Utilization of virtual mono-energetic images (MonoE) derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans.

Tilman Hickethier1, Jonathan Byrtus1, Myriam Hauger1, Andra-Iza Iuga1, Gregor Pahn2, David Maintz1, Stefan Haneder1, Jonas Doerner3.   

Abstract

OBJECTIVES: To investigate the utilization of virtual mono-energetic images (MonoE) at low kiloelectron volt (keV) levels derived from a dual-layer spectral detector CT (SDCT) for the assessment of abdominal arteries in venous contrast phase scans using arterial phase imaging as an internal reference standard.
MATERIALS AND METHODS: A total of 50 patients who received arterial and venous phase imaging of the abdomen on a SDCT system were included in this study. Absolute attenuation, noise, signal- and contrast to noise ratios (SNR; CNR) as well as arterial diameters in defined landmarks were assessed. In arterial phase, conventional reconstructions (CRART) as well as MonoEART at 40keV and in venous phase, conventional reconstructions (CRVEN) as well as MonoEVEN at 70 and 40keV were investigated and intra-individual comparisons were performed. If an artery stenosis (10 patients) was present, the degree of stenosis was assessed according to the system of the North American Symptomatic Carotid Endarterectomy Trial (NASCET).
RESULTS: MonoE 40keV yielded significantly higher attenuation values (in arterial as well as in venous phase) compared to CRART (p<0.001) while noise levels were substantially low. This resulted in markedly superior SNR and CNR in large vessel compared to CRART. Luminal diameters were significantly smaller in MonoE 40keV in both contrast phases compared to CRART (p<0.001), whereas no significant differences were found between both MonoE reconstructions (p≥0.92). The degree of vessel stenosis was significantly higher in MonoE 40keV of both contrast phases compared to CRART (p≥0.02).
CONCLUSION: MonoE at low keV of venous contrast phase scans derived from a novel SDCT are suitable for the assessment of arteries in the abdomen and subsequent stenosis assessment. However, MonoE at 40keV constantly showed significant smaller luminal diameters than the corresponding conventional reconstructions (including the reference standard). This is possibly due to an improved differentiation of the vessel lumen from the wall and raises the question, which imaging technique should be used as an appropriate reference standard for vascular SDCT imaging studies.
Copyright © 2017 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Abdominal imaging; Artery stenosis; Dual energy CT; Spectral detector CT; Venous contrast phase

Mesh:

Year:  2017        PMID: 29362148     DOI: 10.1016/j.ejrad.2017.12.007

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  6 in total

1.  Improved Sensitivity and Reader Confidence in CT Colonography Using Dual-Layer Spectral CT: A Phantom Study.

Authors:  Markus M Obmann; Chansik An; Amanda Schaefer; Yuxin Sun; Zhen J Wang; Judy Yee; Benjamin M Yeh
Journal:  Radiology       Date:  2020-07-28       Impact factor: 11.105

2.  The application of dual-layer spectral detector computed tomography in solitary pulmonary nodule identification.

Authors:  Qingyun Wen; Yong Yue; Jin Shang; Xiaomei Lu; Lu Gao; Yang Hou
Journal:  Quant Imaging Med Surg       Date:  2021-02

3.  Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma.

Authors:  Yasunori Nagayama; Shota Tanoue; Taihei Inoue; Seitaro Oda; Takeshi Nakaura; Daisuke Utsunomiya; Yasuyuki Yamashita
Journal:  Eur Radiol       Date:  2019-07-16       Impact factor: 5.315

Review 4.  Dual-energy CT of acute bowel ischemia.

Authors:  Markus M Obmann; Gopal Punjabi; Verena C Obmann; Daniel T Boll; Tobias Heye; Matthias R Benz; Benjamin M Yeh
Journal:  Abdom Radiol (NY)       Date:  2021-06-30

5.  Utilization of virtual low-keV monoenergetic images generated using dual-layer spectral detector computed tomography for the assessment of peritoneal seeding from ovarian cancer.

Authors:  Taek Min Kim; Sang Youn Kim; Jeong Yeon Cho; Seung Hyup Kim; Min Hoan Moon
Journal:  Medicine (Baltimore)       Date:  2020-06-05       Impact factor: 1.817

6.  Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting.

Authors:  Thuy Duong Do; Stephan Rheinheimer; Hans-Ulrich Kauczor; Wolfram Stiller; Tim Weber; Stephan Skornitzke
Journal:  Eur Radiol       Date:  2020-05-11       Impact factor: 5.315

  6 in total