR (liver) | 41.7 ± 12.7 | 38.6 ± 12.3 | 39.8 ± 11.8 | 0.05 | 0.10 | 0.71 |
| Lower abdomen | ||||||
| Noise (fat) (HU) | 5.1 ± 1.5 | 5.1 ± 1.2 | 4.6 ± 1.1 | 0.51 | ||
| CNR (psoas) | 37.3 ± 10.8 | 35.0 ± 8.6 | 39.5 ± 10.4 | 0.18 | 0.12 | |
Results of the quantitative image quality evaluation: CT numbers of evaluated organs and vessels in the thorax, upper abdomen, and lower abdomen. Statistical significance of results is indicated by italics
| Quantitative parameters | SLCTlow (*) | DLCTlow (**) | DLCThigh (***) | |||
|---|---|---|---|---|---|---|
| Anatomic location | * vs. ** | * vs. *** | ** vs. *** | |||
| Thorax | ||||||
| Subcutaneous fat (HU) | − 115.7 ± 12.0 | − 110.5 ± 12.9 | − 113.0 ± 12.1 | 0.23 | ||
| Lung (HU) | − 875.5 ± 39.8 | − 877.1 ± 43.6 | − 877.1 ± 49.2 | 0.54 | 0.80 | 0.42 |
| Trachea (HU) | − 987.8 ± 14.0 | − 991.9 ± 7.2 | − 993.9 ± 7.6 | 0.58 | ||
| Upper abdomen | ||||||
| Subcutaneous fat (HU) | − 114.4 ± 10.4 | − 104.7 ± 14.3 | − 106.6 ± 16.2 | 0.81 | ||
| Liver (HU) | 113.0 ± 18.7 | 106.7 ± 18.8 | 99.3 ± 22.3 | |||
| Portal vein (HU) | 173.9 ± 21.3 | 151.5 ± 20.3 | 152.2 ± 22.4 | 0.27 | ||
| Aorta (HU) | 161.9 ± 23.2 | 143.8 ± 19.2 | 142.7 ± 22.6 | 0.51 | ||
| Spleen (HU) | 118.1 ± 11.9 | 107.3 ± 11.7 | 106.8 ± 16.0 | 0.62 | ||
| Lower abdomen | ||||||
| Subcutaneous fat (HU) | − 118.0 ± 8.7 | − 109.4 ± 10.2 | − 111.5 ± 10.4 | 0.51 | ||
| Psoas muscle (HU) | 59.3 ± 8.2 | 58.6 ± 5.4 | 59.5 ± 6.0 | 0.16 | 0.93 | 0.21 |
Fig. 5Results of the qualitative analysis showing the percentage of amount of received ratings, with best image quality observed for DLCThigh. Image quality for DLCTlow was equal to SLCTlow